ARTHOSPico
1 / 3Pages

Catalog excerpts

ARTHOSPico - 1

INTERVENTIONAL CARDIOLOGY Stent Delivery System Cobalt-Chromium with Hydrophilic Coating

Open the catalog to page 1
ARTHOSPico - 2

Stent Delivery System Cobalt-Chromium with Hydrophilic Coating COMPLIANCE TABLE Pressure (bar) Excellent outward radial force in a low profile stent Minimal strut thickness (65 μm) for optimal performance and small vessel stenting Co-Cr stent provides strength and stability, while maintaining a high level of flexibility and vessel conformity The ARTHOS Stent System is a multicellular Cobalt-Chromium stent that offers strength, stability, and flexibility for treating the most difficult lesions. Pico Nominal pressure DELIVERY SYSTEM Highly flexible delivery catheter to facilitate navigation...

Open the catalog to page 2
ARTHOSPico - 3

Stent Delivery System Cobalt-Chromium with Hydrophilic Coating TECHNICAL SPECIFICATIONS Stent Description Recommended Guidewire Recommended Guiding Catheter PTCA Entry Profile Crossover Profile Nominal Pressure Rated Burst Pressure (RBP) CoCr Stent Delivery System 0.014” (0.36 mm) 5F min 0.45 mm (0.017”) 0.85 mm (0.035’’) 9 bar 16-18 bar • 5 Micron (0.0026”) small stent 6 • 5 Micron (0.0037”) large stent 9 Strut Thickness • ean recoil of CS-large (QM580): < 3% M • ean recoil of CS-small (QM144): < 1.5% M Stent Recoil Values • ean Foreshortening of CS-large (QM580): < 5% M • ean...

Open the catalog to page 3

All Amg International catalogs and technical brochures

  1. ISTAR

    2 Pages

  2. FACILE

    3 Pages

  3. ITRIXII

    2 Pages