EDUCATIONAL GUIDE How to Measure Uncertainty QUALITY CONTROL
カタログの1ページ目を開く
How to Measure Uncertainty “The laboratory shall determine measurement uncertainty for each measurement procedure, in the examination phases used to report measured quantity values on patients’ samples. The laboratory shall define the performance requirements for the measurement uncertainty of each measurement procedure and regularly review estimates of measurement uncertainty.” ISO 15189:2012 - Section 5.5.8.3 What is Measurement Uncertainty? Measurement Good Practice Guide No. 11 It relates to the doubt that exists for the result of any measurement. For every measurement there is always a...
カタログの2ページ目を開く
AACB recommend labs use at least 6 months QC data to calculate uncertainty. When measuring uncertainty we look at two things intra assay precision and inter assay precision. Intra precision refers to precision within a run; it is normally measured by running 20 or more replicates of the same sample at the same time.This process will help to identify any random uncertainties within a test system. Inter precision on the other hand refers to precision over a number of different runs, it is normally measured by running 20 or more replicates of the sample over several days e.g. one replicate...
カタログの3ページ目を開く
Other Factors Affecting Uncertainty It is important to consider bias when calculating uncertainty. Bias must be measured and if it is significant removed or minimised. If bias is not removed the uncertainty of the bias correction must be calculated and included in the overall uncertainty measurement. To calculate the uncertainty of bias we must first determine uRef and uRep. uRef – Uncertainty value of the analyte assigned to the reference material/EQA. This can be obtained from the reference material or EQA report. uRep – Uncertainty value of the analyte in the reference material/EQA when...
カタログの4ページ目を開く
Conclusion The uncertainty of a measurement relates to the doubt that exists for the result of any measurement. If uncertainty is measured it is no longer uncertainty, but the confidence interval within which the result falls. Uncertainty should be assessed regularly and attempts made to improve the value. Glossary Error – The difference between the measured value and the true value. Uncertainty – A quantification of the doubt relating to a measured value. Standard Deviation (SD) – A measure of how spread out a set of numbers are. Approximately two thirds of all readings will fall between ±...
カタログの5ページ目を開く
ACUSERA True third party quality controls As a world leading manufacturer of multi-analyte true third party controls, thousands of laboratories rely on Randox to accurately assess test system performance and ultimately empower them with the confidence required to release patient test results. With more than 390 analytes available, the number of individual controls required to cover your test menu is significantly reduced while simultaneously reducing costs, time and storage space. A choice of formats is available, including liquid or lyophilised, which ensures flexibility and suitability...
カタログの6ページ目を開く
ACUSERA 24•7 Interlaboratory Data Management Designed for use with the Acusera range of third party controls, the Acusera 24•7 software helps laboratories monitor and interpret their QC data. Access to an impressive range of features, including interactive charts, the automatic calculation of Measurement Uncertainty & Sigma Metrics and live peer group data generated from our extensive database of laboratory participants, ensures Acusera 24•7 is the most comprehensive package available. • Advanced statistical analysis with automatic calculation of performance metrics including; Sigma, UM, TE...
カタログの7ページ目を開く
Copyright © 2019 Randox Laboratories Ltd. All rights Reserved. VAT number: GB 151682708. Product availability may vary from country to country. Some products may be for Research Use Only. For more information on product application and availability, please contact your local Randox Representative.
カタログの8ページ目を開くRandox Laboratories/ランドックスのすべてのカタログと技術パンフレット
-
LT735 Vivalytic Overview
32 ページ
-
Acusera New Controls
8 ページ
-
Acusera 24.7
24 ページ
-
Acusera Third Party Controls
108 ページ
-
Acusera SMART Controls
12 ページ
-
Company Overview
40 ページ
-
Reagents Brochure
64 ページ
-
sdLDL-C
8 ページ
-
RX modena
20 ページ
-
RX misano
20 ページ
-
RX Daytona Plus
20 ページ
-
Rx Imola
20 ページ
-
RX Monaco
20 ページ
-
2024 Product List
64 ページ
-
VeraSTAT-V
16 ページ
-
LT033 RIQAS Explained
64 ページ
-
VERASTAT
16 ページ
-
Endocrine Array
4 ページ
-
Cerebral Arrays
4 ページ
-
Thyroid Arrays
4 ページ
-
Basic QC Stastics
8 ページ
-
Commutability Guide
4 ページ
-
Qnostics
52 ページ
-
Adiponectin LT519
28 ページ
-
Specific Proteins
40 ページ
-
Linearity sets
12 ページ
-
Antioxidants
16 ページ
-
Diabetes Portfolio
28 ページ
-
Total Bile Acids
4 ページ
-
VIVALYTIC
30 ページ
-
RANDOX DISCOVERY
36 ページ
-
HbA1c
2 ページ
-
Preparing QC
1 ページ
-
Point of Care Testing
12 ページ
-
Guide to running QC
1 ページ
-
Tumour Marker Arrays
4 ページ
-
LT169 Cardiac Array
4 ページ
カタログアーカイブ
-
ACUSERA
108 ページ
-
Evidence Evolution
28 ページ
-
Evidence
16 ページ
-
STI Multiplex Array
8 ページ
-
Molecular Testing
16 ページ
-
Evidence Investigator
20 ページ
-
Fertility Array
4 ページ
-
Rx Daytona
16 ページ
-
Evidence Investigator
16 ページ
-
Evidence
16 ページ