EDUCATIONAL GUIDE QC Multi-Rules QUALITY CONTROL

カタログの1ページ目を開く
QC Multirules Educational Guide - 2

QC Multi-Rules What are QC Multi-Rules? Based on statistical process control, QC Multi-rules use a combination of control rules to decide whether an analytical run is in-control or out-of-control. Control Data What is Measurement Uncertainty? Recognising out-of-control QC results using a single rule A laboratory favourite is the 1 result outside 2SD rule. This is a great rule for alerting lab staff to possible out-of-control events, however, you must be careful as it does have a high false rejection rate. Remember, it is normal for 1/20 results to be outside 2SD! It is important that you...

カタログの2ページ目を開く
QC Multirules Educational Guide - 3

Types of Analytical Errors QC Multi-Rules can be used to detect systematic or random errors. Recognising if you have either a systematic or random error can help you identify the source of the problem. Random Errors The characteristics of Random Errors are as follows: • Error which varies in an unpredictable manner (in magnitude and sign) whenever a large number of measurements of the same quantity are made under effectively identical conditions. • Difficult to eliminate but repetition can reduce the influence of random errors.between ± one SD of the average and around 95% of all readings...

カタログの3ページ目を開く
QC Multirules Educational Guide - 4

Explained QC Multi-Rules There are 6 basic QC Multi-Rules. Some of these detect random error and some detect systematic error (which may indicate a bias in the system). Rule 12S (1 result exceeds 2SD) This rule is designed as a warning rule only. If a single QC result is more than ±2 standard deviations from the mean, this run along with previous analytical runs should be evaluated before accepting the run and reporting results. The 12S rule is designed to warn that random error or systematic error may be present. Using the 12S rule alone may lead to frequent rejection of valid runs (false...

カタログの4ページ目を開く
QC Multirules Educational Guide - 5

Rule 2 of 32S (2 of 3 results exceed 2SD) This is a variation of the 22S rule and is used to detect systematic error. The rule is violated when any two levels of control in a run exceed 2 standard deviations on the same side of the mean. Rule R4S (Range exceeds 4SD) This rule is violated if there is at least 4 standard deviations difference between control values within a single run. In this example the level 1 control is +2S above the mean and the level 2 control is -2S below the mean. The total difference is 4s. This rule identifies random error only. Rule 41S (4 results exceed 1SD) This...

カタログの5ページ目を開く
QC Multirules Educational Guide - 6

Rule 10x (10 results same side of the mean) This rule is violated if ten consecutive control results regardless of level fall on the same side of the mean. The rule detects systematic error and can be applied both within and across runs. Violation of this rule may not require rejection of the run but rather indicate that instrument maintenance or calibration is needed. The 10x rule can be modified to 7,8,9 or 12 consecutive results on the same side of the mean, each giving varying levels of sensitivity to systematic bias. This rule is violated when a group of seven consecutive results for a...

カタログの6ページ目を開く
QC Multirules Educational Guide - 7

Tools to Assist Labs Apply QC Multi-Rules To apply QC Multi-Rules you firstly collect your control measurements in the same way that you would for a Levey-Jennings chart. From this you establish a mean and standard deviation (SD). All you are changing is the control limits and how the data will be interpreted. Online QC software with real-time peer group statistics Acusera 24•7 will apply QC multi-rules automatically Compatible for use with the Acusera range of third party controls, the Acusera 24•7 software is designed to help laboratories monitor and interpret their QC data. Access to an...

カタログの7ページ目を開く
QC Multirules Educational Guide - 8

Copyright © 2019 Randox Laboratories Ltd. All rights Reserved. VAT number: GB 151682708. Product availability may vary from country to country. Some products may be for Research Use Only. For more information on product application and availability, please contact your local Randox Representative.

カタログの8ページ目を開く

Randox Laboratories/ランドックスのすべてのカタログと技術パンフレット

  1. Acusera 24.7

    24 ページ

  2. Company Overview

    40 ページ

  3. Reagents Brochure

    64 ページ

  4. sdLDL-C

    8 ページ

  5. RX modena

    20 ページ

  6. RX misano

    20 ページ

  7. RX Daytona Plus

    20 ページ

  8. Rx Imola

    20 ページ

  9. RX Monaco

    20 ページ

  10. 2024 Product List

    64 ページ

  11. VeraSTAT-V

    16 ページ

  12. VERASTAT

    16 ページ

  13. Endocrine Array

    4 ページ

  14. Cerebral Arrays

    4 ページ

  15. Thyroid Arrays

    4 ページ

  16. Basic QC Stastics

    8 ページ

  17. Qnostics

    52 ページ

  18. Adiponectin LT519

    28 ページ

  19. Specific Proteins

    40 ページ

  20. Linearity sets

    12 ページ

  21. Antioxidants

    16 ページ

  22. Diabetes Portfolio

    28 ページ

  23. Total Bile Acids

    4 ページ

  24. VIVALYTIC

    30 ページ

  25. RANDOX DISCOVERY

    36 ページ

  26. HbA1c

    2 ページ

  27. Preparing QC

    1 ページ

カタログアーカイブ

  1. ACUSERA

    108 ページ

  2. Evidence Evolution

    28 ページ

  3. Evidence

    16 ページ

  4. Molecular Testing

    16 ページ

  5. Fertility Array

    4 ページ

  6. Rx Daytona

    16 ページ

  7. Evidence

    16 ページ