1. Catalogs
  2. ACEA Biosciences Inc.
  3. Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source
video corpo

Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source

Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source

Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source

Product catalog summary
Overview: The Agilent 6470B Triple Quadrupole LC/MS is designed for high reliability, sensitivity, and robustness in trace-level analysis. It features advanced components such as the octopole ion guide, hyperbolic quadrupoles, and a curved hexapole collision cell, all aimed at enhancing performance and reducing downtime.
Specifications:
  • MRM Sensitivity: IDL for ESI positive and negative modes is less than 4.0 fg with 99% confidence.
  • Signal-to-Noise Ratio: Greater than 750,000:1 for both ESI positive and negative modes.
  • Mass Resolution: Automatically adjustable during autotune (0.7 Da, 1.2 Da, 2.5 Da) and manually adjustable down to 0.5 Da.
  • Mass Range: m/z 5 to 3,000.
  • Mass Accuracy: 0.1 Da from m/z 5 to 1,000, 0.2 Da from m/z 1,000 to 2,000, and 0.3 Da from m/z 2,000 to 3,000.
  • Dynamic Range: Greater than 6.0 × 106.
  • Polarity Switching: Less than 25 ms.
  • Acquisition Modes: Includes MS1 scan, MS2 scan, product ion scan, and more.
  • Maximum Scan Rate: 17,000 Da/sec.
  • MRM Transitions: Up to 13,500 ion transitions per method.
System Features:
  • Ionization Sources: Includes Agilent Jet Stream, Electrospray, Nanospray, APCI, and more.
  • Autotune: Automated optimization for ion optics and mass axis calibration.
  • Vacuum System: Two turbomolecular pumps and one mechanical pump.
  • VacShield: Allows maintenance without breaking vacuum.
Ordering Information: The system includes the mass spectrometer, Agilent MassHunter Workstation software, and installation services.
Disclaimer: Performance specifications are reviewed for accuracy but do not represent installation tests. S/N specifications are specific to certain conditions and cannot be extrapolated.
See more

Catalog excerpts

Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source-1

Data Sheet Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source The Agilent 6470B triple Quadrupole LC/MS delivers superior reliability, sensitivity, and robustness for targeted trace-level analysis. Designed with robustness, versatility, and throughput in mind, it contains Agilent's signature octopole ion guide, hyperbolic quadrupoles, curved and tapered hexapole collision cell, and advanced detector geometry. Built with VacShield to reduce instrument downtime providing faster maintenance without breaking vacuum. The end result is a rugged and versatile instrument suitable for high-throughput targeted screening or trace-level quantitative research applications. Parameter 10 fg of reserpine injected on column, quantifying on the transition m/z 609.3 to 195.1 IDL <4.0 fg, with 99% confidence (verified onsite during installation) 10 fg of chloramphenicol injected on column, quantifying on the transition m/z 321.0 to 152.0 1 pg of reserpine injected on column, quantifying on the transition m/z 609.3 to 195.1 1 pg of chloramphenicol injected on column, quantifying on the transition m/z 321.0 to 152.0 Mass resolution (autotune) Automatically adjusted during Autotune procedure 0.7 Da, 1.2 Da, and 2.5 Da (Unit, Wide, Widest) Mass resolution (manual tune) Manually adjusted by the user during Manual tune procedure Mass range Mass accuracy 0.1 Da from m/z 5 to 1.000 0.2 Da from m/z 1,000 to 2,000 0.3 Da from m/z 2,000 to 3,000 Mass stability <0.1 Da in 24 hours up to m/z 2,122 in Positive mode and m/z 2,234 in Negative mode Dynamic range >6.0 × 106, resulting in up to 6 orders of linear dynamic range from the LOD Polarity switching (electronics) Acquisition modes MS1 scan, MS2 scan, product ion scan, neutral loss scan, neutral gain scan, precursor ion scan, SIM, and MRM (static, dynamic, triggered) Maximum scan rate Maximum MRM acquisition rate Minimum MRM dwell time 450 per time segment, up to 13,500 ion transitions per method 4,000 dynamic MRM transitions per method Up to 10 MRM transitions (primary and secondary) for library search and compound confirmation Collision cell ion clearance

 Open the catalog to page 1
Agilent 6470B Triple Quadrupole LC/MS with Agilent Jet Stream Source-2

General system specifications Parameter Single point of control Specification Single-point data system method capability with full control of Agilent Infinity I and Infinity II and Agilent 6470B triple quadrupole LC/MS systems Polarity change in time segment Time programming Dedicated scan modes (Full Scan, Product Ion Scan, Precursor Ion Scan, Neutral Loss Scan, and Neutral Gain Scan) or targeted ion monitoring (SIM, MRM, dMRM, and tMRM) Dynamic and triggered MRM aligns MRMs with compound retention time Solvent divert through calibrant delivery system valve within the analysis Agilent Jet Stream...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.