Minimate Installation and use manual

Open the catalog to page 1
Minimate - 2

Via Torino, 3 20030 Senago (MI) Italy www.luzzani.eu E-mail: info@luzzani.com Tel.: +39 029988433 Fax: +39 0299010379 ED. 3 REV. 0

Open the catalog to page 2
Minimate - 3

GENERAL GENERAL CHARACTERISTICS SHAPES CONTROLS CE MARKING 3 — IDENTIFICATION DATA AND WARRANTY 3.1 — MANUFACTURING BATCH 3.2 — WARRANTY 4 — PACKAGING 5 — TECHNICAL CHARACTERISTICS 5.1 — OPERATING CONDITIONS 5.2 — TRANSPORT AND STORAGE CONDITIONS 6 — INSTALLATION AND CONNECTIONS 6.1 — CONNECTION TO HYDRAULIC SYSTEM 6.2 — CONNECTION TO COMPRESSED AIR SYSTEM 6.3 — CONNECTION OF STRESS ABSORBER CORD 6.4 — NOTES FOR CORRECT CONNECTION 7 — NORMAL USE 7.1 — INSUFFLATION OF COLD WATER 7.2 — INSUFFLATION OF COLD AIR 7.3 — COMBINED INSUFFLATION OF COLD WATER AND AIR (SPRAY) 7.4 — FIRST TIME USE AND...

Open the catalog to page 3
Minimate - 4

Water button Handpiece Blue air line Green water line Stress absorber cord

Open the catalog to page 4
Minimate - 5

1 —— WARNINGS 1.1 — Any unauthorised tampering, modification or improper use immediately terminates the warranty and exonerates our company from all liability for injury or damage to persons, animals or property that may be caused by such interference. out in European Council Directive 93/42/EEC: Class II A 1.2 — The stress absorber cord must be anchored to the dental unit. This cord is designed to absorb any strains, thus preventing abnormal traction on the electrical or air/water line connections. 2.1 — GENERAL The Minimate syringe is a medical device designed to blow air and water...

Open the catalog to page 5
Minimate - 6

4 —— PACKAGING The product is shipped in suitable packaging to prevent problems during transport. The packaging consists of a plastic bag containing the Minimate syringe. A protective handpiece sheath and tip-saving tube protect the syringe during transport. Several bags are placed in one box. The Minimate syringe comes ready for connection to the dental unit once all packaging has been removed. For the correct use of the syringe, the protective tube must also be removed from the tip. IMPORTANT NOTE: With each syringe, the package also includes a User’s manual which also includes a “Fault...

Open the catalog to page 6
Minimate - 7

To insufflate cold air into the operating field, just press the right button on the handpiece: 6.1 — CONNECTION TO HYDRAULIC SYSTEM The syringe’s green line must be hooked up to the water supply. NOTE • The water used must be potable water, filtered (<25 μm) and free of bacteria, etc. For the syringe to function properly, the water pressure must not be lower than indicated. 6.2 — CONNECTION TO COMPRESSED AIR SYSTEM 7.3 — COMBINED INSUFFLATION OF COLD WATER AND AIR (SPRAY) The syringe’s blue line must be connected to the compressed air system. To blow a combination of cold air and water...

Open the catalog to page 7
Minimate - 8

Flush out the syringe at the beginning of each work day (minimum flushing time: 2 minutes) and before each patient (minimum flushing time: 20-30 sec.). Immediately upstream of the syringe, install filters able to retain the microorganisms coming from the hydropneumatic circuit. wipe with a damp cloth, removing any stains. Set in a steam autoclave at 134°C for AT LEAST 3 minutes (in compliance with CEI EN 13060). 8 —— CLEANING AND STERILIZATION After each use on a patient, the handpiece and tip of the syringe MUST be cleaned and sterilised to guarantee maximum hygiene. Sterilisation symbol...

Open the catalog to page 8
Minimate - 9

12 —— FAULT REPORT FORM To meet the requirements of Directive 93/42/CEE as amended, the company has established a post-marketing surveillance procedure to monitor any problems generated by the use of our products. This commitment includes the requirement that both user and manufacturer inform the competent authorities of any incident caused to patient or user by malfunction or deterioration in the characteristics and/or performance of a device, as well as any inadequacy in the instructions for use. We kindly ask you to inform us of any anomalies by sending us the sheet attached to the last...

Open the catalog to page 9
Minimate - 10

FAULT REPORT FORM PRODUCT _______________________________________________________________________________________________________ TYPE _________________________________________________ BATCH ___________________________________________________ REPORTED BY ___________________________________________________________________________________________________ COMPANY ______________________________________________________________________________________________________ _________________________________________________________________________________________________________________...

Open the catalog to page 10