Overview: The document details the specifications and features of Laminar Air Flow Ceilings designed for operating rooms to prevent contamination and ensure a sterile environment. These ceilings are crucial in minimizing cross-contamination and nosocomial infections during invasive procedures.
Specifications: The ceilings are compliant with EN ISO 14644-1 and NFS 90 531 standards, achieving ISO 5 particulate classification. They are compatible with CLINICAIR® Air Handling Units and other manufacturers' units. The system includes a user-friendly touch screen interface available in multiple languages.
Filtration and Protection: Equipped with HEPA H14 filters, the ceilings provide fine particulate filtration, eliminating viruses, bacteria, and microorganisms. This ensures a high level of protection in hyper aseptic zones.
Design and Adaptability: Available in various sizes and shapes, the ceilings can be customized to fit specific room configurations. They feature integrated LED lighting and are available in steel or stainless steel versions.
Installation and Maintenance: The design facilitates easy installation and filter replacement. The system includes options for stretch fabric or perforated metal sheet diffusion to ensure uniform air distribution.
Operating Principle: The laminar flow creates a controlled, steady airflow that acts as a protective curtain, removing particles from the operating area. The system includes a 7” touch screen for monitoring temperature, hygrometry, room pressure, and filter status.
Performance: The airflow speed can be adjusted according to different standards and country requirements, with flow rates ranging from 1341 to 13590 m3/h.
Contact Information: ATA Medical, 16 rue Jules Verne, 44700 Orvault, France. Phone: +33 (0) 2 40 92 03 00, Email: [email protected], Website: www.ata-medical.com
See more

Catalog excerpts

LAMINAR AIRFLOW CEILING-1

The range of Laminar Air Flow Ceilings provide effective protection against contamination and are dedicated to operating rooms. They create a clean area around the patient, medical personnel and equipment and provide effective protection against aerobiocontamination, possible during invasive acts. They thus fight cross-contamination and nosocomial infection. Suitable for use in hyper aseptic zone, in accordance with EN ISO 14644-1/ NFS 90 531 to achieve an an ISO 5 particulates classification. They are mainly used with CLINICAIR® Air Handling Unit but can also be applied with other manufacturers' unit. EASY INSTALLATION Installation and replacement of filters facilitated Simple and intuitive use of a control touch screen with interface languages in French, English and Spanish (only with the CLINICAIR range) Fine particulate filtration according to ISO 5 thanks to HEPA H14 FIlters (Supplied with or without filter) SURGICAL LIGHTING Central passage for surgical lighting (depending on range) Steel or stainless steel version Elimination of viruses, bacteria and micro-organisms ADAPTABILITY Available in different sizes in rectangular and square shape. standard size up to 13m² or customized to fit the configuration of each room Integrated white or color light - LED ribbon installed on rails (only in stretch fa

 Open the catalog to page 1
LAMINAR AIRFLOW CEILING-2

OPERATING PRINCIPLE LAMINAR FLOW A laminar flow is a controlled airflow, with a steady and low velocity resulting in a parallel airflow acting like a protection curtain and removing the particles from the operating area (source: NF S 90-351 April 2013) OPTION • LED lighting Integrated white or color light - LED ribbon installed on rails up to 3600 lumens/m for 24 W/M. • Clean room or hallway touch screen and allows to control: Fresh air - Temperature / Hygrometry - Room pressure, - Clogging of filters - Setpoint deviation +/-3°C, - Mode of operation of CLINICAIR, - Secure remote access - Real-time...

 Open the catalog to page 2

All ATA MEDICAL catalogs and technical brochures

  1. LAMINAR FLOW

    2  Pages

  2. CLINICAIR 3

    2  Pages

  3. Multizone 800

    2  Pages

  4. DOPAIR 1000

    2  Pages

  5. Dopair 2000

    2  Pages

  6. Dopair 3000

    2  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.