TESTING FITOCLIMA CURING WALK-IN TEMPERATURE AND HUMIDITY CURING TEST CHAMBER.
Open the catalog to page 1FITOCLIMA CURING WALK-IN TECHNICAL SPECIFICATIONS AND PERFORMANCES • • • • FITOCLIMA WALK-IN CURING CHAMBERS TEMPERATURE RANGE [1] [1] Temperature range extension available from 5°C to 45°C. Please contact Aralab for other technical requirements. [3] Aralab Stability rooms are pre-configured with shelves to provide the most efficient use of the available space (5 levels with approximately 470mm or 560mm of depth each, depending on chamber model). Also available stainless steel AISI 304 shelves as option. [4] Additional trays/levels can be added to shelving configuration. • • • • TEMPERATURE &...
Open the catalog to page 2FITOCLIMA CURING WALK-IN EXTERNAL DIMENSIONS (HxWxD) (mm) Electrical cabinet installation needs: 0 Supply power: 400VAC, 50Hz (60Hz upon request), 16A / 3-Phase + Neutral + Ground Electrical protection: Circuit breaker 3 x 16A + N with 300mA differential 3-Phase electrical cable RV-K 5G2.5 on the top 0 RJ45 Ethernet socket communication port on the top 1. Standard refrigeration system is air cooled 3. 2. Services installation needs: 0 V2” demineralized water supply Condutivity: <40pS/cm Pressure: 1-5 Bar 0 20mm water drain at floor level female connection TEMPERATURE AND HUMIDITY CURING TEST...
Open the catalog to page 3FITOCLIMA CURING WALK-IN EXTERNAL DIMENSIONS (HxWxD) (mm) 2 610 x 2 225 x 3 760 (4 350 with door opening) Minimum Clearence Required Standard refrigeration system is air cooled 3. Services installation needs: 0 V2” demineralized water supply Condutivity: <40^S/cm Pressure: 1-5 Bar 0 20mm water drain at floor level female connection Electrical cabinet installation needs: 0 Supply power: 400VAC, 50Hz (60Hz upon request), 16A / 3-Phase + Neutral + Ground Electrical protection: Circuit breaker 3 x 16A + N with 300mA differential 3-Phase electrical cable RV-K 5G2.5 on the top 0 RJ45 Ethernet socket...
Open the catalog to page 4FITOCLIMA CURING WALK-IN CONSTRUCTION Exterior: EN 14509 sandwich type high density injected polyurethane foam modular panels, 80mm thick (other thicknesses available for different specific insulation requirements) with galvanized steel in light gray plastic finish. Floor: slip resistant stainless steel floor. Walls: Galvanized steel with white epoxy paint, optional stainless steel. Shelving and trays: aluminium (optional stainless steel) structures with polypropylene trays (5 racks). New generation multi-color touch-screen ClimaPlus© controller. Door with observation window with key and safety...
Open the catalog to page 5FITOCLIMA CURING WALK-IN I6] • Programmable Logic Controller exclusively developed by Aralab for FitoClima chambers. • Multi language (PT, ES, EN, FR, DE, IT) easy to use touch-screen interface. • 90mm x 155mm (7 inch) multicolor display. • Controls every environmental variable available for any specific FitoClima model (Temperature, Humidity, Lights, Airflow, CO2 and connected external devices). • Friendly program editor for creating 32 programs of 24 segments each, allowing the design of complex and comprehensive climatic simulation programs. • Password protection of the controller functions....
Open the catalog to page 6FITOCLIMA CURING WALK-IN FITOLOG SOFTWARE The FitoLog software pack is a set of applications designed to facilitate the managing, monitoring and recording of programs and data from the FitoClima chambers. It consists of 3 applications: FitoLog, FitoLogView and FitoProgram. FITOLOG Records and displays in real time all data and details related to the set-points, running variables and equipment behaviour. It also retrieves information about the active components of the chamber, running processes, errors, alarms and allows the configuration of periodic or alarm triggered remote notifications (by...
Open the catalog to page 7FITOCLIMA CURING WALK-IN COMMON ACCESSORIES OTHER ACCESSORIES AVAILABLE. PLEASE CONSULT ARALAB FOR OTHER ITEMS FitoLog® software pack for PC/Laptops, enabling data monitoring, logging and managing operations directly on a computer. IQ, OQ, PQ validations and documentation. CO2 Controlling and monitoring system. Wireless / portable data loggers for temperature and humidity control of external areas. Additional shelves and racks. Shelves in AISI 304 stainless steel. Ethernet / Wireless / Other connections for computer connection. Additional entry port. Extended temperature ranges. Integrated water...
Open the catalog to page 8Let’s meet! [email protected] www.aralab.pt T: +351 219 154 960 Configure your chamber FACEBOOK-SQUARE/AralabChambers Linkedin/company/aralab youtube/user/AralabChambers SQUARE-X-TWITTER/Aralab_ /aralabchambers Control the environment Your own cl