Patient Monitoring Solutions
12Pages

{{requestButtons}}

Catalog excerpts

Patient Monitoring Solutions - 1

Patient Monitoring Solutions

Open the catalog to page 1
Patient Monitoring Solutions - 2

For Out-Patient Department, Spot-check, Transport, Ward and other Basic Monitoring. Configuration Optional SpO2 + NIBP, Li-ion battery Features 8" color TFT LCD Screen (touch screen optional) Portable, lightweight and sturdy design Flexible parameters configuration for different clinical environments Rechargeable Li-ion Battery (up to 12 hours uninterruptable work) Big font and font color display setting Spot-check and continuous monitoring mode Selectable for Adult, Pediatric and Neonatal patients Wired/Wireless CMS, support HL7 protocol to HIS Barcode scanner support Thermal recorder...

Open the catalog to page 2
Patient Monitoring Solutions - 3

PAVO Vital Sign MonitorTechnical Specifications NIBP Measurement method: Automatic oscillometric method Operating mode: Manual, automatic, continuous Measurement unit: imimHg/kPa selectable Typical measurement time: 20~40 s Measurement type: Systolic, Diastolic, Mean Measurement range (mmHg) Range of Systolic pressure: Adult 40-270 Pediatric 40-230 Neonatal 40-135 Range of Diastolic pressure: Adult 10-210 Pediatric 10-150 Neonatal 10-100 Range of Mean pressure: Adult 20-230 Pediatric 20-165 Neonatal 20-110 Measurement accuracy Maximum average error: ±5 mmHg Maximum standard deviation: 8...

Open the catalog to page 3
Patient Monitoring Solutions - 4

Features • 12.1" color TFT LCD screen • 8 waveform display, up to 12-lead ECG analysis • Powerful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation) • Pacemaker detection • ST & arrhythmia analysis • OxyCRGs screen • Wired/Wireless CMS, support HL7 protocol to HIS • SpO2 pulse-tone modulation (Pitch Tone) • MEWS (Modified Early Warning Score) • Graphical & tabular trend review (120 hours) • Rechargeable Lithium-Ion Battery (2600 mAh) 12.1”color TFT LCD screen, wide and flat screen design,ecnomic and reliable Configuration: ECG+SpO2+NIBP+2TEMP+PR+RESP, Li-ion battery Optional:...

Open the catalog to page 4
Patient Monitoring Solutions - 5

CETUS X12 Patient Monitor Technical Specifications NIBP Measurement method: Automatic oscillome-tric method Operating mode: Manual, automatic, continuous Measurement unit: imimHg/kPa selectable Typical measurement time: 20~40 s Measurement type: Systolic, Diastolic, Mean Measurement range (mmHg) Range of Systolic pressure: Adult 40-270 Pediatric 40-200 Neonatal 40-135 Range of Diastolic pressure: Adult 10-210 Pediatric 10-150 Neonatal 10-95 Range of Mean pressure: Adult 20-230 Overpressure protection: Software and hardware, double safety protection Cuff pressure range: 0-280...

Open the catalog to page 5
Patient Monitoring Solutions - 6

CETUS x15 Critical Care Patient Monitor Multiple parameter options satisfy the need for ICU, CCU, NICU. Technical Specifications Configuration: 5-lead ECG, SpO2, NIBP, TEMP, Resp, PR; Li-ion battery Optional: Touch-Screen, 12-lead ECG, Masimo/Nellcor SpO2, IBP, C.O., EtCO2, Multi-Gas, BIS, NMT, VGA, Thermal Recorder, Wired/Wireless CMS Notch: 50/60 Hz notch filter can be set to on or off Differential input impedance >5MΩ Electrode polarization voltage range: ±400mV Baseline recovery time <3s after defibrillation (in monitor and surgery mode) Calibration signal: 1 mV (peak - peak), accuracy...

