LabPad-Evolution

LabPad-Evolution
1 / 88 PagesView full catalog

LabPad-Evolution

Product catalog summary
Introduction
  • Intended Use: The LabPad® Evolution is a Point of Care device for healthcare professionals to conduct rapid In Vitro Diagnostic tests using SmartChip technology.
  • Contents: The package includes the LabPad® Evolution reader, charger, carrying case, and user manual.
  • Operating Conditions: The device should be used in well-lit areas with temperatures between 15-32°C, following safety precautions.
  • Internal Quality Controls: The device performs automatic quality checks and displays error messages if issues are detected.
Overview
  • Reader: Portable and autonomous with multi-measurement capabilities and connectivity options.
  • User Interface: Features a 3-button navigation system and a color screen for status and settings.
  • Power Supply: Operates on rechargeable batteries and can be charged via USB.
  • SmartChip Compatibility: Compatible with Tsmart® and Ksmart® SmartChips for specific tests.
Starting Up
  • First Use: Initialization involves setting language, date, and time. Ensure the device is charged before use.
  • Switching On/Off: Use the middle button for power operations. The device has an auto-off feature.
  • Reinitialization: Restores default settings and erases stored results.
Settings
  • Menu Access: Access settings by holding the validation button. Adjust language, date/time, brightness, and more.
  • SmartChip Configuration: Configure specific tests and measurement units.
  • Memory Management: Default is automatic; manual management allows selective result deletion.
  • Accessories: Pair with Bluetooth devices like barcode readers and printers.
Workflow Integration
  • Connections: Supports USB and Bluetooth for integration with information systems.
SmartChip Rapid Test Performance
  • Procedure: Includes patient identification, test sampling, reading, and result display.
Retrieving Saved Results
  • Browsing and Display: Navigate through saved results and select for display.
Quality Control
  • Liquid Quality Control (LQC): Instructions for performing and storing quality control tests.
Error Messages and Troubleshooting
  • Error Handling: Lists common error messages and troubleshooting steps.
Software Updates and Online Support
  • Updates: Instructions for updating software and accessing online support.
Cleaning and Warranty
  • Maintenance: Guidelines for cleaning the device and warranty information.
Technical Specifications
  • Details: Lists technical specifications and symbols used in the manual.
Overview:
The LabPad Evolution manual provides detailed instructions for integrating the device with information systems, performing tests, and managing results. It covers workflow integration, connection methods, test procedures, and error handling.

1. Workflow Integration:
The LabPad Evolution can integrate with information systems via USB or Bluetooth to manage operator and patient identification, quality control, and results validation. Upon first connection, the device inherits configurations from the system, which may include test permissions and data recording settings. Configuration changes are irreversible without reinitialization.

2. Connection Methods:
USB Connection: Used for connecting to a PC service. Ensure the service is running before connecting.
Bluetooth Connection: Used for connecting to smartphones or tablets. Follow the pairing process using a PIN code.

3. SmartChip Rapid Test Performance:
Preparation: Ensure the device is on a stable surface and all materials are ready, including compatible SmartChips and sampling tools.
Patient Identification: Required before testing, using a barcode reader or connected device.
Test Execution: Insert the SmartChip, perform blood collection, and apply the sample correctly to avoid errors.

4. Ksmart Test Sampling and Reading:
Follow specific instructions for sample collection and insertion timing to ensure accurate results.

5. Result Display and Storage:
Results are displayed and stored automatically. Up to 1000 results can be stored, and they can be browsed and selected for detailed viewing.

6. Quality Control:
Quality control tests can be performed using Liquid Quality Controls specific to each SmartChip. Results are stored similarly to routine tests.

