Daily QA of the CyberKnife using the QA BeamChecker+ Gibbs Cancer Center & Research Institute
カタログの1ページ目を開く
Speaker is the owner of Spectrum Medical Physics, LLC, a company which has contracts with; - Standard Imaging, Inc - Siemens Healthcare USA c Gibbs Cancer Center SPECTRUM & Research Institute Medical Physics, LLC SPECTRUM Medical Physics, LLC
カタログの2ページ目を開く
Current Clinical Use Varian TrueBeam • Daily — Photon Output (Gantry Mount) — Photon Energy Constancy (Gantry Mount) — Photon Flatness and Symmetry (Gantry Mount) — MV Imaging (Gantry Mount) — kV Imaging (Gantry Mount) • Weekly — Electron Output (Couch-Placed) — Electron Energy Constancy (Couch-Placed) — Electron (Couch-Placed) — EDW (Gantry Mount) — RapidArc (Gantry Mount) • Monthly — Respiratory Gating (Gantry Mount) • Patient-Specific — Electron Cut-out Measurements (Couch-Placed) Gibbs Cancer Center SPECTRUM & Research Institute Medical Physics, LLC
カタログの3ページ目を開く
Gibbs Cancer Center & Research Institute Lovable Aspects of the QABC+ • Robust • Simple Hardware Interface • Simple Software Interface • Simple-to-traverse database • Easily Scalable as clinic changes • Stable readings • Easy out-of-tolerance notifications • Very few problems • Excellent Support
カタログの4ページ目を開く
• Machine Isocenter matches the Room Isocenter • Room Lasers provide demarcation of machine isocenter • Alignment crosshair • Optical SSD • Can be delivered as an automated clinical beam Gibbs Cancer Center SPECTRUM & Research Institute Medical Physics, LLC
カタログの5ページ目を開く
Output Verification (Current Methods) Requires manual manipulation of the robot Beam delivered in a non-clinical mode Requires manual correction for temperature and pressure Requires manual numeric entry for baseline comparison Setup uncertainty associated with setup following Iris QA Gibbs Cancer Center & Research Institute
カタログの6ページ目を開く
Output Verification (Current Methods) Machine Isocenter matches the Room Isocenter Room Lasers provide demarcation of machine isocenter Alignment crosshair (Sort of) Optical SSD Can be delivered as an automated clinical beam
カタログの7ページ目を開く
• Crosshair used for "rough alignment" • Four fiducial markers for alignment and offset calculation using TLS Gibbs Cancer Center & Research Institute
カタログの8ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの9ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの10ページ目を開く
• Place 5 SINLGE beams (4 peripheral and 1 center) • 500 MU Each Gibbs Cancer Center & Research Institute
カタログの11ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの12ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの13ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの14ページ目を開く
Readiness Technique ■u XRS A kv I mA | ms I Estimated Imaging Dose (cGy) Exposed 1.0 Expected 1,3 Gibbs Cancer Center & Research Institute c SPECTRUM Medical Physics, LLC
カタログの15ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの16ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの17ページ目を開く
Gibbs Cancer Center & Research Institute
カタログの18ページ目を開く
Thank You Gibbs Cancer Center & Research Institute
カタログの19ページ目を開くStandard Imagingのすべてのカタログと技術パンフレット
-
QA BeamChecker Plus
7 ページ
-
QA CrossChecker
2 ページ
-
QA StereoChecker
2 ページ
-
SuperMAX
8 ページ
-
Luthmann QA Tool
2 ページ
-
Blue Water Phantom
2 ページ
-
ARC QA Phantom
2 ページ
-
IMRT
2 ページ
-
MIMI
2 ページ
-
Lucy 3D QA
12 ページ
-
DoseView 1D
4 ページ
-
DoseView 3D
12 ページ
-
RAy Datasheet
2 ページ
-
RT Workspace Brochure
10 ページ
-
IMSure Brochure
10 ページ