1. Catalogs
  2. AQUALABO
  3. C4E: CONDUCTIVITY /SALINITY
video corpo

C4E: CONDUCTIVITY /SALINITY

C4E: CONDUCTIVITY /SALINITY

C4E: CONDUCTIVITY /SALINITY

Product catalog summary
Applications
  • Urban wastewater treatment
  • Industrial effluent treatment
  • Surface water monitoring
  • Sea water
  • Drinking water
Mounting at 4 Electrodes

The sensor uses a 4-electrode technology with graphite and platinum electrodes. An alternating current of constant voltage is applied between the primary graphite electrodes, while the secondary platinum electrodes regulate the voltage to account for fouling. Conductivity is measured based on the resistance between the primary electrodes.

Digital Technology

The Digital C4E sensor stores calibration and history data, enabling a plug-and-play system without re-calibration. It uses the Universal Modbus RS485 protocol for connectivity with various devices.

  • 4 electrodes (2 graphite, 2 platinum)
  • Range: 0 to 200 mS/cm
  • Digital sensor with Modbus RS-485
  • Robust and watertight design
Technical Characteristics
  • Measure Principle: Conductivity sensor with 4 electrodes
  • Conductivity Ranges: 0-200.0 μS/cm, 0-2000 μS/cm, 0.00-20.00 mS/cm, 0.0-200.0 mS/cm
  • Resolution: 0.01 to 1 depending on the range
  • Accuracy: ± 1% of the full range
  • Salinity Range: 5-60 g/Kg
  • TDS Range (KCl): 0-133,000 ppm
  • Response Time: < 5 s
  • Working Temperature: 0°C to +50°C
  • Temperature Compensation: NTC
  • Stocking Temperature: -10°C to +60°C
  • Signal Interface: Modbus RS-485 (standard) and SDI-12 (optional)
  • Maximum Refreshing Time: Max < 1 s
  • Sensor Power Supply: 5 to 12 volts
  • Electric Consumption: Standby: 25 μA, Average RS485: 6.3 mA, Average SDI12: 9.2 mA, Current Pulse: 500 mA
  • Dimensions: Diameter: 27 mm, Length: 157 mm
  • Weight: 300 g (sensor + 3-meter cable)
  • Material: PVC, DELRIN, stainless steel
  • Maximum Pressure: 5 bars
  • Connection: 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector
  • Protection: IP68
Wiring Diagram

For cable lengths up to 15 meters and 15 to 100 meters, the wiring includes power supply, SDI-12, RS-485 connections, and cable shielding.

Contact Information

AQUALABO
Tel: +33 (0)1 55 09 10 10
Fax: +33 (0)1 55 09 10 39
Email: [email protected]
Website: www.aqualabo.fr

See more

Catalog excerpts

C4E: CONDUCTIVITY /SALINITY-1

C4E: CONDUCTIVITY /SALINITY DIGITAL TECHNOLOGY FOR OPTIMIZED MEASURES RANGE DIGISENS / DATASHEET APPLICATIONS • Urban wastewater treatment • Industrial effluent treatment • Surface water monitoring • Drinking water MOUNTING AT 4 ELECTRODES The electrode works with a technology in 4 electrodes: an alternating current of constant-voltage is established between a primary’s pair of electrodes in graphite. The secondary’s electrodes in platinum allow of regulate the voltage imposed to primary’s electrodes to reflect of the fouling. The voltage measured between the primary’s electrodes is in function of the resistance of place and so, of the conductivity. DIGITAL TECHNOLOGY The “smart” Digital C4E sensor stores calibration and history data within the sensor. This allows you a “plug and play” system without re-calibration. Thanks to the Universal Modbus RS485 protocol, the PONSEL Digital C4E can be connected to all devices commonly used (Datalogger, Controller, Automat, Remote System…). • 4 electrodes (2 graphic, 2 platinum) • Range 0 to 200 mS/cm • Digital sensor / Modbus

 Open the catalog to page 1
C4E: CONDUCTIVITY /SALINITY-2

TECHNICAL CHARACTERISTICS Cable lenght 15 to 100 meters © RED, PURPLE, YELLOW, ORANGE, PINK : POWER SUPPLY V+ K BLUE : SDI-12 © BLACK : POWER SUPPLY V-W GREEN : B " RS-485 " © WHITE : A " RS-485 " © GREEN / YELLOW : CABLE SHIELD PONSEL BYAQUALABO