EDUCATIONAL GUIDE Troubleshooting QC Errors QUALITY CONTROL
カタログの1ページ目を開く
Troubleshooting QC Errors So you ran QC this morning and realised that one of your analytes has been flagged as “out-of-control”, what do you do next? Do you ignore the warning and continue patient testing, repeat the control until it’s within range or do you halt patient testing and investigate the source of the error? Your QC Has Failed, What Do You Do Next? When it comes to troubleshooting QC errors, unfortunately there is no easy path to take. However, it’s important that you have standard operating procedures in place, outlining what to do in the event of an out-of-control error....
カタログの2ページ目を開く
2. Review Your Levey-Jennings Charts To Understand The Type Of Error That Has Occurred There are two types of analytical errors: Systematic Errors Random Errors Both of these types of errors can be recognised on a standard Levey-Jennings chart and by using QC multi-rules. Identifying the type of error will help you relate the error to a possible cause. Systematic Error Systematic errors create a characteristic bias and can be identified by a change in the mean of control values. The change in the mean may be either a gradual trend or an abrupt shift in control results. Standard Deviation...
カタログの3ページ目を開く
Using QC Multi-Rules To Identify Systematic & Random Error You can use QC multi-rules to differentiate between systematic and random error. +3s Standard Deviation Standard Deviation This is a warning rule that is broken when a single control observation is outside the +/-2s limits. This rule warns that either a random or systematic error is present in the test system. If no other unacceptable results are apparent in the test system, it must be assumed that this is simply a random error and no further troubleshooting action is necessary. Any QC result outside +/-3s breaks this rule. This...
カタログの4ページ目を開く
3. Relate The Type Of Error To Possible Causes Ask yourself questions in order of likely relevance and review the most common solutions to the problem. Dependent on whether you have identified a systematic or random error in your system, this can help you determine the possible root cause of the error. See below for some common causes of systematic and random errors. Systematic Error: Shift 1. Change or failure in light source 2. Major instrument maintenance 3. Failure in sampling system 4. Change in temperature 5. Failure in reagent dispense system 6. Change in reagent formulation 7. New...
カタログの5ページ目を開く
ACUSERA True third party quality controls As a world leading manufacturer of multi-analyte true third party controls, thousands of laboratories rely on Randox to accurately assess test system performance and ultimately empower them with the confidence required to release patient test results. With more than 390 analytes available, the number of individual controls required to cover your test menu is significantly reduced while simultaneously reducing costs, time and storage space. A choice of formats is available, including liquid or lyophilised, which ensures flexibility and suitability...
カタログの6ページ目を開く
ACUSERA 24•7 Interlaboratory Data Management Designed for use with the Acusera range of third party controls, the Acusera 24•7 software helps laboratories monitor and interpret their QC data. Access to an impressive range of features, including interactive charts, the automatic calculation of Measurement Uncertainty & Sigma Metrics and live peer group data generated from our extensive database of laboratory participants, ensures Acusera 24•7 is the most comprehensive package available. • Advanced statistical analysis with automatic calculation of performance metrics including; Sigma, UM, TE...
カタログの7ページ目を開く
Copyright © 2019 Randox Laboratories Ltd. All rights Reserved. VAT number: GB 151682708. Product availability may vary from country to country. Some products may be for Research Use Only. For more information on product application and availability, please contact your local Randox Representative.
カタログの8ページ目を開くRandox Laboratories/ランドックスのすべてのカタログと技術パンフレット
-
LT735 Vivalytic Overview
32 ページ
-
Acusera New Controls
8 ページ
-
Acusera 24.7
24 ページ
-
Acusera Third Party Controls
108 ページ
-
Acusera SMART Controls
12 ページ
-
Company Overview
40 ページ
-
Reagents Brochure
64 ページ
-
sdLDL-C
8 ページ
-
RX modena
20 ページ
-
RX misano
20 ページ
-
RX Daytona Plus
20 ページ
-
Rx Imola
20 ページ
-
RX Monaco
20 ページ
-
2024 Product List
64 ページ
-
VeraSTAT-V
16 ページ
-
LT033 RIQAS Explained
64 ページ
-
VERASTAT
16 ページ
-
Endocrine Array
4 ページ
-
Cerebral Arrays
4 ページ
-
Thyroid Arrays
4 ページ
-
Basic QC Stastics
8 ページ
-
Commutability Guide
4 ページ
-
Qnostics
52 ページ
-
Adiponectin LT519
28 ページ
-
Specific Proteins
40 ページ
-
Linearity sets
12 ページ
-
Antioxidants
16 ページ
-
Diabetes Portfolio
28 ページ
-
Total Bile Acids
4 ページ
-
VIVALYTIC
30 ページ
-
RANDOX DISCOVERY
36 ページ
-
HbA1c
2 ページ
-
Preparing QC
1 ページ
-
Point of Care Testing
12 ページ
-
Guide to running QC
1 ページ
-
Tumour Marker Arrays
4 ページ
-
LT169 Cardiac Array
4 ページ
カタログアーカイブ
-
ACUSERA
108 ページ
-
Evidence Evolution
28 ページ
-
Evidence
16 ページ
-
STI Multiplex Array
8 ページ
-
Molecular Testing
16 ページ
-
Evidence Investigator
20 ページ
-
Fertility Array
4 ページ
-
Rx Daytona
16 ページ
-
Evidence Investigator
16 ページ
-
Evidence
16 ページ