1. Catalogs
  2. ATS Scientific Products
  3. General Purpose Freeze Dryer

General Purpose Freeze Dryer

General Purpose Freeze Dryer

General Purpose Freeze Dryer

Product catalog summary
General Purpose Freeze Dryer Specifications
Key Features:
  • Compact, freestanding, and mobile design.
  • Clear acrylic doors for product and condenser chambers.
  • Optional programmable heat for chamber temperature control up to 60°C.
  • Intellitronics™ controller with manual and automatic drying capabilities.
  • Smooth-wall condenser and 304 stainless steel trays with open mesh.
Electrical Requirements:
  • Voltage: 208/230 VAC 1 Φ 60 Hz, 220 1 Φ 50Hz, 230/240 VAC 1 Φ 50 Hz.
  • Breaker Amperage: 30 A for 25L ES, 40 A for 35L EL.
  • Plug Type: NEMA L6-30-P for 25L ES, NEMA 6-50-P for 35L EL.
Specifications:
  • Chamber Temperature Control Range: -29°C to 0°C (without heat), -29°C to 60°C (with heat).
  • Lowest Condenser Temperature: -50°C / -53°C for 25L ES, -82°C / -85°C for 35L EL.
  • Maximum Condenser Capacity: 25 L for 25L ES, 35 L for 35L EL.
  • Maximum Ice Condensing Capacity in 24 Hours: 12 L for 25L ES, 20 L for 35L EL.
  • Vacuum Time to 100 mTorr: ≤ 30 minutes.
Utility Requirements:
  • Ambient Room Temperature: 20°C.
  • Approx. Peak Heat Generated: 7,200 BTU/h (without heat) for 25L ES, 9,800 BTU/h for 35L EL.
Refrigerant Information:
  • Refrigerants: MO89, R404A, R407C, R508B.
  • Safety Class: A1.
  • Total CO2e: 4.481 t for 25L ES, 7.832 t for 35L EL.
Dimensional Data:
  • Width: 34 in (86 cm), Depth: 55 in (140 cm), Height: 79 in (201 cm).
  • Approximate Weight: 950 lb (432 kg).
  • Minimum Clearance: 10 in (25.4 cm) on all sides.
Additional Information:
  • Construction: 316L stainless steel product and condenser chambers.
  • Vacuum Pump: Two-stage rotary vane.
  • Defrost Type: Hot gas.
  • Refrigerant Type: CFC-free.
See more

Catalog excerpts

General Purpose Freeze Dryer-1

General Purpose Freeze Dryer control range (without heat) control range (with heat) Maximum ice condensing capacity 12 L 20 L Maximum deposition rate* 0.50 L/hour 0.83 L/hour Vacuum time to 100 mTorr§ < 30 minutes < 30 minutes Vacuum rate of rise§ < 30 mT/hour < 30 mT/hour Lowest system vacuum§ < 50 mT < 15 mT Note: Performance specifications are based on SP test data from units operating at an ambient room temperature of approximately 20°C. SP recommends an operating range of 15-25°C (59-77°F). Utility Requirements Ambient room temperature 20°C 20°C Approx. peak heat generated 7,200 BTU/h 9,800 BTU/h Approx. peak heat generated 8,900 BTU/h 11,500 BTU/h • Compact, freestanding and mobile design • Clear acrylic product and condenser chamber doors • Optional programmable heat allows for chamber temperature control to 60°C • Intellitronics™ controller with manual and automatic drying capabilities • Smooth-wall condenser • 304 Stainless steel trays with open mesh Electrical Requirements Refrigerant Information Breaker amperage* Plug type (60 Hz Systems) Recommended receptacle (60 Hz Systems) EPA SNAP — IPR/VLTR IPR/VLTR IPR/VLTR IPR/VLTR Note: The refrigerants and insulating foam contain fluorinated greenhouse gases. For more information please visit scientificproducts.com US 800-431-8232 | International +1 (845) 255-5000 | [email protected] | sp-scientificpr

 Open the catalog to page 1
General Purpose Freeze Dryer-2

GPFD 24Dx48 General Purpose Freeze Dryer Front View (With Programmable Heat) Front View (Without Programmable Heat) Side View Dimensional Data Additional Information Width 34 in (86 cm) Minimum clearance on all sides 10 in (25.4 cm) Construction 316L stainless steel product and condenser chambers Vacuum pump§ Two-stage rotary vane Refrigerant type CFC-free * Voltage supply should be ±5% of the specified nominal voltage. SP VirTis units are highly customizable and SP can configure any unit to conform to the service requirements of a wide range of international voltage and phase configurations....

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.