Miura 200
1Pages

{{requestButtons}}

Catalog excerpts

Miura 200 - 1

■ Up to 420 tests per hour with ISE module on board. ■ 240 Bi-reagent photometric tests/hour constant throughput. ■ Up to 360 tests per hour with ISE module on board. ■ Up to 200 mono reagent tests per hour. Common Specifications ■ Analytical methods: end point mono or bichromatic, fixed time, kinetic, differential sample blank. ■ ISE module for: Na+, K+, Li+, CI- in serum, plasma ■ 30/43 reagents on board plus diluent. ■ Mono, double reagents method. ■ Automatic dilution of the samples. ■ Calibration with linear/non linear curves up to 8 points, auto dilution of the calibration curve (linear regression, linear piece-wise, cubic spline, polynomial). ■ Control menu with Levy-Jennings plot. ■ Bar-code reader integrated for samples ■ Arm equipped with liquid level detector (capacitive) and shock sensor. ■ High precision micro-metering pump to dispense reagents and samples (no syringe). ■ Reagent volume: 1+400 microliters. ■ Sample volume: 1+300 microliters. ■ Typical reaction volume: 200-250 microliters. ■ Wavelengths on board: 8 from 340nm to 700nm plus one free position. ■ Light source: Long life halogen lamp (2000 hours). ■ Reading cuvette: 80 on board re-washable ■ Cuvette washing: automatic washing station. ■ Water consumption: 3/4 liters per hour. ■ Reagent cooler: 14°C below ambient temperature. ■ All in one touch screen computer. ■ Operative system: Windows© 8.1

Open the catalog to page 1

All ISE catalogs and technical brochures

  1. Reagents

    8 Pages

  2. Hemo One

    8 Pages

  3. Miura Family

    20 Pages

Archived catalogs

  1. Hemo One

    4 Pages