MFA0500 Datasheet

MFA0500 Datasheet

MFA0500 Datasheet

Product catalog summary
Technical Specifications
  • Static Pressure: 110 hPa at maximum 24V.
  • Free Flow Rate: 380 l/min at maximum 24V.
  • Power Consumption: 12.5 W at 30 hPa and 30 l/min; 22.1 W at 60 hPa.
  • Lifetime: Over 40,000 hours under standard operating conditions.
  • Dynamic Transient Time: Less than 113 ms at 3A max, 85 ms at 6A max, and 68 ms at 10A max.
Design and Functionality
  • Blower designed for blended or 100% oxygen use.
  • Features a patented motor cooling system independent of air inlet flow.
  • Ensures homogeneous air-oxygen blend with patient safety during temperature increases.
Performance
  • Noise Levels: Less than 29 dBa at 10 hPa, 40 dBa at 50 hPa, and 46 dBa at 100 hPa.
  • Includes segregation to avoid patient air contamination.
Customization Options
  • Outlet port options include male or female connectors (22Ø ISO5356-1) or additional auxiliary ports.
  • Inlet filter for motor cooling and O-ring for air inlet.
  • Low leakage sealing as per ISO80601.
  • Customizable wire length and labeling.
Technical Data
  • Voltage Supply: 24 VDC nominal, 28 VDC maximum.
  • Motor Type: Brushless with Hall-effect sensors.
  • Dimensions: 130 x 100 x 61 mm (5.3 x 3.9 x 2.4 in).
  • Weight: 0.45 kg (approximately 1.00 lb).
  • Integration: Operates in any position or orientation.
  • Temperature Range: +10 to +50°C ambient.
  • Humidity Range: 15 to 95% RH non-condensing.
  • Atmospheric Pressure: 660 to 1060 hPa.
  • Oxygen Compatibility: 100% O2 compatible.
See more

Catalog excerpts

MFA0500 Datasheet-1

Blower for Blended or 100% Oxygen Specific blower architecture with patented motor cooling system completely independent from the air inlet flow. The MFA0500 guarantees a fully homogeneous Air-Oxygen blend with complete patient safety in the case of an increase in temperature. static pressure Dynamic transient time * Noise Level * Casing integration * with Blower inlet @1m * LifeTime expectancy based on standard operating conditions Segregation to avoid patient air contamination motor cooling air inlet flow to patient (with blended O2) cooling flow out

 Open the catalog to page 1
MFA0500 Datasheet-2

TECHNICAL DATA Voltage supply Motor type Brushless with Hall-effect sensors Dimensions (casing) Atmospheric pressure Oxygen compatibility CUSTOMISATION  Outlet port male or female connector (22Ø ISO5356-1) or additional auxiliary ports Outlet port  Inlet filter for motor cooling  Low Leakage sealing as per ISO80601  O-ring for air inlet  Custom label  Wire Length 150 or custom All values listed are typical values with tolerance classification grade 4 according to standard ISO 13448

 Open the catalog to page 2

All Airfan catalogs and technical brochures

  1. MFA0114

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.