カタログの抜粋
TomoFix Medial High Tibial Plate (MHT). For Medial High Tibial Technique Guide
カタログの1ページ目を開くImage intensier control Warning This description alone does not provide sufficient background for direct use of the product. Instruction by a surgeon experienced in handling this product is highly recommended. Reprocessing, Care and Maintenance of Synthes Instruments For general guidelines, function control and dismantling of multi-part instruments, please refer to: www.synthes.com/reprocessing
カタログの2ページ目を開くntroduction TomoFix Medial High Tibial Plate (MHT) 2 Surgical Technique General Remarks 5 Positioning and Fixation of the Plate 23 Postoperative Treatment and Implant Removal 43 Product Information Plates 44 Also Available from Synthes: chronOS Wedges 53 TomoFix Medial High Tibial Plate (MHT) Technique Guide Synthes 1
カタログの3ページ目を開くTomoFix Medial High Tibial Plate (MHT). For Medial High Tibial Osteotomies. Features and Benefits Higher stability Four locking screws in the plate head Long shaft portion Evenly transmits the occurring forces into the tibial shaft Compression of the lateral hinge A lag screw pulls the distal osteotomy segment towards the plate… 2 Synthes …and forces the plate into suspension, creating an elastic preload… TomoFix Medial High Tibial Plate (MHT) Technique Guide …which imposes pressure upon the lateral hinge.
カタログの4ページ目を開くEasy plate insertion Compression of the lateral hinge TomoFix technique allows pretension- ng of the plate using cortex screw and TomoFix Knee Osteotomy System TomoFix Tibial Head Plate medial, proximal wedge high tibial - Increased plate strength preload technique stable bridging small stature versions TomoFix Tibial Head Plate lateral, proximal wedge osteotomies stable fixation - Available in right and left TomoFix Femoral Plate medial, distal stable fixation - Available in right and left TomoFix Femoral Plate lateral, distal wedge osteotomies stable fixation - Available in right and left...
カタログの5ページ目を開くIndications and Contraindications Indications: − Open-wedge and closed-wedge osteotomies of the medial proximal tibia − Unicompartmental medial or lateral gonarthrosis with malalignment of the proximal tibia − Idiopathic or posttraumatic varus or valgus deformity of the proximal tibia Contraindication: − Inflammatory arthritis 4 Synthes TomoFix Medial High Tibial Plate (MHT) Technique Guide
カタログの6ページ目を開くGeneral Remarks Open-wedge corrective tibia osteotomies are increasingly being used for joint-preserving surgery of the knee joint. The TomoFix knee osteotomy system is based on the Locking Compression Plate system (LCP) and enables angular-stable connections between the screw and plate. This angular stability allows the stable fixation of an open-wedge osteotomy and hence its rapid osseous healing. Note: Plan the type and position of the osteotomy. The TomoFix medial high tibial plate is suitable for both open and closed-wedge osteotomies. This technique guide will explain the procedure of...
カタログの7ページ目を開くPreparation and Approach 1 Preoperative Planning A precise preoperative plan is crucial to the success of this procedure. The recommended method for planning is that of Miniaci. It must be done on the basis of the weight-bearing x-ray of the full leg in AP view, either on paper or at a digital workstation. − Determine the mechanical axis of the leg: Draw a straight line from the center of the femoral head to the center of the ankle joint (a). − Draw the new weight-bearing line from the center of the femoral head, passing the knee through the desired position (a’). − Determine a hinge point...
カタログの8ページ目を開く- Determine the entry point of the transverse osteotomy. It ies just above the pes anserinus. Make sure there is still enough space for the plate head, so that the screw in hole D can be inserted without protruding into the wedge. De- pending on the determined opening angle and the length of the osteotomy cut (mediolateral diameter of the os- teotomy) the corresponding opening height can be de- rived from Hernigou's trigonometric chart. Trigonometric chart Correction angle Note: These instructions alone do not replace in-depth train- ng in planning for osteotomies. It only serves as a...
カタログの9ページ目を開く312.924 Guiding Block for TomoFix Tibial Head Plate, small, medial, proximal 440.831 TomoFix Tibial Head Plate, small, medial, proximal, shaft 4 holes, head 4 holes, length 112 mm. Pure Titanium 312.926 TomoFix Guiding Block for TomoFix Tibial Head Plate, medial, proximal 440.834 TomoFix Tibial Head Plate, medial, proximal, 4 holes. Pure Titanium 323.042 LCP Drill Sleeve 5.0, for Drill Bits 0 4.3 mm Choose the corresponding guiding block for either the stan- dard sized or small TomoFix plate. Place the guiding block on the plate. The guiding block serves as an aid for attaching the LCP...
