Magnet Specifications
Field Strength: 0.32 T +/- 5%
Type: Permanent
Gradient Strength: Max 22 mT/m
Slew Rate: Max 44 mT/m/ms
Gradient Rise Time: Min 0.5 msec from 0 to +/-22mT/m
Patient Handling
Ergonomics: The patient table is rail-mounted and slides smoothly into the magnet.
Net Patient Space: 34 cm
Table Length: 240 cm
Max. Patient Weight: 200 kg
Patient Positioning: Includes LED centering device and Flash Localizer System for real-time imaging.
RF System
Digital Spectrometer: Manages up to 8 receiving channels, expandable to 16, with multi-array receiving coils using MAX technology.
RF Amplifier: Max RF Power 3.0kW
Preamplifiers: Integrated in the magnetic unit
Extremity Coils: Multi Array (MAX) for shoulder, knee, ankle, hand, wrist, elbow, and hip applications.
Spine Coils: MAX L-Spine and C-Spine for lumbar and cervical spine applications.
Optional Coils: TMJ, Knee Large, and GP solenoidal for various clinical applications.
MRI System Technical Specifications
Pulse Sequences: Includes Spin Echo T1 and T2, Double Echo, Inversion Recovery, and more.
RF Shielding: Requires a Faraday cage with minimum attenuation of 80 dB and dimensions of 3m x 3m x 2.31m.
Imaging Techniques
Acquisition Matrix 2D: From 128 x 128 to 512 x 512
Acquisition Matrix 3D: From 128 x 128 x 24 to 256 x 256 x 128
Visualized FOV: Up to 260 mm
Number of Slices: Up to 64
Artifact Suppression: Includes flow compensation, oversampling, ABS, and reduced FOV.
Imaging Modalities: Half Scan and Half Echo.
Siting Requirements
Installation: Can be installed in a 4m x 4m room with a system footprint of 16m2.
Weight: 9200 Kg
Power Requirement: Single phase 230 VAC 50Hz (USA 110 VAC 60Hz)
Power Consumption: 4 KVA (0.8 KVA on stand-by)
Additional Requirements: Customer to provide cabling, interface devices, network connection, telephone, and high-speed Internet for Remote Diagnostics Assistance.
See more

Catalog excerpts

MRJ3300 Technical Specifications-1

MRI System Technical Specifications MrJ 3300 - Magnet Field Strength • Type • Gradient Strength - Max • Slew Rate - Max • Gradient Rise Time - Min 0.32 T +/- 5% Permanent 22 mT/m 44 mT/m/ms 0.5 msec from 0 to +/-22mT/m MrJ 3300 - RF System Digital Spectrometer MrJ 3300 is equipped with an innovative MRI digital spectrometer that can manage up to 8 receiving channels (expandable to 16) MrJ 3300 provides multi-array receiving coils with MAX technology (Multi Array Cross Tuned). RF Amplifier Max. RF Power 3.0kW Preamplifiers integrated in the magnetic unit Extremity Coils Clinical applications • Multi Array (MAX) Shoulder • Multi Array (MAX) Knee - Ankle • Multi Array (MAX) • Flat linear coil Spine Coils • MAX L-Spine large • MAX L-Spine small • MAX C-Spine Optional Coils • TMJ • Knee Large • GP solenoidal MrJ 3300 - Patient Handling Ergonomics The patient table - rail mounted - can slide into the magnet with a smooth and effortless movement. Hand - Wrist - Elbow Hip • Net Patient Space: • Table Length: • Max. Patient Weight: Clinical applications Lumbar Spine Lumbar Spine Cervical Spine Patient positioning • LED centering device • Flash Localizer System (FLS) - images are continuously acquired and displayed in real time Clinical applications Temporo Mandibular Joint Knee G

 Open the catalog to page 1
MRJ3300 Technical Specifications-2

MRI System Technical Specifications MrJ 3300 - Pulse Sequences • Spin Echo T1 e T2 (SE T1, SE T2) • Rapid Imaging Spin Echo/Double Echo (RISE, RIDE) • Short TI Inversion Recovery GFE (STIR GFE) • FAST-RISE echo train length 5, 7, 9 (ETL 5, ETL 7, ETL 9) • Fast Inversion Recovery (FIR) • 3D Spoiled Gradient Field Echo (3D SPGFE) • Fluid Attenuated Inversion Recovery (FLAIR T1) • 3D Time Reversed Gradient Field Echo (3D EMIT) • Fluid Attenuated Inversion Recovery (FLAIR T2) • Short TI Inversion Recovery (STIR) • 3D Gradient Balanced Steady State (3D GBASS) • Echo Reduced Acquisition Spin Echo (ERASE)...

 Open the catalog to page 2

All ASG superconductors catalogs and technical brochures

  1. MRJ3300

    6  Pages

  2. MROpen Evo

    24  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.