• Products
  • Catalogues
  • News & Trends
  • Exhibitions

Specification Sheet QBC STAR - 1 Pages

  1. P. 1

Catalogue excerpts

QBC STAR Dry Hematology Analyzer TM Mean Corpuscular Hemoglobin Concentration (MCHC) 25 - 37.3 g/dL White Blood Cell Count (WBC) Granulocyte Count (Gran #) Granulocyte % (Gran %) Lymphocyte and Monocyte Count (Lymph/Mono #) Lymphocyte and Monocyte % (Lymph/Mono %) Mean Value White Blood Cell Count (WBC) (× 109/L) Granulocyte % (Gran %) Lymphocyte and Monocyte % (Lymph/Mono %) The precision data represent the analysis of eleven whole blood specimens, each assayed in replicates of ten. Electrical Specifications (Power Pack) Voltage Input Frequency Current Output Power Output General Specifications Dimensions Weight Noise Sample Volume Display W 16” x D 16.3” x H 16.3” (40.6 cm x 41.4 cm x 41.4 cm) 19 lbs (8.6 kg) <70 dB @ 3 ft 70 µL LCD Display Resolution 160 x 160 58 mm thermal recorder paper Connectivity USB Ethernet (Wired) Operating Environment Non-Operating Temperature Requirements Operating Temperature Requirements -2 °F to 122 °F (-20 °C to 50 °C) 16 °C – 32 °C unrestricted (at > 32 °C use may be limited by instrument temperature shutdown or some results may be suppressed) Part Number Innovative Solutions for a Healthier World QBC Diagnostics and QBC STAR are trademarks of QBC Diagnostics. © QBC Diagnostics 2012, All Rights Reserved 168 Bradford Drive Port Matilda, PA 16870 U.S.A. +1-814-692-7661 www.qbcdiagnostics.com Form No. 584 Rev. C Specification Sheet Parameters and Operating Ranges

 Open the catalogue to page 1

All QBC Diagnostics catalogues and technical brochures

  1. 842VES

    2 Pages

    En
  2. APEX 12

    4 Pages

    En
  3. APEX 24

    4 Pages

    En
  4. APEX 6

    4 Pages

    En