TESTA e-TS30055W

TESTA e-TS30055W

TESTA e-TS30055W

Product catalog summary
Equipment Description:
The Testa_e TS3055W is a 3-zone thermal shock chamber designed for testing samples under varying temperature conditions. It consists of a high-temperature room for heat storage, a test room for sample placement, and a low-temperature room for cold energy storage. The chamber includes a general power panel, operation panel, power distribution room, and machine room, with air inlets and outlets for temperature regulation.
Technical Specifications:
  • High Temperature Chamber: Pre-heat range of +60ºC to +200ºC with a heat-up rate of ≤20 minutes.
  • Low Temperature Chamber: Pre-cooling range of -78ºC to 0ºC with a pull-down rate of ≤80 minutes.
  • Volume and Dimensions: Interior volume of 300L, with inside dimensions of W970mm × H460mm × D670mm and outside dimensions of W2030mm × H1935mm × D1845mm. The chamber weighs approximately 1510kg.
  • Test Room: Operates in 2-zone or 3-zone mode with temperature exposure ranges of +60ºC to +150ºC and -55ºC to -10ºC. Temperature fluctuation is 1.0ºC with a deviation of ±2.0ºC.
  • Temperature Recovery: Recovery time is ≤5 minutes under specified conditions.
Construction:
The chamber is constructed with galvanized steel and AISI 304 stainless steel, insulated with polyurethane foam and glass wool. It features a thermal generator system with fans, heaters, and temperature sensors for both high and low-temperature areas. The chamber is equipped with adjustable pins and castors for support and mobility, and a manual sliding door with an unlock button.
Refrigeration System:
The system uses mechanical compression cascade refrigeration with components like a mechanical compressor, plate heat exchanger, and finned tube heat exchanger. It employs refrigerants R449a/R23 and includes a PLC controller with a 10.4-inch display for program management and data recording.
Installation Requirements:
The chamber requires a power supply of 380 V 50Hz with specific water quality and environmental conditions for optimal operation. It should be installed in a well-ventilated area free from strong vibrations, electromagnetic fields, and hazardous gases.
See more

Catalog excerpts

TESTA e-TS30055W-1

TESTING Testa_e TS3055W 3-ZONES THERMAL SHOCK CHAMBER

 Open the catalog to page 1
TESTA e-TS30055W-2

• • • • 3 ZONES TEMPERATURE SHOCK TEST CHAMBER• This system features a three-zone structure enabled by damper switching, allowing samples to remain in place and be tested via cable connection. Easily switch between two-zone and three-zone operation, and expose samples to ambient temperature when needed. Air outlet Machine room High temperature room Test room Power distribution room Operation panel General power panel High temperature room: for the storage of heat. Test room: for the placement of test sample and execution of test. Low temperature room: for the storage of cold energy. General power...

 Open the catalog to page 2
TESTA e-TS30055W-3

• • • • HIGH TEMPERATURE CHAMBER PRE-HEAT TEMPERATURE RANGE +60°C ^ +200oC TEMPERATURE HEAT-UP RATE NoteTemp. heat-up rate is the performance when high temp. +60°C -e +200°C <20min chamber operates independently. • • • • LOW TEMPERATURE CHAMBER PRE-COOLING TEMPERATURE RANGE -78°C ^ 0°C TEMPERATURE PULL-DOWN RATE Note Temperature pull-down rate is the performance of low +20°C -e -70°C <80 min • • • • TEST ROOM (TEST SAMPLE AREA) TEST METHOD 2-zone or 3-zone system by means of damper switching HIGH TEMP EXPOSURE RANGE +60°C~+150°C LOW TEMP EXPOSURE RANGE -55°C ~-10°C • TEMPERATURE RECOVERY PERFORMANCE...

 Open the catalog to page 3
TESTA e-TS30055W-4

CONSTRUCTION• • • • STRUCTURE AND INSULATION OUTER WALL Galvanized steel sheet with plastic-sprayed surface Interior INTERIOR AISI 304 stainless steel plate INSULATION Polyurethane foam + glass wool DOOR INSULATION Glass wool HIGH TEMPERATURE AREA Fan, heater, fan door, temperature sensor LOW TEMPERATURE AREA Fan, heater, evaporator, regenerator, fan door, temperature sensor One set of manual up and down sliding door with unlock button. OBLONG CABLE PORT 25mm x 100mm x1, at the left side of the chamber Controller display, fault indicating lamp, hour meter, USB interface, over- temperature protection...

 Open the catalog to page 4
TESTA e-TS30055W-5

• Nickel chrome electric strip heater. Control method of heater: equivalent periodic pulse-width modulation without contact, SSR (solid-state relay). • Evaporator cooling. • • • • REFRIGERATION SYSTEM WORKING MODE • Mechanical compression cascade refrigeration. • Compressor: Mechanical compressor. • Condenser: Plate heat exchanger, Water Cooling. • Condensing evaporator: Plate heat exchanger. • Evaporator: Finned tube heat exchanger. • Throttling device: Capillary, thermostatic expansion valve. • Refrigerant: R449a/R23 (ozone zero). • Refrigerating machine control method: The control system will...

 Open the catalog to page 5
TESTA e-TS30055W-6

• • • • TEST SAMPLE POWER SUPPLY CONTROL TERMINAL • Relay contact control, AC240V, within 2A (contact is closed in normal operation; contact will be opened when the equipment shuts down or fault occurs). • The use of this terminal for controlling the power on or off is strongly recommended, to ensure the power supply connected the test sample being reliably cut off when the equipment shuts down or fault occurs. • • • • GENERAL POWER SUPPLY LEAKAGE CIRCUIT BREAKER • Power input. • For the leakage circuit and overload circuit. • • • • HOUR METER • 0~99999h zero reset is not available. • • • • SAFETY...

 Open the catalog to page 6
TESTA e-TS30055W-7

araldb 3-ZONES thermal shock chamber

 Open the catalog to page 7
TESTA e-TS30055W-8

INSTALLATION REQUIREMENTS Power supply: 380 V 50Hz 3-phase four wires + protective grounding wire, 63 A (39 kW). Circulating cooling water: • The quality of circulating cooling water: chlorine ion content <300PPm, PH:7.2 to 8.5, total alkali <800mg/L • Water pressure: 0.32 MPaA0.45MPa (The cooling water pipeline system shall be designed and constructed to ensure the pressure at the inlet of refrigerating machine at rated flow is 0.32MPaA0.45MPa, and the pressure falls between the outlet of refrigerating machine and the cooling tower does not exceed 0.05MPa.) Ground flat <5mm/2m Well-ventilated...