1. Catalogs
  2. ATS Scientific Products
  3. Ultra Prime Pilot Freeze Dryer

Ultra Prime Pilot Freeze Dryer

Ultra Prime Pilot Freeze Dryer

Ultra Prime Pilot Freeze Dryer

Product catalog summary
Specifications Overview
The Ultra Prime Pilot Freeze Dryer is available in both air-cooled and water-cooled configurations. It requires compressed air at 80 psig and inert gas for backfilling at 1 PSIG. The ambient room temperature for optimal operation is between 15-25°C. The water-cooled version requires cooling water at 15-22 Lpm and 5-25°C.

Key Features
  • Compact design for easy installation.
  • Improved refrigeration for faster condenser pull-down.
  • Single product chamber design for larger batches and uniformity.
  • Available with Merlin or LyoLogic control systems.
  • Optional hydraulic stoppering system.
  • Narrow footprint with cleanroom configuration available.
  • Uses refrigerant with lower Global Warming Potential (GWP).

Performance Specifications
  • Lowest shelf temperature: ≤ -67°C / -70°C.
  • Shelf temperature control range: -55 to 65°C with a tolerance of ±0.5°C.
  • Condenser temperature: ≤ -82°C / -85°C.
  • Maximum condenser capacity: ≥ 50 L.
  • Ice condensing capacity in 24 hours: ≥ 20 L.
  • Vacuum time to 100 mTorr: ≤ 45 minutes.

Utility Requirements
  • Voltage options include 208 VAC and 230 VAC for single-phase, and 400 VAC for three-phase systems.
  • Construction materials include 316L stainless steel for shelves and chambers.
  • Defrost type is hot gas, and the refrigerant is CFC-free.

Dimensional Data
  • Standard configuration width: 64 cm; cleanroom configuration width: 89 cm.
  • Depth and height are consistent at 198 cm and 192 cm, respectively.
  • Maximum weight is 909 kg, with crated weight estimated at 1,089 kg.

Shelf Configuration
  • Available in configurations from 7 to 15 shelves, with varying shelf areas and clearances.

Additional Information
  • Performance specifications are based on SP test data at approximately 20°C ambient temperature.
  • Clearance requirements increase if machines are placed side by side.
See more

Catalog excerpts

Ultra Prime Pilot Freeze Dryer-1

SCIENTIFIC PRODUCTS VIRTIS Ultra Prime Pilot Freeze DryerKey Features• Compact design for easy installation • Improved refrigeration for faster condenser pull-down and overall performance • Single product chamber design allows for larger batches and product uniformity • Available with a Merlin or LyoLogic control system • Optional hydraulic stoppering system available • Narrow footprint (cleanroom configuration available) • Refrigerant with lower GWP Specifications note: Note: Performance specifications are based on SP test data from units operating at an ambient room temperature of approximately 20°C. SP recommends an optimum operating range of 15-25°C (59-77°F). SCIENTIFIC PRODUCTS

 Open the catalog to page 1
Ultra Prime Pilot Freeze Dryer-2

## ATS SCIENTIFIC PRODUCTS VIRTIS Electrical Requirements Shelf Configuration Voltage Phase Hertz Breaker Recommended Shelf Area Shelf Clearance Refrigerant Information Shelf size (W x D): Bulk and Stoppering Moveable Bulk/ Stoppering Note: If machines are placed side by side, increase the minimum clearance to 121.9 cm (48 inches).Additional Information * Shelf fluid inlet temperature controlled to within ± 0.5°C of the setpoint within the Shelf Temperature Control Range when at 100 mTorr. t Shelf Pull-Down times are based on units with one (1) to eight (8) shelves. The increased mass of stainless...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.