Flowkit Support Pack
20Pages

{{requestButtons}}

Catalog excerpts

Flowkit Support Pack - 1

by Armstrong Medical Support Pack Respiratory care, simplified 1

Open the catalog to page 1
Flowkit Support Pack - 3

Support Pack High Flow Oxygen Therapy, Instructions for use p4 AquaNASE®, Fitting guide p5 Universal Face Mask, Fitting guide p8 MAXBIend™, Oxygen sensor calibration procedure pi 0 AquaVENT"-' Heater setup pi 1

Open the catalog to page 3
Flowkit Support Pack - 4

High Flow Oxygen Therapy 1. Select correct size of AquaNASE® prongs. 2. Apply AquaNASE® and secure neck strap as shown. 3. Insert humidification chamber onto heater plate and spike water bag. 4. Attach circuit as shown a-d. 5. Connect heater cable and temperature probes. 6. Turn on heater and select NIV mode. 7. Set flow at 20l/min and select prescribed FiO2 for 2 minutes. 8. Increase flow to 30l/min and monitor PaO2, SpO2 and respiratory rate.

Open the catalog to page 4
Flowkit Support Pack - 5

Fitting guide Step 1 Fit the looped neck support strap by placing over patients head and sliding down around the neck as shown. Step 2 Adjust the looped neck support strap by pulling one strap through the clamp as shown. Step 3 Place the elasticated strap around the patients head. Step 4 Adjust the ends of the elasticated strap to secure the device.

Open the catalog to page 5
Flowkit Support Pack - 6

CPAP Instructions for use 1. Select correct size of face mask using sizing gauge, DO NOT APPLY MASK UNTIL POINT 7. 2. Insert humidification chamber onto heater plate and spike water bag. 3. Attach circuit as shown a-f. 4. Connect heater cable and temperature probes. 5. Turn on heater and select NIV mode. 6. Set flow at 60l/min and select prescribed FiO2. 7. Apply mask as per IFU. 8. Connect breathing circuit to mask. 9. Adjust flow to achieve CPAP, confirmed by manometer reading. Align point with bridge of nose Align point with bridge of nose

Open the catalog to page 6
Flowkit Support Pack - 7

1. Select correct size of face mask using sizing gauge, DO NOT APPLY MASK UNTIL POINT 7. 2. Insert humidification chamber onto heater plate and spike water bag. 3. Attach circuit as shown a-d. 4. Connect heater cable and temperature probes. 5. Turn on heater and select NIV mode. 6. Select mode of ventilation as prescribed. 7. Apply mask as per IFU. 3a Align point with bridge of nose Align point with bridge of nose

Open the catalog to page 7
Flowkit Support Pack - 8

Universal Face Mask Instructions for use Step 1 Use sizing gauge to select appropriate size of mask. Place gauge tip to the bridge of patient’s nose. Align chin to scale to indicate correct size for patient. If in between 2 sizes, use smaller size for a better fit. Step 2 Remove mask from packaging and release the bottom two quick release strap clips. Label should be facing outwards and readable. Step 3 Place patient’s chin in chin cup and gently roll mask up to cover mouth and create a seal over the bridge of the nose. Step 4 Slide headgear over patient’s head and reconnect bottom strap...

Open the catalog to page 8
Flowkit Support Pack - 9

Step 5 Adjust top and bottom velcro straps by pulling slightly to hold the mask in place. Adjust the straps at the crown of the head to create a secure, comfortable fit. Step 6 Mask and headgear should look like this when fitted correctly on patient. Step 7 Mask and headgear should look like this when fitted correctly on patient. Step 8 Universal face mask can be used for CPAP and BPAP therapies. By adding a CO2 exhalation valve the mask can be used for BPAP. Detachable CO2 exhalation connector

Open the catalog to page 9
Flowkit Support Pack - 10

AquaVENT® Heater Setup AquaVENT® heater function indicators Red lights will flash to indicate warning. MODE MODE MUTE MUTE An alarm will sound to Red lightsuser toto alert a will flash a indicate warning.lights flashflash to Red will will to lights fault. Red indicate warning. indicate warning. TEMP PROBE INSP. HEATER WIRE TEMP TEMP PROBE EXP. HEATER WIRE PROBE INSP. HEATER INSP. HEATER WIRE WIRE CHAMBER TEMP MUTE HIGH EXP. HEATER EXP. HEATER WIRE WIRE CHAMBER TEMP LOW CHAMBER TEMP CHAMBER TEMP HIGH HIGH CHAMBER TEMP LOW TEMP PROBE AIRWAY TEMPCHAMBER TEMP LOW HIGH INSP. HEATER WIRE AIRWAY...

