カタログの抜粋
Syringe Pump The JMS SP-500 Syringe Pump provides accurate and continous delivery of medication with a disposable syringe. Features ♦ Occlusion pressure is monitored by means of a load cell sensor built into the slider. ♦ The pressure sensor ensures the correct delivery, preventing potential risk of an unintended infusion caused by a negative venous pressure. ♦ The high quality stepper motor enables a non-pulsatile flow, even at an extremely low flow rate. ♦ The pump features two CPUs that monitor all functions in order to ensure safe, breakdown-free operation. ♦ The syringe can be loaded in a single action via the exclusive gate opening mechanism. Applications 1) Infusion of cancer-fighting drugs, hormones and anesthetic agents. 2) Infusion of nutrition, vasodilators and vasopressors in ICU, CCU and NICU. 3) Administration of milk and enteral nutrition to prematurely born babies and ne
カタログの1ページ目を開くSP-500 Syringe Pump Technical Specifications Name Model Syringes Flow Rate Range Rapid Flow Rate Accuracy Total Volume Display 0.1~ 200 ml/h 10 ml-syringe 0.1~ 300 ml/h 20, 30 and 50 ml-syringes (0.1 ml/h increments) Approx. 200 ml/h (10 ml syringe) Approx. 350 ml/h (20 ml syringe) Approx. 450 ml/h (30 ml syringe) Approx. 800 ml/h (50 ml syringe) Drive Accuracy: ±1%, with JMS syringes: ±3% (After 1 hr with flow rate of 1 ml/h or higher) 0.1-999.9 ml (resettable) Three levels can be selected: Adjustment of Occlusion Level 2 [L][low pressure] 0.2 ~ 0.6 Kgf/cm (150 ~ 440 mmHg, 20 ~ 60 kPa)...
カタログの2ページ目を開くJMS North America Corporationのすべてのカタログと技術パンフレット
-
OT-701
2 ページ
-
Blood Management System
24 ページ
-
NEOSHIELD®
2 ページ
-
BasicLine
4 ページ
-
SDS-50
2 ページ
-
SysLoc® MINI
2 ページ