Introduction
The document provides technical specifications and operational details for the Tethys HMD, a handpiece thermal disinfector accessory compatible with the Tethys H10 model. It highlights the ability to wash, thermally disinfect, and dry up to six instruments simultaneously, enhancing the functionality of existing Tethys H10 units with a firmware update.

Cycle Description
The handpiece cycle is completed in 34 minutes and includes washing, thermal disinfection, and drying phases. The thermal disinfection is effective on external surfaces, verified according to EN15883-1 and EN15883-2 standards.

Standard Equipment
The standard equipment includes:
  • 4 turbine adapters
  • 3 micromotor handpiece adapters
  • 1 scaler tip adapter (with 3 tips per adapter)
  • 4 plugs


Technical Specifications
The cycle operates at a thermal disinfection temperature of 90°C, with a washing time of 13 minutes and a thermal disinfection time of 4 minutes, achieving an A0 value of 3,000.

Phases of Operation
  1. Washing with detergent
  2. Rinsing
  3. Thermal disinfection
  4. Drying


Maximum Flexibility
The support for the handpiece cycle can be easily assembled and removed, allowing the Tethys H10 to retain its original function as an ultrasound thermal disinfector.

Adapters for Every Requirement
Specific adapters are available for treating turbines, micromotor handpieces, and scaler tips. Proper instrument treatment requires fitting the appropriate adapter for each instrument family.
See more

Catalog excerpts

Tethys HMD | Anthos-1

Thermal disinfection time Total cycle time Handpiece cycle Standard equipment - 4 turbine adapters - 3 micromotor handpiece adapters - 1 scaler tip adapter (3 tips per adapter) - 4 plugs BU Medical Equipment Sede legale ed amministrativa Headquarters Cefla s.c. Via Selice Provinciale, 23/a 40026 Imola - Bo (Italy) tel. +39 0542 653111 fax +39 0542 653344 Stabilimento Plant Via Bicocca, 14/c 40026 Imola - Bo (Italy) tel. +39 0542 653441 fax +39 0542 653601 According to the standards in force, in extra-EU areas the availability and specifications of some products and/or characteristics may vary. Please contact your local distributor for further information. Washing time Thermal disinfection temperature Cycle description Due to our policy of constant technological upgrading, technical specifications may be subject to change without prior notice. Technical specifications Revolutionary Evolut

 Open the catalog to page 1
Tethys HMD | Anthos-2

Thanks to the new HMD, Tethys H10 is now also a handpiece thermal disinfector Adapters for every requirement Users can treat the following instruments: - Turbines - Micromotor handpieces - Scaler tips This latest accessory makes it possible to wash, thermally disinfect and dry up to 6 instruments (turbines or micromotor handpieces) simultaneously. The HMD is compatible with all Tethys H10, including those that have already been installed. All the user has to do is update the firmware to make the thermal disinfector more complete than ever. To ensure proper instrument treatment all you need to...

 Open the catalog to page 2

All Anthos catalogs and technical brochures

  1. A | Anthos

    6  Pages

  2. Cart | Anthos

    8  Pages

  3. R7 | Anthos

    24  Pages

  4. C9 | Anthos

    23  Pages

  5. A7 | Anthos

    20  Pages

  6. A6 | Anthos

    18  Pages

  7. A5 | Anthos

    16  Pages

  8. A3 | Anthos

    12  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.