Open the catalog to page 10
Flowkit Support Pack - 11

To unlock the front panel To lock the front panel Change between Invasive and Non-invasive modes MODE TEMP PROBE INSP. HEATER WIRE EXP. HEATER WIRE MUTE CHAMBER TEMP HIGH CHAMBER TEMP LOW TEMP PROBE AIRWAY TEMP HIGH INSP. HEATER WIRE AIRWAY TEMP LOW EXP. HEATER WIRE CHAMBER TEMP HIGH SEE MANUAL CHAMBER TEMP LOW AIRWAY TEMP HIGH AquaVENT® AIRWAY TEMP LOW Heater Humidifier AIRWAY TEMP HIGH INSP. HEATER WIRE AIRWAY TEMP LOW PressHEATER WIREhold for EXP. and CHAMBER TEMP HIGH SEE MANUAL 3 secs to unlock for CHAMBER TEMP LOW 60 AIRWAY TEMP HIGH will C37.0 secs. UNL.K AquaVENT® AIRWAY TEMP LOW...

Open the catalog to page 11
Flowkit Support Pack - 12

Power indicator ON/OFF button Mode selection (HFOT) helmet CPAP mode, purple for the CPAP mode and blue for the HFOT mode. After selection of a mode, the screen

Open the catalog to page 12
Flowkit Support Pack - 13

Mode selection (CPAP) helmet CPAP mode, purple for the CPAP mode and blue for the HFOT mode. After selection of a mode, the screen

Open the catalog to page 13
Flowkit Support Pack - 14

MaxBlendTM Oxygen sensor calibration procedure MAXBlendTM should be calibrated before its first clinical use. Thereafter, the manufacturer, Maxtec Inc, recommends weekly calibration. (Based on MAXBlendTM operator manual) • Connect the oxygen supply line (pressure alarm will sound). Verify that the oxygen sensor is located in the sensor port and connected to the sensor cable. Do not connect medical air supply line at this time. • Using the ON/OFF key, ensure MAXBlendTM is powered on. • Rotate the FiO2 control to 100%. Allow two minutes for the reading to stabilise. • Unlock the screen. •...

Open the catalog to page 14
Flowkit Support Pack - 15

(Based on MAXBlendTM operator manual) Alternatively, MAXBlendTM can be quickly calibrated to room air (20.9%) using a shortcut command. This function saves time by setting the calibration valve to 20.9% without scrolling the display. • Suspend the oxygen sensor in room air. • Press the LOCK key to unlock the keypad. • Press and hold down the CAL key press the . When the % sign starts to flash, • The unit will automatically enter the LOCKED condition and return to normal operati

Open the catalog to page 15
Flowkit Support Pack - 16

Protected by BioCote® AquaVENT® Heated Breathing Systems contain BioCote® antimicrobial silver additive to limit the numbers of microbes on the surface of the breathing circuit, protecting it from microbial colonisation. Microscopy showing viable P. aeruginosa cells as greencoloured and dead P. aeruginosa cells as red-coloured. Treated polymer (Containing BioCote®) Untreated polymer (Not containing BioCote®)

Open the catalog to page 16

All Armstrong Medical catalogs and technical brochures

  1. Aqua NASE®

    2 Pages

  2. UltraKEEP

    2 Pages

  3. AquaNASE

    4 Pages

  4. Secura

    2 Pages

  5. AquaVENT

    9 Pages

  6. SAFE-Flow

    4 Pages

  7. Combi-flex®

    12 Pages

  8. AMSORB® PLUS

    20 Pages

  9. CPAP

    16 Pages

  10. ultraclear

    6 Pages

Archived catalogs