Product catalog summary
Overview: The AIRFAN blower, specifically the MFA0290 model, is designed for use in ventilators for Acute, Sub-Acute, and Home Care settings. It features a patented motor cooling system and a unique soundproof housing.
Specifications:
  • Derived from the MFA0300 model.
  • Maximum pressure (ΔP max) of 76 hPa at 24V when fully closed.
  • Maximum flow rate (Q max) of 270 l/min at 24V when fully opened.
  • Typical power consumption: 11.5 W at 30 hPa and 30 l/min, 22.6 W at 60 hPa.
  • Expected lifetime of over 40,000 hours under standard operating conditions.
Performance:
  • Dynamic transient time: < 95 ms at 3A max, < 88 ms at 6A max, < 84 ms at 10A max.
  • Noise levels: < 29 dBa at 10 hPa, < 41 dBa at 30 hPa, < 48 dBa at 60 hPa, measured with blower inlet at 1m.
Customization Options:
  • Outlet port options: male or female connector (22Ø ISO5356-1) or additional auxiliary ports.
  • O-ring for air inlet and custom labeling available.
  • Wire length: standard 150 mm or customizable.
Technical Data:
  • Voltage supply: 24 VDC nominal, 28 VDC maximum.
  • Motor type: Brushless with Hall-effect sensors.
  • Dimensions: 130 x 100 x 61 mm (5.3 x 3.9 x 2.4 in).
  • Weight: 0.44 kg (approximately 1.00 lb).
  • Operational in any position or orientation.
  • Temperature range: +10 to +50°C ambient.
  • Humidity: 15 to 95% RH non-condensing.
  • Atmospheric pressure: 660 to 1060 hPa.
  • No oxygen compatibility.
See more

Catalog excerpts

Airfan MFA0290 Datasheet-1

Blower for Performance & Reliability AIRFAN blower architecture with patented motor cooling system and unique soundproof housing. The MFA0290 is derived from the MFA0300 for implementation in ventilators dedicated to Acute, Sub-Acute and Home Care. fully closed Dynamic transient time * > 40 000 hours* Casing integration * LifeTime expectancy based on standard operating conditions Noise Level * * with Blower inlet @1m 02 after the turbine flow to patient cooling flow out

 Open the catalog to page 1
Airfan MFA0290 Datasheet-2

Voltage supply Motor type Dimensions (casing) Atmospheric pressure Oxygen compatibility Brushless with Hall-effect sensors COMPRESSOR MAP Static AP vs Airflow at Constant PWM and Motor Voltage PWM = Pulse Width Modulation AIRFLOW (l/mn) at Sea level conditions - Performance are achieved with AIRFAN controller MFA0900 and nominal Motor Ke Outlet port male or female connector (220 ISO5356-1) or additional auxiliary ports O 0 O-ring for air inlet // 0 Custom label CUSTOMISE TO YOURSPECIFICATIONS Wire Length 150 or custom All values listed are typical values with tolerance classification grade 4...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.