1. Catalogs
  2. Airfan
  3. Airfan MFA0500 Datasheet

Airfan MFA0500 Datasheet

Airfan MFA0500 Datasheet

Airfan MFA0500 Datasheet

Product catalog summary
Specifications:
  • Static Pressure: 110 hPa at maximum 24V.
  • Free Flow Rate: 380 l/min at maximum 24V.
  • Power Consumption: 12.5 W at 30 hPa and 30 l/min; 22.1 W at 60 hPa.
  • Lifetime: Over 40,000 hours (L10) under standard conditions.
  • Voltage Supply: 24 VDC nominal, 28 VDC maximum.
  • Motor Type: Brushless with Hall-effect sensors.
  • Dimensions: 130 x 100 x 61 mm (5.3 x 3.9 x 2.4 in).
  • Weight: 0.45 kg (approximately 1.00 lb).
  • Operating Temperature: +10 to +50°C ambient.
  • Humidity: 15 to 95% RH non-condensing.
  • Atmospheric Pressure: 660 to 1060 hPa.
  • Oxygen Compatibility: 100% O2 compatible.
Performance:
  • Noise Levels: Less than 29 dBa at 10 hPa, less than 40 dBa at 50 hPa, and less than 46 dBa at 100 hPa.
  • Dynamic Transient Time: Less than 113 ms at 3A max, less than 85 ms at 6A max, less than 68 ms at 10A max.
Design Features:
  • Specific blower architecture with a patented motor cooling system independent from the air inlet flow.
  • Ensures a homogeneous Air-Oxygen blend with patient safety in case of temperature increase.
  • Segregation to avoid patient air contamination.
Customisation Options:
  • Outlet Port: Male or female connector (22Ø ISO5356-1) or additional auxiliary ports.
  • Inlet Filter: For motor cooling.
  • O-ring for air inlet and low leakage sealing as per ISO80601.
  • Wire Length: 150 or custom length.
  • Custom Labeling available.
Integration:
  • Works in any position or orientation.
  • Casing Integration Type.
Compliance:
  • All values are typical with tolerance classification grade 4 according to ISO 13448.
See more

Catalog excerpts

Airfan MFA0500 Datasheet-1

Blower for Blended or 100% Oxygen Specific blower architecture with patented motor cooling system completely independent from the air inlet flow. The MFA0500 guarantees a fully homogeneous Air-Oxygen blend with complete patient safety in the case of an increase in temperature. static pressure Dynamic transient time * Noise Level * Casing integration * with Blower inlet @1m * LifeTime expectancy based on standard operating conditions Segregation to avoid patient air contamination motor cooling air inlet flow to patient (with blended O2) cooling flow out

 Open the catalog to page 1
Airfan MFA0500 Datasheet-2

TECHNICAL DATA Voltage supply Motor type Brushless with Hall-effect sensors Dimensions (casing) Atmospheric pressure Oxygen compatibility CUSTOMISATION  Outlet port male or female connector (22Ø ISO5356-1) or additional auxiliary ports Outlet port  Inlet filter for motor cooling  Low Leakage sealing as per ISO80601  O-ring for air inlet CUSTOMISE TO YOUR SPECIFICATIONS  Wire Length 150 or custom All values listed are typical values with tolerance classification grade 4 according to stand

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.