Open the catalog to page 6
Patient Monitoring Solutions - 7

CETUS X15 Critical Care Patient Monitor Technical Specifications Range of Mean pressure: Adult 20-230 Pediatric 20-165 Neonatal 20-105 Measurement accuracy Maximum average error: ±5 mmHg Maximum standard deviation: 8 mmHg Resolution: 1 mmHg Interval: 1,2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes Overpressure protection: Software and hardware, double safety protection Cuff pressure range: 0-280 mmHg IBP (Optional) Channel: 2-channel or 4-channel ART: 0 to 300 mmHg PA: -6 to 120 mmHg CVP/RAP/LAP/ICP: -10 to 40 mmHg Measurement range: P1/P2 -50 to 300 mmHg Resolution:...

Open the catalog to page 7
Patient Monitoring Solutions - 8

Aspect BISx module (Optional) Parameter Measurement: BC: 0~30 (Only limited to the combined use of an external sensor with a BIS module) EMG: 30~55 dB (bar chart) with intensity between 30 dB and 80 dB (tendency chart) BIS: 0~100 SQI: 0%~100% SR: 0%~100% SEF: 0.5 Hz~30 Hz TP: 40~100 Db EEG Measurement: Input impedance >5 MQ Noise (RTI) <2 pV (0.25~50 Hz) Input signal range: ±1 Mv EEG bandwidth between: 0.25 Hz~110 Hz NMT Tof-Watch® SX (Optional) Microprocessor-controlled Stimulation Mode: TOF, TOFS, PTC, 1 Hz Twitch, 0.1 Hz Twitch, DBS DBS3.3 and 3.2 (Double Burst), Tetanic Stimulation...

Open the catalog to page 8
Patient Monitoring Solutions - 9

CETUS xl Advanced Patient Monitor Multiple parameter options satisfy the needs of ICU, CCU, NICU Configuration: 5-lead ECG, SpO2, NIBP, Resp, PR; Li-ion battery Optional: 12-lead ECG, Masimo/Nellcor SpO2, IBP, C.O., EtCO2, Multi-gas, BIS, NMT; VGA, Thermal Recorder, Wired/Wireless CMS Masimo SET® SpO2 Provides anti-motion and anti-low perfusion SpO2 measurement. Masimo Phasein IRMA™/ISA Sidestream/Mainstream EtCO2 Allows selection of the modality best suited to the application, monitoring with infrared absorption technique. Features • 15.6”/17/19” switchable TFT LCD Touch Screen • Aluminium...

Open the catalog to page 9
Patient Monitoring Solutions - 10

Technical Specifications CMRR>100dB Notch: 50/60Hz notch filter can be set to on or off Differential input impedance >5 MQ Electrode polarization voltage range: ±400 mV Baseline recovery time <3s after defibrillation (in monitor and surgery mode) Calibration signal: 1 mV (peak - peak), accuracy ±3% RESP Measurement method: Thoracic electrical bioimpedance Measuring lead: Lead I, II Wave gain: x0.25, x0.5, xl, x2 Respiratory impedance range: 0.5-5 Q Baseline impedance: 500-4000 Q Gain: 10 grades Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s TEMP Accuracy: ±0.1 °C or ±0.2 °F (without...

Open the catalog to page 10
Patient Monitoring Solutions - 11

CETUS xl Advanced Patient Monitor Technical Specifications Standard Mainstream CO2 (Optional) Measurement range: 0-19.7%, 150 mmHg, or 0-20kPa Resolution: 0.1 mmHg Measurement accuracy 0 - 40 mmHg: ± 2 mmHg 41 - 70 mmHg: ± 5% of reading 71 - 100 mmHg: ± 8% of reading 101 - 150 mmHg: ± 10% of reading Respiration rate: 3-150 bpm Respiration rate accuracy: 1% ±1 bpm Warm-up time: 97% within 8s, full accuracy within 20s Standard Sidestream CO2 (Optional) Measurement rage: 0-20% (0-150 mmHg) Accuracy: < 5.0% CO 2: ± 2 mmHg > 5.0% CO 2: < 6% of reading Respiration rate: 2~150 BPM Respiration...

Open the catalog to page 11