OP_medinSINDI_1080_REV14
36Pages

{{requestButtons}}

Catalog excerpts

OP_medinSINDI_1080_REV14 - 1

Operating, cleaning and maintenance instructions medinSINDI® Universal Gas Supply Unit with Pressure Monitor

Open the catalog to page 1
OP_medinSINDI_1080_REV14 - 2

Contact and ordering information medin Medical Innovations GmbH Adam-Geisler-Str. 1 D-82140 Olching Sales and customer service: by authorised partners a detailed list can be found on the website www.medingmbh.com Copyright  This document is the exclusive property of medin Medical Innovations GmbH. No part of this document may be copied or otherwise reproduced or stored in any form on electronic media without prior written consent of the company, except in the case of express authorisation from the company. Contact the manufacturer for further information. Information on the validity and...

Open the catalog to page 2
OP_medinSINDI_1080_REV14 - 3

3. This warranty does not cover: - regular maintenance and repair or replacement of parts due to normal wear and tear - consumables (components which could be expected to need regular replacement over the course of the life of the product) - damage or defects caused by use, operation or handling of the product not in accordance with the intended use specified in the instructions for use - damage or changes to the product caused by: d) failure to follow the manufacturer's instructions for use and e) failure to follow the manufacturer's care and maintenance f) connecting or using...

Open the catalog to page 3
OP_medinSINDI_1080_REV14 - 4

2.8. Overview of the General Technical Description 10 2.9. Accessories for medin medinSINDI® REF 1080 10 3.3. Gas Supply Connection (see Figure 3) 11 3.4.4. Interruption of the External Power Supply 12 4.3.6. Oxygen Concentration Adjustment Trend 18 4.4.3.3. (De-)activation of "Acoustic alarm off" Function 21 5.2. Monitoring Alarms in the medinSINDI® CPU Monitor: 23 5.3. Error messages during operation 24 medin Medical Innovations GmbH Adam-Geisler-Str. 1 D-82140 Olching

Open the catalog to page 4
OP_medinSINDI_1080_REV14 - 5

6.5. Power Supply Unit (Cleaning, Maintenance, Repairs and Modifications) 25 6.6.1. Every 2 months - oxygen calibration: 26 6.7. medinSINDI® - Basic Construction of the Monitor and the Mixer 27 Figure 1: Screws for attachment to the rolling cart rail 11 Figure 2: Symbol to indicate the charging process 12 Figure 3: Connection to Medijet® 14 Figure 4: medinSINDI® front 15 Figure 5: medinSINDI® Monitor 16 Figure 6: Oxygen concentration trend 18 Figure 7: Special function buttons 19 Figure 8: Service Menu 20 Figure 9: Oxygen calibration 20 Figure 10:...

Open the catalog to page 5
OP_medinSINDI_1080_REV14 - 6

Warning - Handling - Do not operate the device in the vicinity of highly flammable gases Do not use any oil or lubricant on or near the oxygen equipment Do not place any containers with liquids on the device Oxygen may only be supplied via the connection provided The oxygen supply must be monitored This equipment may be changed without prior notice The medinSINDI® may not come into contact with flammable anaesthesia gases or other flammable materials while in use. Electromagnetic Compatibility It is confirmed that the device meets the EMC guidelines 89/336/EEC, 93/68/EEC and 93/42/EEC....

Open the catalog to page 6
OP_medinSINDI_1080_REV14 - 7

Product Description medinSINDI® universal gas supply unit with pressure monitor is an electronic unit with microprocessorcontrolled monitoring and is used to display the set parameters. The unit is battery-operated and is charged by an external power supply plug. The screen is backlit. The operating time without mains power is approximately 5 hours. The oxygen concentration is determined in the gas flow delivered. The O 2 cell is delivered ready to use by a known manufacturer. This cell generates an analogous DC voltage proportional to the oxygen concentration. The cell itself does not need...

Open the catalog to page 7
OP_medinSINDI_1080_REV14 - 8

2. Product Specifications2.1. Modes - Automatic shutoff of the pressure monitor if no CPAP use is taking place - 21-100% continuous adjustment of the oxygen concentration - Reset for alarms - 2-minute alarm silencing (make silent or end): Warning - Alarm Silencing Silencing the alarm requires additional monitoring of vital signs! - Confirmation of the nCPAP alarm setting for automatic monitoring - Confirmation of the FiO2 concentration setting for automatic monitoring (± 10% of the O2) - Disconnection or leakage alarm (if the patient pressure falls below 1.5 cm H2O) - Alarm...

Open the catalog to page 8
OP_medinSINDI_1080_REV14 - 9

2.5. Power Supply The electronic system is fully battery-operated and can be recharged with an external power supply plug. Uninterruptible operation for an intensive care unit, with an external power supply plug (automatic 100 to 230 V AC) - Input voltage, external power supply plug, mains adapter: 100 - 230 V AC - Input frequency mains adapter 50 - 60 Hz - Power input: 400 mA - Output voltage 24 V DC/625 mA - Internal battery: rechargeable- NiMh 9.6 volts/1100 mA - Internal battery life: approx. 5 hours (when fully charged) - Battery charging time - max. 5 hours 2.6. Atmosphere and...

Open the catalog to page 9
OP_medinSINDI_1080_REV14 - 10

2.7. Dimensions and Weight - Dimensions (height x width x depth): 180 x 240 x 145 mm (without connections and holder) - Weight: 2.6 kg (without holder) 2.8. Overview of the General Technical Description • Information on these points can be found in the chapters indicated: • Conditions for use, transport and storage: see chapter 3.7 • Characteristics and accuracies of the device: see chapter 2 • Information on the installation of the device: see chapter 3 • Description of the supply gases required: see chapter 3.3 • Description of the power supply: see chapter 3.4 • The device...

Open the catalog to page 10
OP_medinSINDI_1080_REV14 - 11

Attach device to wall/rolling cart rail mounting and secure in place: Loosen the knurled head screws of the rail clamps on the back of the device Hang device on wall rail or rolling cart rail Tighten the knurled head screws. Knurled head screws Figure 1: Screws for attachment to the rolling cart rail The medinSINDI® can be transported when it is not in use; however, when it is in use, it must be securely mounted. However, if the medinSINDI® becomes damaged during this transport or due to rough handling, its function must be checked by a service technician. It is not allowed to be used until...

Open the catalog to page 11

All Medin Medical Innovations GmbH catalogs and technical brochures

  1. Nuflow®

    2 Pages

  2. Pb_1000_long

    3 Pages

  3. OP_Medijet

    103 Pages

  4. Pb_4000_long

    3 Pages

  5. OP_Miniflow

    103 Pages

  6. OP Bubble

    2 Pages

  7. Processing

    4 Pages

  8. Pb_1207_long

    2 Pages

  9. PB_1207MKI

    2 Pages

  10. PB_1207

    2 Pages

Archived catalogs

  1. MEDINsindi

    2 Pages

  2. Medin_Catalog

    16 Pages

  3. OP_Blender

    18 Pages

  4. OP_Medijet_1000

    104 Pages

  5. products_medin

    16 Pages