AmcareMed Oxygen Generator

AmcareMed Oxygen Generator

AmcareMed Oxygen Generator

Product catalog summary
Overview
The document provides detailed information about the AmcareMed medical oxygen generator system, which utilizes PSA (Pressure Swing Adsorption) technology to supply pure oxygen to hospitals and clinics. The system is designed to be safe, economical, convenient, and independent, with the ability to customize installations based on customer site conditions.

Specifications
The oxygen generator system offers a capacity range from 5Nm³/hr to 100Nm³/hr, with an operating temperature between 0℃ and 38℃. The oxygen outlet pressure is adjustable at 0.4Mpa, and the air feeding pressure ranges from 0.7 to 0.8Mpa. The system ensures a purity level of 93%±3%.

Features
The system features automatic and stable operation with real-time monitoring and alarm functions. It boasts higher separation efficiency compared to similar PSA oxygen generators, requiring less air to produce more oxygen. The modular design facilitates easy installation and maintenance.

Working Principle
The system comprises an air compressor, dryer, PSA oxygen generator, and booster compressor. Air is compressed, filtered, and stored before entering the PSA generator, where nitrogen is adsorbed, allowing oxygen to be collected and stored for use.

Oxygen Standards
The oxygen produced meets the European Pharmacopoeia requirements for medical oxygen, with specific limits on impurities such as carbon monoxide, carbon dioxide, nitrogen oxides, sulfur dioxide, oil, and water.

Main Components
The system includes an oxygen tank, PSA oxygen generator, LCD control panel, filters, dryer, air compressor, and booster compressor. Each component plays a crucial role in ensuring the quality and efficiency of oxygen production.

Containerized Oxygen Generator
This mobile, plug-and-play system is pre-completed in a container, offering low cost, small area, and easy maintenance. It is suitable for various industries and can operate under extreme conditions.

Oxygen Cylinder Filling System
AmcareMed provides high-pressure cylinder filling systems capable of filling 10 to 100 medical oxygen cylinders daily. These systems offer autonomy to healthcare facilities and can be equipped with optional features to enhance functionality.

Contact Information
AmcareMed Technology Co., Limited is located in Zhejiang Province, China, with contact details provided for further inquiries.
See more

Catalog excerpts

AmcareMed Oxygen Generator-1

Medical Oxygen Generator PSA technology Professional oxygen generator manufacturer Safe economic convenient and independent

 Open the catalog to page 1
AmcareMed Oxygen Generator-2

OXYGEN GENERATOR AmcareMed medical oxygen generator system could supply pure oxygen for all kinds of hospitals and clinics continuously. The system uses air as raw material and zeolite molecular sieve as adsorbent to separate oxygen from air by pressure swing adsorption (PSA) technology. The purity is 93%±3%. In addition to the existing oxygen generators series, we can also customize the oxygen generator system according to the installation conditions of the customer site, and can be installed on site for customers. FEATURES Capacity: 5Nm³/hr to 100 Nm³/hr Oxygen outlet pressure: 0.4Mpa(adjustable)...

 Open the catalog to page 2
AmcareMed Oxygen Generator-3

WORKING PRINCIPLE Oxygen generator system mainly consists of air compressor, dryer, PSA oxygen generator, and booster compressor. The air is compressed to working pressure by screw air compressor, and then enters dryer through filter to remove oil, water and dust. Finally enter the air tank to store necessary air for PSA generator. Clean compressed air enters PSA generator to separate oxygen and nitrogen. Adsorption tower filled with zeolite molecular sieves(ZMS). ZMS adsorb nitrogen and allow oxygen to stream through. Oxygen accumulates on the top of the tower, then enter to oxygen tank. After...

 Open the catalog to page 3
AmcareMed Oxygen Generator-4

OXYGEN STANDARD European Pharmacopoeia requirements for medical Oxygen Oxygen Carbon Monoxide Carbon Dioxide Nitrogen Monoxide and Dioxide Sulphur Dioxide Working dewpoint MODEL Oxygen Capacity Model Booster Compressor Installation Area Feed air required (m³/min) Single line solution Dual line solution Note:The installation area is not an absolute area, we will design according to the specific location. Note:The installation area is not an absolute area, we will design according to the specific location.

 Open the catalog to page 4
AmcareMed Oxygen Generator-5

MAIN PARTS Air Compressor The air compressor is used to supply compressed air to the oxygen generator. Power: 5.5~132kw. FAD: maintenance. from air ensuring that air quality is in accordance with relevant international standards for oil residual content. Lower temperature air is better for the efficiency of the molecular sieve. Based on PSA technology, molecular sieve adsorbs nitrogen from the air and separates the oxygen. Using two adsorption tower process, two tower cycle alternately, continuous output of high quality oxygen. Oxygen Tank It collects the necessary volume of oxygen from PSA oxygen...

 Open the catalog to page 5
AmcareMed Oxygen Generator-6

Containerized Oxygen Generator The container oxygen generator is a complete set of oxygen generators pre-completed in a container. It is a mobile plug-and-play device, with the advantages of low cost, small area, simple operation, easy maintenance, quick on-site oxygen production, convenient switching and no pollution. The containerized oxygen generator system can be regarded as a mobile oxygen center, which can continuously supply oxygen for different work sites. It can be widely used in medical, petrochemical, ozone production and other industries and fields. The system works well even in extreme...

 Open the catalog to page 6
AmcareMed Oxygen Generator-7

Oxygen Cylinder Filling System AmcareMed offers a full range of high-pressure cylinder filling systems allowing to fill daily 10 to 100 medical oxygen cylinders at a pressure of 150 to 200 bar. Designed to be connected to an oxygen generator and mounted on the wall with fixed brackets.The systems provide full autonomy to healthcare facilities: they can easily manage peak consumption flows and fill their own cylinders. Depending on the use, optional products such as gas heaters, backfire preventers, pressure relief valves, gas leak alarms, pressure alarms, pressure switches, and bottle guard brackets...

 Open the catalog to page 7