CO2 INCUBATORS
20Pages

{{requestButtons}}

Catalog excerpts

CO2 INCUBATORS - 1

CO 2 INCUBATORS The Safer Choice for Your Laboratory NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A. | nuaire.com

Open the catalog to page 1
CO2 INCUBATORS - 2

An accessory port is standard on all models providing power to accessories such as bottle rollers and flask shakers. 2 The inner glass door features a cam action door handle to create a tight seal against the gasket eliminating potential leaks saving gas and lowering operating costs. The gasket flange on the inner door faces towards the laboratory environment keeping unwanted growth isolated from the growth chamber. The gasket may be removed and autoclaved separately if so desired.

Open the catalog to page 2
CO2 INCUBATORS - 3

About NuAire. For more than 45 years, NuAire has been committed to bringing you the highest-quality, most dependable laboratory products on the market. We are universally recognized as one of the world’s leading providers of reliable equipment for the most demanding environments, including Biological Safety Cabinets, CO2 Incubators,Laminar Air Flow workstations, Ultra Low Temperature Freezers, Centrifuges,Animal Transfer Stations, Pharmacy Compounding Isolators, Polypropylene Fume Hoods, Polypropylene Casework, and a variety of complementary products and systems. You can depend on our...

Open the catalog to page 3
CO2 INCUBATORS - 5

NuTouch Electronic Control System The NuTouch Electronic Control System (ECS) is a color 127 x 178 mm (5 x 7-inch) touchscreen designed for high user adoption. Designed after smart-phone technology, NuTouch creates a unique product experience providing an easy method to control, monitor, and notify. Help screens offer a step-by-step guide to assist in procedures such as running a sterilization cycle. User activated help pop-up descriptions are available if an item or icon is unclear. For example, pressing the text “CO2” on the main screen provides the description “This control setting...

Open the catalog to page 5
CO2 INCUBATORS - 6

Rounded interior corners in the growth chamber promote more thorough cleaning by eliminating tight crevasses found in 90-degree corners.

Open the catalog to page 6
CO2 INCUBATORS - 7

Constant Contamination Control Contamination-Free Design Construction The growth chamber is constructed of 16 gauge, type 304L polished stainless steel using a crevice-free design. The growth chamber walls are completely smooth with rounded corners which allow more complete contact between the chamber surface and cleaning solutions than a typical 90-degree corner. Shelving, supports and guide rails are easily removable and can be autoclaved or left in the chamber during a sterilization cycle. Dual Sterilization Cycles In-VitroCell Direct Heat CO2 Incubators offer a 145°C high-temperature...

Open the catalog to page 7
CO2 INCUBATORS - 8

Closed Loop HEPA Filtration Keep Contamination Out In-VitroCell CO2 incubators are designed with characteristics similar to an ISO Class 5 Clean Room. The growth chamber maintains a slightly positive pressure relative to the surrounding lab, preventing entry of contaminants. Recirculated chamber air and supply gas pass through 99.99% HEPA filters, and the air change rate is once every 20 to 30 minutes to minimize cell desiccation. A sensor bay continually samples the chamber environment and makes adjustments when needed. + Positive + Direction of Filtered Outlet Flow HEPA Filtered Air...

Open the catalog to page 8
CO2 INCUBATORS - 9

Optional Surfaces Copper Surfaces Copper Option Add copper surfaces to the incubator growth chamber and shelving to reduce bacteria and minimize potential incubator contamination. Shelves and water pans fabricated from copper are available as options. Reduce the risk of contamination with the option of a growth chamber fabricated from copper

Open the catalog to page 9
CO2 INCUBATORS - 10

Temperature Uniformity and Recovery Because your samples matter, In-VitroCell CO2 Incubators from NuAire offer the fastest recovery times in the industry. Temperature Uniformity and Recovery The Direct heating elements or the water jacket are wrapped in R5 insulation which surrounds the growth chamber. Dual temperature sensor probes relay information back to the NuTouch Electronic Control System and make adjustments as needed. NU-5810 Temperature Recovery After 30 Sec. Door Opening 37.2 37.1 Open Closed Return to 37° C Set Point 15 Min. After Door Closed Time, Minutes Temp. Chamber...

Open the catalog to page 10
CO2 INCUBATORS - 11

Sensitivity and Accuracy of Gas Control A microprocessor-based, non-dispersive, single source dual wave infrared (IR) sensor controls CO2 levels within the chamber. The wavelengths used are absorbed by only CO2 making the measurement insensitive to other components such as water vapor. Advanced design provides a very stable output minimizing drift and requiring less frequent calibration. NU-5800 Series C02 30s Door Opening Recovery 6.0 CO2 Percentage Recovered 5.3 Minutes After Door Opening Meets Recovery Specification Door Open Recovery Delay Ends 0.3 Time, Minutes Range: 0.1 to 20%...

Open the catalog to page 11
CO2 INCUBATORS - 12

O2 and Relative Humidity Control Oxygen Control - Hypoxic Studies For more sensitive cells lines prone to stress and DNA damage, CO2 Incubator models featuring Oxygen control utilize an O2 sensor to monitor oxygen in the growth chamber. Nitrogen (N2) gas is injected to suppress oxygen to set point. An extensive O2 range applies to various research applications including the brain, eyes, liver, heart, kidneys, and more. Featured on O2 Control models only. Range: 0.5 to 21% Accuracy: ± 0.25% Recovery: 5% ± 2.0% within 15 Minutes Humidity Control Relative humidity levels up to 90% are achieved...

Open the catalog to page 12
CO2 INCUBATORS - 13

Selection Guide Chamber Volume Temperature Control Direct Heat Direct Heat Direct Heat Reservoir, Sensor Controlled Direct Heat Direct Heat Reservoir, Sensor Controlled Direct Heat Direct Heat Direct Heat Reservoir, Sensor Controlled Direct Heat Direct Heat Reservoir, Sensor Controlled Water Jacket Water Jacket Sterilization Cycles RH (Humidity) Control Water Pan, Convection Water Pan, Convection Water Pan, Convection

Open the catalog to page 13
CO2 INCUBATORS - 14

Incubator Options Gas Tight Sectioned Inner Doors Sectioned inner doors allow access to separate area of the growth chamber. A latch centered between each set of doors allows access to either with a turn of 90 degrees. [A] Each door is sealed with a gasket [B], and secured with a stainless steel hinge [C]. This option improves recovery times by limiting thermal loss and the amount of chamber atmosphere escaping when the main door is opened. Inner Door Sections Brackets. Safely stack In-VitroCell Direct Heat and Water Jacketed CO2 Incubators. Left Hinge Door Swing

Open the catalog to page 14

All Avidity Science catalogs and technical brochures