Fume cupboards 2018
37Pages

{{requestButtons}}

Catalog excerpts

Fume cupboards 2018 - 1

Köttermann Fume Cupboards. Individually configurable.

Open the catalog to page 1
Fume cupboards 2018 - 2

We exceed standards for you. Individually configurable efficiency. Subject to technical modifications Status: January 2018

Open the catalog to page 2
Fume cupboards 2018 - 3

Contents 02 Overview of fume cupboard types 04 Which fume cupboard for which requirement? 06 Bench-type fume cupboards 09 Energy-efficient fume cupboards 12 Walk-in and distillation fume cupboards 16 Height-adjustable fume cupboards 19 High-performance fume cupboards 22 Radionuclide fume cupboards 25 Pharmacy hood 28 Substructures and accessories 31 Fume cupboard electronics 33 Modularity 35 Contact

Open the catalog to page 3
Fume cupboards 2018 - 4

Fume cupboard typesFlexible and individually configurable When dealing with volatile hazardous substances, dusts or aerosols, the protection of employees is our top priority. Such works must therefore be carried out in a fume cupboard. KOTTERMANN fume cupboards not only achieve top results in all relevant tests, but also offer users ergonomic advantages that make work processes easier and thus even safer. o Complete range the right fume cupboard for every application o Made from steel high quality and long lasting o Perfect safety certified by independent institutes o Ergonomic functions to...

Open the catalog to page 4
Fume cupboards 2018 - 5

Equipment options o Certification of all EXPLORIS® standard fume cupboards according to DIN EN 14175 o Optimized exhaust air system in terms of flow technology o Excellent containment capacity for pollutants o Sash made of laminated safety glass o Especially low operating costs o Lots of space for devices due to particularly large interior depth o Easy operation and safe monitoring through integrated EXPLORIS TouchTronic® * o Support of various applications through individual gas, water and electrical services o Tested for 20 years of operation ° Electric cables free from halogenated...

Open the catalog to page 5
Fume cupboards 2018 - 6

Which fume cupboard for which requirement? KOTTERMANN fume cupboards for every application Fume cupboards can be used universally in the laboratory. They ensure optimum protection against hazardous substances. The selection of the appropriate fume cupboard type depends both on the room conditions of the laboratory and on the type of chemicals used. Especially when using acids, the following recommendations should be followed. Fume cupboards for general use (Bench-type, distillation, walk-in fume cupboard, EXPLORIS VarioTop®, EXPLORIS EcoPlus®) suitable occasional use not suitable Organic...

Open the catalog to page 6
Fume cupboards 2018 - 7

High-performance fume cupboards with PP lining (polypropylene) / fume cupboard for hydrofluoric acid High thermal load

Open the catalog to page 7
Fume cupboards 2018 - 8

Bench-type fume cupboards All-rounders for general use A particularly energy-saving variant of the fume cupboard for general use is the EXPLORIS EcoPlus® More information on page 9

Open the catalog to page 8
Fume cupboards 2018 - 9

07 Dimensions EXPLORIS® Bench-type fume cupboard high / low design EXPLORIS® Bench-type fume cupboards Dimensions in mm | Volume flow in m³ / h Width Front height Housing height Usable space width Usable space height Usable space depth Working height

Open the catalog to page 9
Fume cupboards 2018 - 10

Equipment Options o Certified according to DIN EN 14175 Part 2 and Part 3 o With innovative EXPLORIS TouchTronic® * as standard o Low operating costs due to low airflow o Extra large interior depth due to special worktop shape EXPLORIS® bench-type fume cupboards for general use can be used universally in the laboratory. They are ideal for dealing with organic solvents and are also suitable for working with cold or diluted acids. Depending on the room height of the laboratory, there is a choice between a high and a low design. o Sash options - one-piece sash (fume cupboards with standard...

Open the catalog to page 10
Fume cupboards 2018 - 11

Fume cupboard conventional EXPLORIS EcoPlus® Saving EXPLORIS EcoPlus® Operating time per f.c (width 1 500 mm) Assumption: 1000 m3 of processed laboratory air cost 0.80 Euro.

Open the catalog to page 11
Fume cupboards 2018 - 12

Dimensions in mm | Volume flow in m3 / h Usable space depth 760 760 760 760 760 Minimum permissible volume flow ** 300 375 450 530 530 Volume flow to be adjusted 360 450 540 630 630 ' Alarm value

Open the catalog to page 12
Fume cupboards 2018 - 13

EXPLORIS EcoPlus® reduces the consumption of expensive conditioned air in the laboratory by up to 30 percent compared to conventional fume cupboards. This saves energy, protects the environment and reduces the operating costs per fume cupboard by up to € 1,000 per year. The ventilation system offers further savings potential: Smaller pipe cross-sections and smaller dimensioned fans can save up to 20 percent on investment costs. Bench-type fume cupboards are available as EXPLORIS EcoPlus® and can be retrofitted. Certified according to DIN EN 14175 Part 2 and Part 3 With innovative EXPLORIS...

Open the catalog to page 13
Fume cupboards 2018 - 14

Walk-in and distillation fume cupboards High experimental setups securely contained

Open the catalog to page 14
Fume cupboards 2018 - 15

Dimensions EXPLORIS® walk-in fume cupboard EXPLORIS® walk-in fume cupboards Dimensions in mm | Volume flow in m³ / h Width Front height Housing height Usable space width Usable space height Usable space depth

Open the catalog to page 15
Fume cupboards 2018 - 16

Dimensions EXPLORIS® distillation fume cupboard / V Dimensions in mm | Volume flow in m3 / h Usable space depth 625/750 625/750 625/750 625/750 625/750 Minimum permissible volume flow ** 600 750 900 1050 1050 Volume flow to be adjusted 630 790 950 1 110 1 110 ' Alarm value

Open the catalog to page 16
Fume cupboards 2018 - 17

Equipment options o Certified according to DIN EN 14175 Part 2 and Part 3 o With innovative EXPLORIS TouchTronic® * as standard o Extremely high interior space for particularly large experimental setups EXPLORIS® Walk-in and distillation fume cupboards are used when a large working height is required, for example in distillations or preparative column chromatography. Distillation fume cupboards are equipped with a low worktop. In addition, walk-in fume cupboards have the advantage that big and heavy equipment can be moved on a mobile table directly into the chamber of the fume cupboard. *...

Open the catalog to page 17
Fume cupboards 2018 - 18

Height-adjustable fume cupboards EXPLORIS VarioTop®

Open the catalog to page 18
Fume cupboards 2018 - 19

Dimensions EXPLORIS VarioTop® Dimensions in mm | Volume flow in m3 / h Usable space width ** 910 / 1 100 Usable space depth ** 680 / 770 Minimum permissible volume flow *** 540 Volume flow to be adjusted 600 Maximum permissible volume flow 1 500 to / from 960 mm interior height *** Alarm value

Open the catalog to page 19