VetPro® 5000 - 1

Go to Table of Contents Next Go to Table of Contents Next VetPro® 5000 WHIP Delivery Dental Systems Model Numbers: Stationary Units Service and Parts Manual Mobile Units VetPro® Solo FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY

Open the catalog to page 1
VetPro® 5000 - 2

access procedures Operation & Troubleshooting General Info Section A Testing & Repair - continued GENERAL INFORMATIOn (*) Indicates multiple pages due to model / serial number break (s).

Open the catalog to page 2
VetPro® 5000 - 3

General Information Ordering Parts Indicates an imminently hazardous situation which will result in serious or fatal injury if not avoided. This symbol is used only the most extreme conditions. Non-warranty parts orders may be faxed to Midmark using the Fax Order Form in the back of this manual. The following information is required when ordering parts: • Serial number & model number • Part number for desired part (Refer to Section E: Exploded Views & Parts Lists) Indicates a potentially hazardous situation which could result in serious injury if not avoided. Indicates a potentially...

Open the catalog to page 3
VetPro® 5000 - 4

Specifications Mobile Units Stationary Units cart Cabinet/Table Width WARNING The Acteon Newtron® LED scaler complies with current electromagnetic compatibility standards. However, it must not be used by or on those with cardiac stimulators or other active implants unless adequate electromagnetic interference shielding is present. check with a cardiologist to verify the EMI shielding capabilities of the implant. See enclosed Newtron® Led manual for further advisories. WARNING / • \ Equipment not suitable for use in the presence of a flammable anesthetic mixture. © Midmark Corporation 2008...

Open the catalog to page 4
VetPro® 5000 - 5

WARNING Equipment not suitable for use in the presence of a flammable anesthetic mixture. © Midmark Corporation 2008 SF-1930 [Revised: 5/10/12] v

Open the catalog to page 5
VetPro® 5000 - 6

© Midmark Corporation 2008 SF-1930 [Revised: 5/1/09] vi

Open the catalog to page 6
VetPro® 5000 - 7

Warranty Limited Warranty SCOPE OF WARRANTY Midmark Corporation (“Midmark”) warrants to the original purchaser The VetPro® 5000 delivery systems (except for components not warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark’s obligation under this warranty is limited to the repair or replacement, at Midmark’s option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective. APPLICABLE...

Open the catalog to page 7
VetPro® 5000 - 8

Operation & Troubleshooting [Revised: dd/mo/yr]

Open the catalog to page 8
VetPro® 5000 - 9

Theory of Operation - Mobile and Solo Unit Compressor When the compressor is running, atmospheric air is drawn in through the intake filter to the compressor head where air is compressed and discharged to a storage tank. A check valve is placed between the compressor head and the storage tank to prevent air flow back to the head. The solenoid allows air to escape for approximately two seconds during start-up to relieve pressure from the head, then closes sending pressurized air to the storage tank. When the pressure in the storage tank reaches a predefined level, the pressure switch is...

Open the catalog to page 9
VetPro® 5000 - 10

Theory of Operation continued - Mobile and Solo Unit Electrical • Supply line voltage is always present at the mobile unit power cord receptacle when plugged into a wall receptacle. When the On/ Off switch is turned on, line voltage, 115 VAC, is supplied to the pressure switch, fan and receptacles (accessory, scaler & fiber optics). • The system will continue to run until the pressure inside the system reaches 100 PSI (+ / - 5 PSI) (Cut-out time). It will turn back on when the pressure lowers to 80 PSI (+ / - 5 PSI) (Cut-in time). • The timer keeps the solenoid turned off until the...

Open the catalog to page 10
VetPro® 5000 - 11

Theory of Operation - Stationary Units When the master On/Off switch is turned to On, supply air passes through the On/Off switch valve to the water master pilot valve block, the air master pilot valve block, syringe, water bottle and foot control pressure indicator. Wall Mount Shown Air Flow The air pilot valve block directs air through two air regulators. One regulates air to 30 PSI and supplies the water bottles with air pressure. The other supplies the foot control with air regulated at 80 PSI. When the foot control pedal is depressed, the valve (in foot control) opens and supplies...

Open the catalog to page 11
VetPro® 5000 - 12

Basic Operation - Handpieces The automatic kink valves activate a handpiece when it is removed from the resting position and brought forward of the instrument head. Drive air will be delivered to the withdrawn handpiece when the foot control is depressed. Resting Position Open Position Step 1: Turn master switch on. Step 3: Lift the handpiece and pull it forward of the instrument head. Depress foot control. Note: If necessary, make adjustments with the control knobs as desired. Refer to: Section B, Handpiece Adjustments Step 2: Flip wet/dry toggle on foot control to desired position....

Open the catalog to page 12
VetPro® 5000 - 13

Troubleshooting Chart - WHIP Delivery Head Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 03/10/09] A-6 Always verify model & serial number

Open the catalog to page 13
VetPro® 5000 - 14

Troubleshooting Chart - WHIP Delivery Head Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 03/10/09] A-7 Always verify model & serial number

Open the catalog to page 14
VetPro® 5000 - 15

Troubleshooting Chart - WHIP Delivery Head Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 6/3/13] A-8 Always verify model & serial number

Open the catalog to page 15
VetPro® 5000 - 16

© Midmark Corporation 2008 SF-1930 [Revised: 6/3/13] A-9 Always verify model & serial number

Open the catalog to page 16
VetPro® 5000 - 17

Troubleshooting Chart - WHIP Delivery Head Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 6/3/13] A-10 Always verify model & serial number

Open the catalog to page 17
VetPro® 5000 - 18

Troubleshooting Chart - WHIP Delivery Head Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 6/3/13] A-11 Always verify model & serial number

Open the catalog to page 18
VetPro® 5000 - 19

Troubleshooting Chart Problem Models: Serial Numbers: © Midmark Corporation 2008 SF-1930 [Revised: 6/3/13] A-12 Always verify model & serial number

Open the catalog to page 19

All Midmark catalogs and technical brochures

  1. ANIMAL HEALTH

    138 Pages

  2. imaging

    23 Pages

  3. Synthesis

    2 Pages

  4. VetPro® CR

    4 Pages

  5. iqvitals

    8 Pages

  6. seating

    7 Pages

  7. Anesthesia

    5 Pages

  8. Steri-Center

    6 Pages

  9. Soniclean®

    4 Pages

  10. Midmark M3

    6 Pages

Archived catalogs