Introduction
The Aerogen® Pro System is a portable medical device for aerosolizing medications for inhalation, suitable for patients on and off mechanical ventilation, including neonates, pediatrics, and adults. It operates without compressed gas and is intended for hospital use.
Indications for Use
The device is designed for multiple patient use to aerosolize medications for inhalation across all age groups.
System Components
The system includes a nebulizer, T-piece, controller, controller cable, AC/DC adapter, and mounting brackets. It can hold up to 10 mL of medication and uses the Aerogen Vibronic® aerosol generator.
System Warnings
Only trained medical personnel should operate the device. Ensure proper functioning by checking for visible aerosol and no flashing indicator lights. Avoid using with flammable substances and unauthorized modifications. Follow cleaning and sterilization instructions strictly.
Assembly & Installation
Clean and sterilize components before use. Connect the nebulizer to the T-piece and controller. The system can operate on AC power or its internal battery for up to 45 minutes.
Installation for Use with a Ventilator
Connect the nebulizer to the inspiratory limb of the breathing circuit and perform a leak test. Secure the controller with the universal mounting bracket.
Nebulisation
For doses ≤ 3 mL, initiate a 15-minute cycle. For doses > 3 mL, initiate a 30-minute cycle.
Installation for Use Off-Ventilator
For face mask use, connect the mask and elbows to the nebulizer. For mouthpiece use, connect the nebulizer to the T-piece and then to the mouthpiece.
Functional Test
Perform a functional test before first use and between patients by inspecting parts, adding saline, and verifying operation with the controller.
System Operation and Verification
Ensure the system is operational by checking indicator lights and aerosol visibility. Verify battery operation and discard any remaining liquid before patient use.
Cleaning, Disinfection, and Sterilisation
Components must be cleaned, disinfected, and sterilized before first use. Manual cleaning involves disassembling parts, washing with detergent, and rinsing. Disinfection uses approved agents, and sterilization can be done via steam or hydrogen peroxide gas plasma.
Controller and Mounting Bracket Cleaning
Use disinfectant wipes for cleaning. Do not autoclave or immerse in liquid.
Troubleshooting
Address common issues like low battery or connection errors by recharging, cleaning, or checking connections. Contact Aerogen for persistent issues.
Warranty and Product Life
The nebulizer has a one-year warranty, and the controller and adapter have a two-year warranty. The controller is validated for 1460 doses, and the nebulizer for 730 doses and 26 autoclave treatments.
Specifications
Includes physical dimensions, environmental conditions, power requirements, and performance metrics like flow rate and particle size.
Symbols and Electromagnetic Susceptibility
Explains symbols used and compliance with electromagnetic compatibility standards.
Warnings and Precautions
Use only specified components to avoid emissions or immunity issues. Follow EMC precautions and avoid using near RF communication devices.
Electromagnetic Compatibility (EMC) Information
The system is designed for specific electromagnetic environments with low RF emissions, suitable for all establishments.
Separation Distances
Recommended distances between RF equipment and the nebulizer vary by frequency range.
Electromagnetic Immunity
The system is tested for immunity against various electrical disturbances, ensuring operation in typical environments.
Manufacturer Information
Aerogen Ltd., Galway Business Park, Dangan, Galway, Ireland. Contact: +353 91 540 400, [email protected], www.aerogen.com
See more

Catalog excerpts

Aerogen Pro-1

System Instruction Manual

 Open the catalog to page 1
Aerogen Pro-3

Contents Introduction 2 Indications for Use System Warnings Assembly & Installation Installation For Use With A Ventilator Nebulisation 16 Installation For Use Off-Ventilator Functional Test Cleaning, Disinfection and Sterilisation Troubleshooting 26 Warranty 28 Life Of Product Specifications 29 Performance 30 Symbols 31 Appendix 1 Aerogen® Pro System Instruction Manual

 Open the catalog to page 3
Aerogen Pro-4

Introduction The Aerogen® Pro System is a portable medical device for multiple patient use that is intended to aerosolise physician-prescribed medications for inhalation that are approved for use with a general purpose nebuliser. This device can be used by patients on and off mechanical ventilation or other positive pressure breathing assistance. Aerogen® Pro is suitable for use by neonate, paediatric to adult patients as described in this manual. It incorporates the Aerogen Vibronic® aerosol generator. Aerogen® Pro is intended for hospital use only. It is designed to operate inline with standard...

 Open the catalog to page 4
Aerogen Pro-5

Aerogen Pro System The Aerogen Pro System includes the following components: 1 Figure 1.  Aerogen Pro System 1. Nebuliser with Filler Cap 2. T-Piece (Adult) with Plug 3. Aerogen Pro Controller 4. Controller Cable 5. AC/DC Adapter 6. Universal Mounting Bracket & Equipment Mount Adapter Note:  Visit www.aerogen.com for full parts list. Aerogen® Pro System Instruction Manual

 Open the catalog to page 5
Aerogen Pro-6

1. The nebuliser holds up to 10 mL of liquid medication. The nebuliser is translucent to allow visual monitoring of medication levels and aerosolisation. When the nebuliser is connected into the breathing circuit, the filler cap can be opened or removed from the nebuliser without causing loss of circuit pressure. Within the nebuliser is the Aerogen Vibronic® aerosol generator, which consists of a domed aperture plate with precision-formed holes that control the size of the aerosol droplets and a vibrational element that creates micro-pumping action to aerosolise medication. Gravity brings the...