7. Error Messages:
Error messages are displayed with a code and possible solutions are provided in the manual.
Error Messages and Solutions
  • Low Battery: Insufficient charge to perform a test. Solution: Eject SmartChip, plug in power adapter, and reinsert SmartChip within 1 minute.
  • Memory Full: Not enough space to save results. Solution: Eject SmartChip, erase results, or set memory to automatic.
  • Insufficient Ambient Brightness: Not enough light for the test. Solution: Ensure sufficient lighting and reinsert SmartChip.
  • Data Reading Error: SmartChip not inserted properly or damaged. Solution: Reinsert SmartChip or use another one.
  • Inadequate Ambient Temperature: Temperature too high or low. Solution: Adjust environment temperature and retry with another SmartChip.
  • Expired SmartChip: SmartChip has expired. Solution: Use an unexpired SmartChip.
  • SmartChip Not Inserted Properly: Solution: Reinsert SmartChip correctly.
  • Error During Autotests: Solution: Restart device and try again.
  • Defective SmartChip: Solution: Use another SmartChip.
Troubleshooting
  • Device Not Working: Charge battery for at least 4 hours.
  • Battery Cannot Be Charged: Contact reseller.
  • Device Frozen: Hold power button for over 20 seconds.
Software Updates and Online Support
Software updates are provided regularly. Connect LabPad® Evolution to a PC for updates via a link provided by the reseller.
Cleaning Instructions
Clean with soapy water or ethanol using a lint-free cloth. Avoid liquid in the SmartChip area.
Warranty Information
Device is under warranty. Contact reseller for defects.
Technical Specifications
  • Operating Temperature: 15-32°C
  • Storage Temperature: -20 to +70°C
  • Memory: 1,000 measurements
  • Battery: Lithium-ion Polymer 3.7V 2100mAh
Icons and Symbols
Includes navigation buttons, battery indicators, and connection indicators.
See more

Catalog excerpts

LabPad-Evolution-3

Connected Biology Everywhere Guide d’utilisation LabPad Evolution

 Open the catalog to page 3
LabPad-Evolution-4

1.1/ Intended use 1.2/ Contents of the packaging 1.3/ Operating conditions and precautions of use 1.4/ Internal Quality Controls 2.1/ Reader 2.2/ User interface 2.3/ Color Screen 2.4/ Power Supply 2.5/ Compatibles SmartChip : Tsmart® et Ksmart® 2.6/ SmartChip Insertion 2.7/ Contact-free SmartChip ejection 4.1/ Settings menu 4.2/ Adjustable parameters and functionalities of the settings menu 4.3/ Configuration of compatible SmartChip 4.4/ Addition regarding memory management 4.5/ Accessories 3.1/ First use and initialization 3.2/ Switching on 3.3/ Switching off 3.4/ Reinitialization User manual...

 Open the catalog to page 4
LabPad-Evolution-5

5.1/ Principle 5.2/ USB connection 5.3/ Bluetooth connection 6/ SmartChip rapid test performance 6.1/ Getting ready 6.2/ Patient identification 6.3/ Tsmart® test sampling and reading 6.4/ Ksmart® test sampling and reading 6.5/ Result display and SmartChip ejection 7/ Retrieving saved results 7.1/ Browsing the list of results 7.2/ Selecting one result for display 8.1/ Usage of a Liquid Quality Control (LQC) 8.2/ How to start a quality control test 8.3/ Performing a quality control test: example of the INR 8.4/ Storage of quality control results 9/ Error messages 10/ Troubleshouting 11/ Software...

 Open the catalog to page 5
LabPad-Evolution-6

Caution This user guide provides all required information to operate optimally the LabPad®Evolution. Please read it carefully before using the reader for the first time. User manual LabPad Evolution

 Open the catalog to page 6
LabPad-Evolution-8

1.1/ Intended use LabPad® Evolution is a Point of Care (PoC) multi-measurements testing device, intended for healthcare professionals use only, in order to perform In Vitro Diagnostic (IVD) rapid tests from a small volume of biologic sample. This device must be operated only with a range of single use compatible tests called SmartChip. Each SmartChip in the panel allows for the test of one specific biologic measurement. Depending on the technology, there are two different formats of SmartChip respectively called Tsmart® and Ksmart® due to their geometrical shape. LabPad® Evolution can be used...

