カタログの抜粋

Data Loggers for Routine Control Washer-Disinfectors / Washer-Disinfectors for Endoscopes / Steam Sterilizers / Bedpan Washers CSSD | Hygiene Control
カタログの1ページ目を開く
Electronic Bowie Dick Test System Together with the evaluation software Winlog.med the new EBI 16 data logger forms an easy to use and very reliable electronic measurement system. With it a comprehensive routine control of steam sterilizers can be performed using an Electronic Bowie Dick Test according to EN ISO 11140-4. In addition to the review of the steam penetration, the relevant sterilization parameters are controlled. clear, reproducible measurement results high-resolution graphical cycle display digital data recording and storage Starter sets EBI 16 With the EBI 16 starter sets you...
カタログの2ページ目を開く
What does the EBI 16 accomplish? • Bowie Dick Test: The EBI 16 delivers clear results during daily checks of the air evacuation test and steam penetration test according to ISO 17665, EN ISO 11140-4 and DIN EN 285. • Easy use, evaluation and documentation of the Bowie Dick Test • Early warning system: The EBI 16 provides early identification of possible failures in steam-sterilizers. Even the smallest quantities of residual air that doesn`t lead to a failed Bowie Dick Test yet are detected. • Vacuum check: The EBI 16 allows a reasonable vacuum check also for sterilizers without pressure...
カタログの3ページ目を開く
Routine Control made easy With the EBI 100 data loggers, ebro offers reliable measurement devices for routine control of steam sterilizers, washer-disinfectors and washer-disinfectors for endoscopes according to German guidelines and ISO 15883. The data loggers are available in different versions including temperature or temperature / pressure loggers with and without external probes. They record the temperature and pressure variation over time during the washer-disinfector or sterilizer process and can easily be read out on a computer with the help of Winlog. med software. This software...
カタログの4ページ目を開く
During the routine monitoring of washer-disinfectors, washer-disinfec- tors for endoscopes and steam sterilizers the pressure and temperature variation in time is documented with the help of the data loggers. During the evaluation, the software recognizes and checks all relevant phases of the sterilization. In addition, all of the required parameters are calculated and graphically displayed with the help of the software. Winlog.med software allows the simultaneous start or readout of mul- tiple data loggers. After the routine monitoring process is concluded, Winlog.med delivers a clear and...
カタログの5ページ目を開く
Bowie Dick Test Logger data logger data logger range Temperature accuracy Pressure measurement range Pressure accuracy Protection class Factory calibration ebro Electronic Wissenschaftlich-Technische Werkstätten GmbH Peringerstr. 10 85055 Ingolstadt, Germany Phone +49 841 95478-0 Fax +49 841 95478-80 Internet: www.ebro.com Email: ebro@xyleminc.com All names are registered tradenames or trademarks of WTW GmbH. Technical changes reserved. © 2013 WTW GmbH.October 2013
カタログの6ページ目を開くEbro Electronicのすべてのカタログと技術パンフレット
-
TMX-SERIES
176 ページ
-
TIC 100
4 ページ
-
Evaluation Software
6 ページ
-
Passion for good oil
4 ページ
-
EBI 300 / 310 / 330
12 ページ
-
Ebro Catalog
180 ページ
-
Catalog_2014
156 ページ
-
Set SL 1500
6 ページ
-
EBI 300
4 ページ
-
EBI 300 / 310
6 ページ
カタログアーカイブ
-
Data Logger Systems_2015
20 ページ
-
Data Logger Systems_2013
12 ページ