600-1200 PHCI-1

TESTING FITOCLIMA 600 / 1200 PHCI ENVIRONMENTAL TESTING CHAMBERS FOR CURING APPLICATIONS (MORTAR, CEMENT, LIME AND PLASTER). ASTM C511 Moist Cabinet

 Open the catalog to page 1
600-1200 PHCI-2

ARALAB is a company specialized in designing, developing, manufacturing and servicing of high quality climatic chambers and controlled environment rooms. Since 1985 we have been perfecting ways to create and control temperature, humidity, light, air flow and many other environmental conditions. Only the highest quality components are used to manufacture our chambers so customers can have the best equipment for their research and testing purposes. Control the Environment. KEY FEATURES • Ready to use. No assembly needed • Minimal footprint for efficient use of laboratory space • Adaptive future...

 Open the catalog to page 2
600-1200 PHCI-3

• • • • TEMPERATURE & HUMIDITY WORKING RANGE ENVIRONMENTAL TESTING CHAMBERS FOR CURING APPLICATIONS

 Open the catalog to page 3
600-1200 PHCI-4

Fresh air intake (021 mm) Exhaust air and passthrough (0 80 mm) Standard refrigeration system is air cooled 3. 0 Services hub installation needs: 0 3/” male demineralized water supply valve Conductivity: <50pS/cm , TDS <35PPM Pressure: 1-5 Bar In case of reverse osmosis/cylinder (optional): 4. /” male tap water valve connection 5. 0 20mm water drain at floor level female connection Electrical cabinet installation needs: Supply power: 230VAC, 50Hz (60Hz upon request), 12A / Single Phase + Neutral + Ground Electrical protection: Circuit breaker 16A + N with 30mA differential 0 RJ45 communication...

 Open the catalog to page 4
600-1200 PHCI-5

EXTERNAL DIMENSIONS (HxWxD) (mm)INTERNAL DIMENSIONS (HxWxD) (mm) 1330 1266 1. Standard refrigeration system is air cooled 3. (-20°C model is water cooled. See point 5.) 0 Services hub installation needs: 0 %” male demineralized water supply valve Conductivity: <50pS/cm , TDS <35PPM Pressure: 1-5 Bar In case of reverse osmosis/cylinder (optional): 4. %” male tap water valve connection 5. 0 20mm water drain at floor level female connection Electrical cabinet installation needs: Supply power: 230VAC, 50Hz (60Hz upon request), 13A / Single Phase + Neutral + Ground Electrical protection: Circuit breaker...

 Open the catalog to page 5
600-1200 PHCI-6

EQUIPMENT DESCRIPTION CONSTRUCTION & CONTROL Multi-color 7-inch touch-screen ClimaPlus© controller. Open door alarm with configurable time-out function. Highly resistant stainless-steel. Polyurethane insulation. Exterior zinc plated steel and gray epoxy paint. Pivoting door(s) with spring lock, magnetic gasket and safety lock(s). 4 and 5 built-in casters brakes. 80mm Ø side entry port. Maximum load per chamber: 600 model: 160Kg / 1200: 320Kg. ENVIRONMENTAL CONTROL • Air-cooled, mechanical refrigeration by hermetic compressor. • Dual heating technology with hot gas by-pass and stainless steel...

 Open the catalog to page 6
600-1200 PHCI-7

CLIMAPLUS HMI CONTROLLER Programmable PLC exclusively developed for ARALAB chambers. Multi-language, easy-to-use touch-screen interface (EN, FR, DE, IT, PT, ES and other languages on request). Resolution of 0,1ºC for Temperature and 0,1% for Relative Humidity. High performance temperature and humidity control with value correction in all ranges. Capability for creating 32 programs of 24 segments each. Internal non volatile memory for storing test data. Automatic restart of tests due to power failure, without losing data and restarting test where it was interrupted. Real-time monitoring of all...

 Open the catalog to page 7
600-1200 PHCI-8

FITOLOG SOFTWARE The FitoLog software pack is a set of applications designed to facilitate the managing, monitoring and recording of programs and data from the TESTA chambers. It consists of 3 applications: FitoLog, FitoLogView and FitoProgram. FITOLOG Records and displays in real time all data and details related to the set-points, running variables and equipment behaviour. It also retrieves information about the active components of the chamber, running processes, errors, alarms and allows the configuration of periodic or alarm triggered remote notifications (by email or SMS, depending on existing...

 Open the catalog to page 8
600-1200 PHCI-9

ACCESSORIES OPTIONAL ACCESSORIES FitoLog / FitoView software for data monitoring, logging and programs configuration on the PC. Ethernet cable for PC connection. CO2 Controlling and monitoring system. Wall mounting conductivity meter, for water quality control, with assembling accessories. Water Demineralizer. 20 liter water tank with electric pump and security valve. Glass door with double glazing. Additional Stainless steel wire shelves. Reinforced shelves for heavy samples. ACCESSORY FOR ACCELERATED RADIATION AGING 8 fluorescent and UV lamps will decrease climatic control range, when lights...

 Open the catalog to page 9
600-1200 PHCI-10

Let’s meet! [email protected] www.aralab.pt T: +351 219 154 960 Configure your chamber FACEBOOK-SQUARE/AralabChambers Linkedin/company/aralab youtube/user/AralabChambers SQUARE-X-TWITTER/Aralab_ /aralabchambers Control the environment Your own cl