 Open the catalog to page 8
LabPad-Evolution-9

1.2/ Contents of the packaging In the box you will find: 3 The LabPad® Evolution reader A Charger with a main unit and a micro USB-B cable Check that all these elements are included; if not, contact your reseller. User manual LabPad Evolution Introduction

 Open the catalog to page 9
LabPad-Evolution-10

1.3/ Operating conditions and precautions of use CAUTION Operating conditions and precautions for safety use To ensure the proper operation of the device as well as your own safety, please follow the below instructions: • Use your LabPad® Evolution in a location with sufficient lighting and with ambient temperature between 15 and 32°C (59° to 90°F); • To perform a test, place your device on a stable, flat, and vibration-free surface; • Do not insert any object other than a SmartChip in the insertion area, including during cleaning; • Use only the side black button to eject a SmartChip; • Do not...

 Open the catalog to page 10
LabPad-Evolution-11

1.4/ Internal Quality Controls LabPad® Evolution is a technologically advanced device performing automatically number of internal quality controls prior, during and after each measurement. If a problem subject to interfere with the measurement is detected, an error message such as “Error XX”, where XX is the error number, is displayed on a red background on the screen of the reader (See Section 9 « Error Messages»). As example, the following error screen is displayed when inserting an expired SmartChip: The expiry date is one of information encoded in the Datamatrix of the SmartShip. CAUTION...

 Open the catalog to page 11
LabPad-Evolution-13

2.1/ Reader The LabPad® Evolution is an In Vitro Diagnostic (IVD) device allowing for the measurement of different biologic tests with the same reader. Its design allows for: • Portability (270g weight) • Autonomy (powered by chargeable batteries) • Easy to use (3 buttons interface) • Multi measurements • Connectivity (Bluetooth Low Energy and USB) User manual LabPad Evolution Over

 Open the catalog to page 13
LabPad-Evolution-14

2.2/ User interface 3 buttons allow to navigate the menus and make selections: Use triangular buttons to navigate into the menus. Use middle power button to select. Pressing the middle power button for 2 seconds allows to switch on or off the device and to access to the settings menu (See Section 3 «Starting Up»). 2.3/ Color Screen Status bar User manual LabPad Evolution Overview Battery Charge

 Open the catalog to page 14
LabPad-Evolution-15

Central part Contextual information area Display area Message area Current date 2.4/ Power Supply The LabPad® Evolution is powered by a rechargeable battery and can be operated from an electrical power supply when charging. To charge the battery, connect the micro USB port located on the lower part of the device to the power adaptor using the provided USB cable, then the charger power adapter to a power supply. The provided USB cable complies with CEI 61010 standard. User manual LabPad Evolution Overview

 Open the catalog to page 15
LabPad-Evolution-16

The socket outlet must be installed near the equipment and easily accessible. In order to check the battery indicator when charging, press the middle power button. Battery indicator According to the battery level, the indicator on the upper right of the screen is fully white when the battery is full, and respectively, 75%, half or 25% white. The indicator is red when the reader does not have enough battery power to perform a test. Quickly plug it in power supply. User manual LabPad Evolution Overview

 Open the catalog to page 16
LabPad-Evolution-17

The icon with lightening symbol means the battery is charging. Whatever the battery charge, it is always possible to perform an INR test by plugging the device into a power supply. It is advised not to store the device when the battery is low. When the battery is totally empty, saved data are not lost but the device must be configured again (see Section 3.1 First use and initialization). 2.5/ Compatible SmartChip Tsmart® et Ksmart® The LabPad® Evolution operates with a range of compatible single-use disposable SmartChip tests. Each SmartChip allows for the performance of a specific biology test....

 Open the catalog to page 17