USER MANUAL 16-2090 FUSION CAMERA
Open the catalog to page 1
Description of the front and back panels_________________________________ 16 Suggested decontamination procedures_________________________________ 17
Open the catalog to page 2
1 Preface Thank you for the confidence you have demonstrated by purchasing the Ackermann Fusion Camera System. In order to make the best use of this Fusion device, it is essential for you to become acquainted with this manual. Symbol corresponds to points requiring special attention. Symbol provides advice. To facilitate installation and use of the device, we have attempted to make the device manual more practical. Consequently, references to the product presentation on the last page (like C1 for example) will be provided for easier viewing of the relevant product parts. This user manual is...
Open the catalog to page 3
2 About the device Ackermann Fusion Camera is a Cmos FULL HD medical camera. Its ergonomically designed sensor, automatic shutter, good sensitivity, excellent resolution and faithful color-rendering makes it ideal as a medical tool for rigid or flexible endoscopy. This camera comes with: • A camera head with integrated 15-35 mm HD optional zoom lens, Video cables, Optional accessories: • Ackermann Premium USB stick for archiving. For the United States and Canada, use the “hospital grade” power cord. This must be connected to a “hospital grade” mains socket. This equipment has been delivered...
Open the catalog to page 4
The device must be opened only by a competent technician authorized by the manufacturer. Do not insert metal objects into the device. This is to avoid any risk of electrical shock, fire, shortcircuit or hazardous emissions. Do not expose the device to splashed water or to humidity. Use only the accessories supplied with the device or as recommended by the manufacturer. This camera is not a flame-proof device. It must not be used in the presence of flammable anaesthetics. This device was not designed to be used in an ionising environment. This device is not sterile. The surface temperature...
Open the catalog to page 5
4 Regulatory advice 4.1 Compliance This product is designed and manufactured by Ackermann proven to have a certified quality system. It meets the requirements of European directive 93/42/CEE, on medical devices. Consequently, it particularly meets the standards of electrical safety (IEC) and electromagnetic compatibility (EMC) ad hoc. 4.2 Electromagnetic interferences and electrostatic discharge Although this product complies with EMC standards, it may in very special circumstances interfere with other devices, or itself be the object of interference from other devicees or an unfavorable...
Open the catalog to page 6
The symbol displayed on the device and on the accompanying documentation indicates that this product cannot under any circumstances be treated as household waste. It must therefore be delivered to a waste collection center for the recycling of electrical and electronic equipment. In disposing of, please comply with the waste elimination norms in effect in the country where it is installed. For further details on the treatment, recovery and recycling of this device, kindly contact your nearest distributor who will advise you on the procedure to follow. 5 Installing the device This medical...
Open the catalog to page 7
The DVI-D output produces a progressive 1080 signal. • If needed, connect other monitors or peripherals, preferably using the HDSDI & 3GSDI output ports C8 . HDSDI, 3GSDI output ports produce an interleaved 1080 signal. HDSDI, GSDI output ports are preferred when a very long cable is necessary. • Turn the activation switch S2 to position “1”. The camera takes about 30 seconds to boot. The Ackermann logo is displayed on the screen during this interval. The flashing LED Standby button, and the appearance of “Standby” on the screen indicate that the camera is properly activated. • Connect the...
Open the catalog to page 8
You can at any time during surgery to return to the factory settings for the surgical specialty by clicking the “Specialities presets” button. Press the “STOP” button at the end of the surgery. You can at this stage: Return to the main menu by tapping the “Completed” button; OR Create a user profile from the settings used during the operation; OR Log in if you wish to add this set of settings to your profile. Before connecting an endoscope (or fibroscopes) to the camera lens, check the condition of the endoscope (clean distal lens, setup if necessary). 5.3 Activating the device from a...
Open the catalog to page 9
6 Operating guidelines 6.1 Selecting the surgical specialty This camera can be used in various mini-invasive surgical operations. Select the specialty from the pre-entered list. This loads the video parameters for the application accordingly: LAPAROSCOPY HYSTEROSCOPY CYSTOSCOPY ENT ARTHROSCOPY FLEXIBLE ENDOSCOPY Note 1: Each of these specialties can be accessed by scanning the Fusion Card on its dedicated area Z1 on the control unit. This can be done at any time. If the Fusion Card is scanned while a surgical operation is in progress, then confirmation will be requested. Note 2: If you have...
Open the catalog to page 10
Type of press Short press Long press Light source - Intensity + Light source - Intensity + Light source - Intensity - Light source - Intensity - Freeze frame Freeze frame Recording images on the camera (if equipped with an integrated recorder) Record / Stop video recording on the camera (if equipped with an integrated recorder) Insufflator (RUN / High Flow Rate / STOP) Light source (ON / OFF) Recording images on the camera (if equipped with an integrated recorder) (b) Record / Stop video recording on the PC Record / Stop video recording on the camera (if equipped with an integrated...
Open the catalog to page 11
You should proceed as following: • Pair the camera to the endoscope, activate the ligt source, reduce ambient light and film a white surface • Set the white balance; the “BALANCING IN PROGRESS” message is displayed on the screen. • Continue to film the white surface until the “BALANCE OK” message is displayed on the screen. Note 1: Setting the white balance causes the chronometer to be activated, signaling the start of the surgery and therefore the availability of additional functionalities, particularly those linked to the image quality settings. Note 2: If the light source is connected to...