PROCEDURE FOR THE PROPER INSERTION AND DISINSERTION OF THE INTERCHANGEABLE TIPS
1 / 3Pages

Catalog excerpts

PROCEDURE  FOR THE PROPER INSERTION AND DISINSERTION OF THE INTERCHANGEABLE TIPS - 1

PROCEDURE FOR THE PROPER INSERTION AND DISINSERTION OF THE INTERCHANGEABLE TIPS

Open the catalog to page 1
PROCEDURE  FOR THE PROPER INSERTION AND DISINSERTION OF THE INTERCHANGEABLE TIPS - 2

It is recommended to follow the ensuing instructions carefully in order to avoid irreversible damage to the handpiece and/or tip. Unscrew the upper part the laser Orange handpiece counetrclockwise Take the laser tip from its box (200μm/300μm/400μm) #weare8853 8853 S.p.A. Sales Office - Administrative offices - Works Via Pitagora, 11 - 20016 Pero (MI) Tel.+39 02 8853.501 :: laser@8853.it :: www.88dent.com

Open the catalog to page 2
PROCEDURE  FOR THE PROPER INSERTION AND DISINSERTION OF THE INTERCHANGEABLE TIPS - 3

Remove the protective cap from the tip. Insert the tip inside the handpiece all the way in avoiding the in terposition of whatever foreign material between the connection of the tip and handpiece Reverse the operation to replace and/or change the tip. After use tip must be cleaned with disinfecting solution (with H2O2 solution soacket gauzes/Clorhexidine solution). Any residual tissue on tip must be removed. Tips can be sterilized by steam sterilization after appropriate packaging (rubber program or 134°c x2 atm x 5-5 min). PRODOTTO ITALIANO #weare8853 8853 S.p.A. Sales Office -...

Open the catalog to page 3

All 88Dent catalogs and technical brochures

  1. Software

    14 Pages

  2. Scanner

    7 Pages

  3. Z1 Series

    12 Pages