Amaro 5000 - Biosafety BL3 - 1

Product Description Steam Sterilizer AMARO 5000 Series

Open the catalog to page 1
Amaro 5000 - Biosafety BL3 - 2

Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 2
Amaro 5000 - Biosafety BL3 - 3

These equipments have a high quality standard, are fully tested in order to grant a high service level and a good performance. The steam sterilizers AMARO 5000 series have been designed to answer the most demanding requirements of steam sterilization of medical devices and in accordance with the European and ISO standards and regulations. The steam sterilizers AMARO 5000 series are also manufactured in a version for laboratory purposes. The steam sterilizers AMARO 5000 LAB BIO series is particularly directed to provide sterilization service to the following market segments:  Laboratories...

Open the catalog to page 3
Amaro 5000 - Biosafety BL3 - 4

Instruments and glassware Sterilization of utensils, instruments and laboratory glassware. The new line of steam sterilizers AMARO 5000 LAB BIO responds also to the concerns of today: - Saving water and energy; - Security and ease of use; - Ergonomics and safety at work. - Allow the validation of sterilization processes required for the certification and quality management system of sterilization services; The AMARO 5000 LAB BIO series is a free floor standing, front loading sterilizer. The equipment, optionally, has a built in steam generator. The available models in the steam sterilizers...

Open the catalog to page 4
Amaro 5000 - Biosafety BL3 - 5

The dimensions and capacities of each chamber models of sterilizers AMARO 5000 LAB BIO are listed in the following table: Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 5
Amaro 5000 - Biosafety BL3 - 6

Sterilization Programs The steam sterilizers AMARO 5000 LAB BIO series have the following available working programs, accessible to the operator: Executable cycles and corresponding phases  CYCLE 1: Heating up cycle. Cycle for the pre-heating of the sterilizer chamber Steam admission Post treatment Pressure [mbar] Product Description – AMARO 5000 LAB BIO Series

Open the catalog to page 6
Amaro 5000 - Biosafety BL3 - 7

• CYCLE 2: Bowie & Dick test cycle Cycle to detect air pouches, vacuum pump malfunction and verification of steam penetration conditions Phases: - Pre-conditioning - Steam admission - Sterilization at 134° C - Exhaust - Post treatment - Airing - End Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 7
Amaro 5000 - Biosafety BL3 - 8

• CYCLE 4: Cycle for wrapped Instruments Cycle for glassware, utensils and wrapped surgical instruments. Phases: - Pre-conditioning - Steam admission - Sterilization at 134° C - Exhaust - Post treatment - Airing - End Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 8
Amaro 5000 - Biosafety BL3 - 9

• CYCLE 10: Cycle for biologic risk level 3 Cycle for materials of bio hazard risk level 3 Phases: - Pre-conditioning - Steam admission - Sterilization at 134° C - Rapid Exhaust - Post treatmen - Airing - End Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 9
Amaro 5000 - Biosafety BL3 - 10

CYCLE 12(*): Laboratory cycle. Pre vacuum and fast exhaustion - Pre-conditioning - Steam admission - Sterilization 121° C - Exhaust - Post treatment - End Product Description - AMARO 5000 LAB BIO Series

Open the catalog to page 10
Amaro 5000 - Biosafety BL3 - 11

CYCLE 14(1): Laboratory cycle. Pre vacuum and slow exhaustion - Pre-conditioning - Steam admission - Sterilization 121° C - Exhaust - Post treatment - End Product Description - AMARO 5000 LAB BIO Series Cycles 12 to 15 are specially designed to meet laboratory specific applications. Time and sterilization temperature band can be selected by authorized supervisor.

Open the catalog to page 11
Amaro 5000 - Biosafety BL3 - 12

CYCLE 16: Cycle for cleanness of the chamber (maintenance) Cycle for the cleanness of the chamber, at day start, when chamber is cold. - Open the load door - Close the load door Note: As option it is possible to provide other sterilization cycles, as client specification. The authorized supervisor has permission to modify parameters (time and sterilization temperature band, designation and description of laboratory cycles. Default parameters as mentioned above. Mechanical features The sterilization body consists of a chamber and integral jacket that assures a high heating efficiency and...

Open the catalog to page 12
Amaro 5000 - Biosafety BL3 - 13

Door locking safety system, preventing door opening in case of power or compressed air failure. The sterilizer is equipped with a 0,01 pm (micron) bacteriological filter to equalize the chamber pressure to atmospheric at the end of drying phase, for admission of sterile air into the chamber. For bio hazard laboratory cycles a special sterilized bacteriological filter element is used for gaseous treatment before release to atmospheric vent. 5. 1 Safety and safety devices The sterilizer chamber and the jacket are each equipped with safety valve against overpressure with automatic differential...

Open the catalog to page 13
Amaro 5000 - Biosafety BL3 - 14

The steam inlets into the sterilization chamber are placed asymmetrically in its lateral side. Pressure and temperature are separated and individually controlled in the chamber and jacket within the pressure and temperature range. The Sterilizer is equipped with emergency pushbuttons, in the load and unload side, that cut out the power supply to the machine. The AMARO 5000 is equipped with a mechanical lock device that keeps the door from opening, while there is pressure inside the sterilizer chamber. In the cooling and condenser tank, the water level safety device prevents the system...

Open the catalog to page 14
Amaro 5000 - Biosafety BL3 - 15

Pressure transducers are used one in the sterilization chamber and another at the double body (jacket), have a scale of 0 to 4000 hPa (absolute). Normalized validation ports for external temperature and pressure sensors for validation and qualification purposes, as per EN 285 standard. Information available in the operator-interface console of the unload door side:  Closed doors Cycle complete Failure occurrence (Error Message) Door open Chamber temperature Jacket temperature (Optional) 5.3 Performance and control The sterilizer is equipped with a microprocessor that is programmed in a way...

Open the catalog to page 15
Amaro 5000 - Biosafety BL3 - 16

Failure and statistics are permanently registered in memory. Note: The sterilizer can process various fixed cycles and cycles with programmable parameters by the user, or the client’s specifications. For the cycles programmed by user, once the cycle is defined it is kept in the microprocessor’s memory, remaining the active cycle until the next modification. The operator does not have to alter the active cycle unless it is desired. The temperature of the sterilization plateau is calculated by F0 factor, allowing, in this way, to assure that, in the beginning of the sterilization time count,...

Open the catalog to page 16