Vacutron® Suction Regulators - 1

Vacutron Suction Regulators ® & Accessories Guide

Open the catalog to page 1
Vacutron® Suction Regulators - 2

Vacutron® Suction Regulators by Chemetron® Chemetron has been a leader in suction regulators for more than 40 years. The Vacutron line offers safety and performance features unlike any other regulator on the market. Vacutron provides four distinct advantages over other suction regulators.  Superior Patient Safety Vacutron suction regulators have a secondary internal vacuum relief valve. This valve ensures that under no circumstance, i.e. component failure, can the vacuum pressure delivered to the patient exceed a pre-set safe level so the patient is not inadvertently exposed to the full...

Open the catalog to page 2
Vacutron® Suction Regulators - 3

Vacutron® Suction Regulators by Chemetron® Easy access mode selection knob gives instant access to regulated, intermittent and full-line modes, depending on the model. High-impact polycarbonate case for strength and durability. Color-coded glow-in-the-dark analog gauge provides easy readability under typical hospital lighting conditions. Oversized vacuum regulating knob ensures precise vacuum control. Weight-bearing brass posts prevent case from cracking. A full line of inlet and outlet fittings ensures compatibility with existing fixtures in any medical facility. Model-specific color-coded...

Open the catalog to page 3
Vacutron® Suction Regulators - 4

Continuous Model The continuous suction regulator is designed to deliver constant vacuum for general suction procedures, including routine maintenance of the patient airway, oropharyngeal and endotracheal suctioning. Key Features  Internal safety relief valve limits the maximum vacuum level to 320 mm Hg in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available Specifications Operating Modes: OFF and Regulated Regulated Vacuum Range: 0-300 mm Hg Safety Relief Valves: 1...

Open the catalog to page 4
Vacutron® Suction Regulators - 5

Pediatric Continuous Model The pediatric continuous suction regulator is designed to deliver constant vacuum for general suction procedures on pediatric patients, including routine maintenance of the airway, oropharyngeal and endotracheal suctioning. Key Features  Internal safety relief valve limits the maximum vacuum level to 140 mm Hg in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available Specifications Operating Modes: OFF and Regulated Regulated Vacuum Range:...

Open the catalog to page 5
Vacutron® Suction Regulators - 6

Low Vacuum (Neonatal) Continuous Model The low vacuum continuous suction regulator is designed to deliver constant vacuum for general suction procedures on neonatal patients, including routine maintenance of the airway, oropharyngeal and endotracheal suctioning. Key Features  Internal safety relief valve limits the maximum vacuum level to 100 mm Hg in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available Specifications Operating Modes: OFF and Regulated Regulated...

Open the catalog to page 6
Vacutron® Suction Regulators - 7

High Vacuum Continuous Model The high vacuum suction regulator is designed for procedures that require constant vacuum pressures in excess of 300 mm Hg. There are two operational modes: regulated suction and full line. Regulated suction allows the clinician to set the vacuum level between 0 and the vacuum line pressure. On full line mode, the vacuum pressure delivered is unregulated and equal to the vacuum line pressure. Key Features  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available Specifications Operating...

Open the catalog to page 7
Vacutron® Suction Regulators - 8

Surgical Continuous Model The surgical suction regulator is designed for use in surgical suites. There are two operational modes: regulated vacuum and full line. Regulated suction allows the clinician to set the vacuum level between 0 and 300 mm Hg. On full line mode, the vacuum pressure delivered is unregulated and equal to the vacuum line pressure. Key Features  Internal safety relief valve limits the maximum vacuum level on regulated mode to 320 mm Hg in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in...

Open the catalog to page 8
Vacutron® Suction Regulators - 9

Continuous/Intermittent Model The continuous/Intermittent suction regulator is designed to support a variety of procedures requiring constant or intermittent suction. Constant suction is used primarily to maintain a patient’s airway, while intermittent suction is used for drainage procedures. Key Features  Dual safety relief valves limit the maximum vacuum level on both constant and intermittent modes in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available...

Open the catalog to page 9
Vacutron® Suction Regulators - 10

Pediatric Continuous/Intermittent Model The pediatric continuous/intermittent suction regulator is designed to support a variety of procedures requiring constant or intermittent suction on pediatric patients. Constant suction is used primarily to maintain the patient airway, while intermittent suction is used for drainage procedures. Key Features  Dual safety relief valves limit the maximum vacuum level on both constant and intermittent modes in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source...

Open the catalog to page 10
Vacutron® Suction Regulators - 11

Intermittent Model Intermittent suction regulators are designed to deliver intermittent vacuum only. This regulator is used when clinicians want to ensure that constant suction is never delivered. Key Features  Internal safety relief valve limits the maximum vacuum level to 170 mm Hg in the event of regulator failure or inaccuracy  Superior regulation mechanism ensures stable vacuum settings even with fluctuations in source vacuum  ISO configurations available Specifications Operating Modes: OFF and Intermittent Regulated Vacuum Range: 0-150 mm Hg Cycle Times: 17 seconds of suction...

Open the catalog to page 11