Product catalog summary
Overview: The document provides detailed technical specifications and operational guidelines for the STERRAD™ 100NX Sterilization System with ALLClear™ Technology. This system is designed to enhance compliance, reduce workflow interruptions, and minimize human error through its user-friendly design and connectivity features.
Product Descriptions: The STERRAD™ 100NX is available in various configurations, including single and double-door units, with STANDARD, FLEX, and DUO cycles. Each unit includes installation, and additional accessories such as cassettes, disposal boxes, and cycle kits are available.
Technical Specifications: The system operates with a heat generation of 4574 BTU/hour when idle and 6187 BTU/hour at maximum. It requires specific electrical power configurations depending on the region (US/Canada, International, Japan). Installation space requirements include specific clearances from walls and service access space.
Networking and Data Recording: The system supports network connectivity for instrument tracking and offers electronic data storage for up to 200 cycles. It includes an internal printer and supports external USB devices for data management.
Operational Environment: The system operates within a temperature range of 18°C - 35°C and humidity levels of 10% - 85% RH. It uses hydrogen peroxide as a sterilant, delivered in cassettes with automatic expiration detection.
Cycle Specifications: The system offers various sterilization cycles: Standard, Flex, Express, and Duo, each with specific time durations and lumen claims. The sterilization process ensures a Sterility Assurance Level (SAL) of 10-6 through a double-kill cycle.
System Dimensions and Layout: Detailed dimensions and layout configurations are provided for both one-door and two-door systems, including space planning and service clearance requirements.
Compliance and Standards: The system complies with ISO 14937 standards, ensuring safety and efficacy in sterilization processes.
Additional Information: For further details, users are directed to visit the ASP website or contact local ASP representatives. The document emphasizes the importance of referring to the complete instructions for use for proper handling and safety precautions.
See more

Catalog excerpts

STERRAD™ 100NX with ALLClear™ Technology Data Sheet-1

REDUCES WORKFLOW INTERRUPTIONS ENHANCES COMPLIANCE AUTOMATICALLY* DESIGNED WITH SIMPLICITY IN MIND TO REDUCE THE POTENTIAL FOR HUMAN ERROR *STERRAD™ Systems with ALLClear™ Technology have features that may enhance compliance, including greater adherence to device IFUs and improved record handli

 Open the catalog to page 1
STERRAD™ 100NX with ALLClear™ Technology Data Sheet-2

Installation and Electrical Requirements HEAT GENERATION ELECTRICAL POWER SPECIFICATIONS • US/Canada: 208 VAC, 60 Hz, NEMA L21-30 5-wire grounding twist-lock outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors • International: 380-415 VAC, 50/60 Hz, 5-wire grounding outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors • Japan: 200 VAC, 50/60 Hz: 4-wire grounding outlet attached to a dedicated 30-amp, 3-phase delta configuration with separate ground conductors INSTALLATION...

 Open the catalog to page 2
STERRAD™ 100NX with ALLClear™ Technology Data Sheet-3

Cycle Specifications* CYCLES Standard: sterilizes most general surgical instruments Flex: sterilizes up to 2 single-channel flexible endoscopes Express: sterilizes da Vinci® 3-D endoscopes and other delicate instruments without lumens Duo: sterilizes cameras and up to 2 single-channel flexible endoscopes LUMEN CLAIMS Standard: diameter: ≥0.7 mm; length: ≤500 mm Flex: diameter: ≥1 mm; length: ≤850 mm Express: nonlumen sterilization only Duo: diameter: ≥1 mm; length: ≤875 mm Technical Specifications CYCLE TEMPERATURE Hydrogen peroxide STERILANT DELIVERY Delivered in closed system in cassettes with...

 Open the catalog to page 3
STERRAD™ 100NX with ALLClear™ Technology Data Sheet-4

Space Planning Dimensions LEGEND Millimeters [Inches] TOP VIEWS ONE-DOOR CONFIGURATION 865.0 +/- 6.4 [34.1 +/- 0.25] TOP VIEW TWO-DOOR CONFIGURATION

 Open the catalog to page 4
STERRAD™ 100NX with ALLClear™ Technology Data Sheet-5

System Dimensions LEFT SIDE INPUT SIDE LEFT SIDE SYSTEM LAYOUT TWO-DOOR CONFIGURATION SYSTEM LAYOUT ONE-DOOR CONFIGURATION 865.0 +/- 6.4 [34.1 +/- 0.25] Clearance on each side of the system should be ~39.4” For recessed system, clearance should be ~1.8” on each side: (865-775)/2 = 45 mm (~1.8”) 1842 +/- 6.4 [72.5 +/- 0.25] WALL OPENING SERVICE CLEARANCE IN-WALL INSTALLATION INPUT OR OUTPUT SIDE

 Open the catalog to page 5
STERRAD™ 100NX with ALLClear™ Technology Data Sheet-6

asp.com ASP International GmbH, Zug Branch Bahnhofstrasse 2, Zug 6300, Switzerland ©ASP 2021. All Rights Reserved. Capitalized product names and ALLClear™ are trademarks of ASP™ Global Manufacturing, GmbH. Important information: Prior to use, refer to the complete instructions for use supplied with the device(s) for proper use, indications, contraindications, warnings and precautions. ADVANCED STERILIZATION PRODUCTS, INC. 33 Technology Drive, Irvine CA 92618, USA EC REP ASP, The Netherlands BV BIC 1, 5657 BX Eindhoven, The Netherlands