カタログの10ページ目を開く3 Positioning of patient Perform the surgery with the patient in a supine position. Attach a lateral support and foot pad to the operating table so that the leg can be easily positioned in 90° flexion and in full extension. Position the patient so that the hip, knee and ankle joint can be visualized with the image intensifier. Lower the contralateral leg at the hip joint to facilitate access to the medial proximal tibia. The sterile draping also exposes the iliac crest so that the leg axis can be checked intraoperatively. A sterile tourniquet can be used, but is not mandatory. Note: Allow...
カタログの11ページ目を開くPreparation and Approach 4 1 Approach Position the knee in 90° flexed position. Mark the anatomic landmarks (medial joint line, cranial border of pes anserinus, course of the medial collateral ligament, and tibial tuberosity) on the skin. Make a 6 – 8 cm long skin incision running from a point anterior to the insertion of the pes in a posterocranial direction. The incision ends over the posteromedial corner of the medial tibial plateau.(1) Note: This incision runs essentially in line with the skin lines and the saphenous nerve. First, divide the subcutaneous tissues and the fascia at the...
カタログの12ページ目を開くOsteotomy 1 1 Determine the position of the osteotomy Instrument 292.260 Kirschner Wire л 2.5 mm with trocar tip, length 280 mm, Stainless Steel >15 mm Alternative instrument 338.002 Guide Wire л 2.5 mm with drill tip, length 300 mm, Stainless Steel >30 mm Position the leg in full extension and adjust the knee joint exactly into AP view under fluoroscopy. Align the medial and lateral compartments in AP projection. Rotate the leg in a position which locates the patella exactly anteriorly (one third of the fibular head is then usually covered by the tibia).(1) Caution: A correct view of the...
カタログの13ページ目を開くDepuy Synthesのすべてのカタログと技術パンフレット
-
Building on Success
16 ページ
-
RADIUS OF CURVATURE
3 ページ
-
HEALIX ADVANCE
4 ページ
-
3.5 mm LCP™ Medial
15 ページ
-
MatrixRIB®FixationSystem
86 ページ
カタログアーカイブ
-
2.4 mm VA LCP™
4 ページ
-
Power line II
4 ページ
-
Concorde
28 ページ
-
LCP Intercarpal
31 ページ
-
LCS® COMPLETE™
2 ページ
-
Synthes TPLO.
8 ページ
-
SynFix-LR System
56 ページ
-
CONDUIT™
15 ページ
-
Brochure_FINAL
2 ページ
-
DePuy Synthes
81 ページ
-
Anspach
3 ページ
-
PINNACLE® Hip Solutions
12 ページ
-
Corail
24 ページ
-
S-ROM® NOILES™
68 ページ
-
Speed
2 ページ
-
attune
80 ページ
-
HAMMERLOCK® 2
2 ページ
-
Polar
4 ページ
-
Alveolar Distractor.
4 ページ
-
Piezoelectric System
4 ページ
-
Air Power Line II
6 ページ
-
TruMatch Pin Guides
16 ページ
-
P F N A
8 ページ
-
SKILL, DEDICATION,
16 ページ
-
Orthopaedics. Overview
20 ページ
-
DURALOC
16 ページ
-
Marathon Cemented Cup
20 ページ
-
REEF Surgical Technique
16 ページ
-
MatrixNEURO
8 ページ
-
Anspach XMax
4 ページ
-
Anspach eMax 2 Plus
4 ページ
-
Small Electric Drive
4 ページ
-
Air Pen Drive
4 ページ
-
Colibri II
4 ページ
-
Spine
25 ページ
-
LCP Distal Fibula Plates
32 ページ
-
Expert Tibia Nail
84 ページ
-
Sacral Bars
16 ページ
-
Pelvic C-Clamp
20 ページ
-
LCP
24 ページ
-
PFNA
112 ページ
-
HCS 1.5, 2.4, 3.0
36 ページ
-
LCP Wrist Fusion
32 ページ
-
LCP Compact Hand
28 ページ
-
VA-LCP Elbow
48 ページ
-
Distal Radius
44 ページ
-
Olecranon
30 ページ
-
LCP Hook Plate
28 ページ
-
DHP & Olecranon
4 ページ
-
LCP S-A
4 ページ
-
Epoca
4 ページ
-
Philos
32 ページ
-
MultiLoc
68 ページ