 Open the catalog to page 6
Aerogen Pro-7

System Warnings Read and study all instructions before using the Aerogen Pro System and accessories. Only trained medical personnel should operate the device. During use observe for correct functioning of the nebuliser by regularly verifying aerosol is visible and no flashing indicator lights. Do not use a filter or heat-moisture exchanger (HME) between the nebuliser and patient airway. Do not attach a continuous supply of medication to the nebuliser; the device operates in 15 or 30 minute cycles. Clean, sterilise, assemble and perform a functional test (page 19) according to the instructions...

 Open the catalog to page 7
Aerogen Pro-8

Inspect all parts before use, and do not use if any parts are missing, cracked or damaged. In case of missing parts, malfunction or damage, contact your sales representative. Do not immerse or autoclave the controller or AC/DC adapter. Disassemble all parts before autoclaving. Use only with components specified by Aerogen. Do not use or store outside of specified environmental conditions. To avoid mechanical or electrical damage, do not drop the nebuliser or the controller. Do not use in the presence of devices generating high electromagnetic fields such as magnetic resonance imaging (MRI) equipment....

 Open the catalog to page 8
Aerogen Pro-9

Aerogen Pro Controller On/Off Power • 15 Min. - Press & Release • 30 Min. - Press & hold for 3 Sec. Error Indicator 30 Minute Mode Indicator light 15 Minute Mode Indicator light Battery Status Indicator 9V DC Input Controller Cable Input Figure 2.  Aerogen Pro Controls & Indicators Aerogen® Pro System Instruction Manua

 Open the catalog to page 9
Aerogen Pro-10

Table 1.  Aerogen Pro Controls & Indicators Control / Indicator Green (steadily lit) = 15 Minute nebulisation cycle on Green (flashing) = Low battery power Nebuliser automatically powers off after 15 minutes have elapsed Green (steadily lit) = 30 Minute nebulisation cycle on Green (flashing) = Low battery power Nebuliser automatically powers off after 30 minutes have elapsed On/Off Power Button Amber = Faulty electrical connection Error Indicator To operate in 15 Minute Mode, press and immediately release the On/Off button To operate in 30 Minute Mode, press and hold the On/Off button for at...

 Open the catalog to page 10
Aerogen Pro-11

Assembly & Installation Aerogen Pro System Set-Up Clean and sterilise the nebuliser and T-piece(s) as described in the Cleaning, Disinfection and Sterilisation section of this manual. Note:  The nebuliser and T-piece, as packaged, are not sterile. 1 Perform a functional test of Aerogen Pro before use and between patients as described in the Functional Test section of this manual (see page 19). 2 Insert the filler cap into the opening on the nebuliser. 3 Connect the nebuliser to the T-piece by pushing the nebuliser firmly onto the T-piece (Figure 3). Figure 3.  Connecting nebuliser to T-piece...

 Open the catalog to page 11
Aerogen Pro-12

Connect the Aerogen Pro controller and the nebuliser together using the controller cable (Figure 4). Figure 4.  Connecting controller and nebuliser 5 To operate on AC power (the primary mode of operation), connect the Aerogen Pro AC/DC adapter to the Aerogen Pro controller and plug the adapter into an AC power source (Figure 5). Figure 5.  Connecting the AC/DC Adapter

 Open the catalog to page 12
Aerogen Pro-13

The Aerogen Pro can be battery-operated for portable applications. The rechargeable battery can power the System for up to 45 minutes when fully charged. In the case of AC power failure the controller will automatically switch to battery operation. Note:  Allow a minimum of four hours for the internal battery to fully recharge. Note:  To ensure uninterrupted operation of the Aerogen Pro, secure both the AC/DC adapter cable and the controller cable so they cannot become disconnected during treatment. If clips are available on patient circuits, run the cables through the eyes of the clips. If clips...

 Open the catalog to page 13
Aerogen Pro-14

Installation for use with a Ventilator Connection to a Breathing Circuit 1. For adult breathing circuits, connect the nebuliser with adult T-piece into the inspiratory limb of the breathing circuit before the patient Y (Figure 6). Figure 6.  Connecting the Aerogen Pro to an adult breathing circuit Note:  Figure 6 shows adult configuration only 2. For paediatric breathing circuits, connect the nebuliser with the paediatric T-piece into the inspiratory limb of the breathing circuit before the patient Y (Figure 7). Figure 7.  Connecting to a paediatric breathing circuit

 Open the catalog to page 14