SANITIZATION DEVICES
16Pages

{{requestButtons}}

Catalog excerpts

SANITIZATION DEVICES - 3

THE COMPANY Founded in 1994, ATENA LUX is a dynamic and young reality recognized in Italy and abroad as a manufacturer of lighting systems for the technical and medical sectors. The wide range of lighting fixtures, which are manufactured and manufactured within the production department located in Gruaro (VE), are synonymous with guarantee, comfort and safety. In fact, Atena Lux stands out for the production of bed head beams, certi fi ed as electro-medical devices, and lighting fixtures for operating theaters and clean rooms. In recent months, the health situation has prompted the company...

Open the catalog to page 3
SANITIZATION DEVICES - 4

UV-C TECHNOLOGY Germicidal Lamps UV-C ultraviolet rays are electromagnetic waves with germicidal properties. Their wavelength ranges from 100 to 280 nm and their maximum effectiveness occurs at the wavelength of 265 nm. UV-C rays have a destructive effect on RNA and DNA of bacteria, mold, yeasts and viruses, preventing their proliferation. Germicidal lamps emit light at 254 nm, approximately 85% of maximum efficiency and are ozone-free. What are the advantages of using UV-C germicidal lamps? Ultraviolet (UV) purification is a very effective method to clean the environment of biological...

Open the catalog to page 4
SANITIZATION DEVICES - 5

How germicidal UV-C lamps work? Precautions for the use of germicidal lamps UV-C radiation has a photolytic effect on DNA and RNA, thus preventing microorganisms such as bacteria, mold, yeasts and viruses from reproducing. The purification effect is obtained with wavelengths less than 320 nm, with maximum efficiency at 260nm. For the elimination of microorganisms with UV-C rays, they must be on the surface of an object or transported by air. Each microorganism needs a different UV-C dose for its inactivation or elimination. Among the various universally recognized documents, the one...

Open the catalog to page 5
SANITIZATION DEVICES - 6

PURIFICATION CYCLE MOTION DETECTOR Ignition at 1 minute from START If you enter the room, the UV cycle will be paused Sheet-steel body powder coated by means of opaque white (RAL 9005) polyester powders, UV resistant CEL version with electronic ballast EEI=A2, 220-240V 50-60Hz, heat ignition. Sensore di presenza e temporizzatore inclusi Luminous source Luminous source FC-L compact lamp, UV-C lamp with wavelength 254 nm, lamp coupling 2G1 Life time of UV-C sources Irradiation time * Appliance compliant with Operating temperature electrical safety electromagnetic compatibility degree of...

Open the catalog to page 6
SANITIZATION DEVICES - 7

PURIFICATION CYCLE MOTION DETECTOR Asymmetrical recuperator in satin aluminium Sheet-steel body powder coated by means of opaque white(RAL 9010) polyester powders, UV resistant. Installation through ABS spacers which are supplied with the luminaire. CEL version with electronic ballast EEI=A2, 220-240V 50-60Hz, heat ignition Motion detector with timer Luminous source FL T26 linear UV-C lamp with wavelength 254 nm, lamp coupling G13. 1x30W UV-C Rendering Dimensions [mm] UV-C Irradiation Operating temperature Irradiation time * Appliance compliant with electrical safety electromagnetic...

Open the catalog to page 7
SANITIZATION DEVICES - 8

PURIFICATION CYCLE MOTION DETECTOR Symmetrical recuperator in satin aluminium Sheet-steel body powder coated by means of opaque white(RAL 9010) polyester powders, UV resistant. Installation through ABS spacers which are supplied with the luminaire. CEL version with electronic ballast EEI=A2, 220-240V 50-60Hz, heat ignition. Motion detector with timer Luminous source FL T26 linear UV-C lamp with wavelength 254 nm, lamp coupling G13 1x30W UV-C Rendering Dimensions [mm] UV-C Irradiation Irradiation time * Appliance compliant with Operating temperature electrical safety electromagnetic...

Open the catalog to page 8
SANITIZATION DEVICES - 9

How to calculate the irradiation time: The amount of UV-C radiation needed to destroy microorganisms is given by a mathematical formula: UV DOSE (J / m²) = Exposure time x UV-C irradiation Using the data reported by the CIE (Internation Commission on Illumination) body on the effect of UV-C radiation in microorganisms, it is possible to obtain what is the necessary dose of UV-C to eradicate 99% of the most common viruses, bacteria, molds and spores. A recent research, led by a team of scientists from the University of Milan, in collaboration with experts from the National Institute of...

Open the catalog to page 9
SANITIZATION DEVICES - 10

PRACTICAL EXAMPLES Vision UV For a 18 m² room we use VISION UV 1X55W, installing it at a height of 2.5 m. To destroy 99% of SARS-COV-2 (169 J / m²) we have to let it run 1207,14 seconds (about 20 minutes). 169 (J / m²) / 0,14 = 1207,14s = 20mins For particularly resistant organisms (see table) it is sufficient to increase the exposure time, for example after about 28 minutes even the most resistant spores will be knocked down. To identify the number of appliances required for air treatment in relation to the size of the premises in m², consult the following tables:

Open the catalog to page 10
SANITIZATION DEVICES - 11

Moovi UV For a 25m² room we use MOOVI UV 2X55W, installed it in the center of the room. At a distance of 2,5 m, to destroy 99% of SARS-COV-2 (169 J / m²) we must leave it running. 169 (J / m²) / 0,35 = 482,85s = 8,04mins For particularly resistant organisms (see attached table) it is sufficient to increase the exposure time, for example after about 12 minutes even the most resistant spores will be knocked down. Time needed to break down 99% of the most resistant microorganisms: Impact UV For an 18 m² room we use IMPACT UV 1X36W, installing it on the ceiling at a maximum height of 3 m. To...

Open the catalog to page 11
SANITIZATION DEVICES - 12

SANITIZING LIGHT Sanitation according to nature The air in nature is purified from mold, bacteria and viruses thanks to the active action of the ions generated by the sun, wind and the motion of the seas. On average, in the mountains we find a concentration of 15,000 ions / cm3, at the sea 50,000 ions / cm3, while in homes the value is drastically reduced to less than 100 ions / cm3. Thanks to the introduction of bipolar ion generation technology, Atena Lux combines the lighting experience with the sanitization of environments, creating a new product concept, created to be installed in...

Open the catalog to page 12
SANITIZATION DEVICES - 13

Ottica composta da lenti modulari ad emissione luminosa controllata Type of installation Body RECESSED CEILING or SUSPENDED (fixing bracket not included) Steel sheet (6/10) painted with matt white polyester powders (RAL 9003 matt) Fixed output electronic LED driver (ON / OFF), included LED DALI standard digital dimmable electronic driver, included. Electronic ballast EEI = A2, 220-240V ; 50-60Hz, hot start Double sanitizing technology: UV-C RAYS for the elimination of viruses and bacteria BIPOLAR ION GENERATOR for odor control and the elimination of VOCs Luminous source Color temperature...

Open the catalog to page 13

All ATENA LUX catalogs and technical brochures

  1. HEALTH CARE

    53 Pages

  2. AIRWALL

    13 Pages

  3. TAURUS

    12 Pages

  4. Armonia

    3 Pages

  5. SENSE

    2 Pages

  6. AUXILIA

    3 Pages

  7. TRIAL

    3 Pages

Archived catalogs

  1. HEALTHCARE

    94 Pages

  2. PLAY/N

    2 Pages

  3. PLAY

    4 Pages