EasyOne Pro LAB - 2 / 2 Pages

Features
EasyOne Pro LAB offers improved efficiency and quality
of care by providing complete pulmonary function
testing on site

Specifications
Flow/Volume Sensor
Type: Ultrasonic transit-time

• Multiple-Breath N2 washout for FRC, RV, TLC, LCI
• Single Breath CO Diffusion (DLCO)

Range: ±16 l/s
Resolution: 4 ml/s

• Spirometry

Accuracy: ±2% or 0.02 l/s

• Flexible EMR capabilities

Resistance: < 1.5 cmH2O/l/s at 12 l/s

• No warm-up time
• High-resolution color touch screen
• Automatic and fast calibration for each test performed
• Automated ATS/ERS interpretation
• Challenge testing with manitol and methacholine protocols
• Wide range of data export options

CO Sensor
Type: Non-dispersive infrared
Range: 0 to 0.35%
Resolution: 0.0001%
Accuracy: < 0.001%

• Easy raw data export for scientific use
• No service contracts

Tracer Gas Sensor (Helium, N2)

• Easy software upgrades

Type: Ultrasonic transit-time

• Remote support capability

Range: 0 to 50%,
Resolution: 0.01%,

ATS/ERS Interpretation Tree

Accuracy: 0.05%,

FEV1/VC > LLN

Yes
TLC > LLN

No

TLC > LLN

Yes

No

TLC > LLN

TLC > LLN

No

Normal
DLCO > LLN
Yes

Normal

No

Restriction

No

PV
disorders

Obstruction

DLCO > LLN
Yes

CW and NM
disorders

Mixed
defect

ILD
Pneumonitis

Yes

Asthma
CB

0.2%

Type: Non-dispersive infrared
Range: 0 to 15%
Resolution: 0.005%
Accuracy: ±0.05% (from 0 to 5%)

DLCO > LLN
No

0.05%

CO2 Sensor

No
Yes

0 to 100%

No

Emphysema

Standards
Quality System: ISO 9001, ISO 13485
FDA: 510(k) market clearance
MDD 93/42/EEC: CE marked
Electrical: EN 60601-1, EN 60601-1-2
Electrical and Environmental Requirements
Supply: 100 to 240 VAC, 50/60 Hz
Temperature: 10 to 40°C

Single breath DLCO

Humidity: 30 to 75% (non condensing)

measurement

nddmedical

nddmedical

nddmed

nddmedical

ndd Medical Technologies, Inc. • Two Dundee Park, Andover, Massachusetts 01810 • Phone: 877-904-0090 • www.nddmed.com
© Copyright 2013 ndd Medical Technologies, Inc. All rights reserved. rev. 3/13

EasyOne Pro LAB
  1. P. 1

    A<|><|>EasyOne Pro® LABMultiple-breath nitrogen washout to obtain complete lung volumes,FRC and LCI (lung clearance index)
  2. P. 2

Text version

All ndd catalogues and technical brochures

  1. EasyOne Pro

    2 Pages

    En
  2. EasyOne Plus

    2 Pages

    En
  3. Easy on-PC

    2 Pages

    En