video corpo

ZEEnit P series
5Pages

{{requestButtons}}

Catalog excerpts

ZEEnit P series - 1

Atomic Absorption Spectrometer Technical Data ZEEnit® series | Update 07/2014 | OBue Analytik Jena AG | Konrad-Zuse-Str. 1 | 07745 Jena/ Germany | www.analytik-jena.com | info@analytik-jena.com

Open the catalog to page 1
ZEEnit P series - 2

ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with “Third Generation” Magnetic Field Control. ZEEnit 650 P ZEEnit 700 P Flame technique Graphite furnace technique Technique HydrEA technique Graphite furnace technique Hydride technique HydrEA technique Direct solid sampling technique Wavelength range Optical bench Optimized Czerny-Turner-Design for maximum energy throughput with two focal lengths, automated wavelength selection, peaking and slit selection Real double beam mode Free choice of single and double beam mode Encapsulated optics...

Open the catalog to page 2
ZEEnit P series - 3

Flame technique Spray chamber All-titanium single slot burner Reproducible burner rotation Optional automatic cleaning device for 5 cm burner head (Scraper) PPS spray chamber with mixing wing for aqueous and organic solutions Fully computer controlled Total Flow Gas box Computer controlled automated flame ignition Automatic input of additional gases for organic solvents Safety functions Sensor control of the burner head Check of the siphon system Automatic shut down of gases if the flame is not detected or in case of a system power failure or as a defined action during a multi routine...

Open the catalog to page 3
ZEEnit P series - 4

Graphite furnace technique Integrated computer-controlled transverse heated Zeeman graphite furnace Graphite tube is transverse heated to provide a constant temperature profile Independent gas controls for the external and internal gas flows around and through the tube allow easy removal of volatile matrixes whilst protecting the tube against interference from outside air during the internal gas stop hence maximizing tube life The advanced furnace concept, utilising the adaptive sensor-less temperature control and emission independent temperature control ensures constant, precise,...

Open the catalog to page 4
ZEEnit P series - 5

Direct solid sampling technique – solidAA® Analysis of the original solid sample, directly and without any laborious sample preparation Wide measurement range True micro-method Avoiding hazardous reagents Easy handling Immediate results High sensitivity Additional technical data Dimensions (W x H x D) Environmental requirements Power requirements 230 V (±10 %); 50/60 Hz, slow fuse with 35 A, 2100 VA, single phase alternating current Temperature +10 oC up to 35 oC Rel. Humidity max. 90 % at +30 oC Non condensing Technical Data ZEEnit® series | Update 07/2014 | OBue Analytik Jena AG |...

Open the catalog to page 5

All Analytik Jena GmbH+Co. KG catalogs and technical brochures

  1. UVP GelSolo

    8 Pages

  2. contrAA800

    8 Pages

  3. multi N/C

    16 Pages

  4. novAA800

    12 Pages

  5. multi X 2500

    12 Pages

  6. compEAct

    8 Pages

  7. TOPwave

    8 Pages

  8. novAA series

    12 Pages

  9. PlasmaQuant MS

    12 Pages

  10. multi N/C

    24 Pages

  11. multi EA 5000

    20 Pages

Archived catalogs

  1. multi EA 4000

    12 Pages

  2. SPECORD PLUS