Intensive Care and Transport Ventilator Solutions
Open the catalog to page 1LYRA x1 Best performing and versatile ventilation for hospital applications A compact turbine driven ventilator with multi-function, covers the non-invasive and invasive ventilation, and is suitable for treatment of most patient type. LYRA x1 is versatile throughout hospital and transport. Comprehensive ventilating modes, including APRV, PRVC, NIV are available for all your demands and for all type of patients from neonatal to adult. A collapsible high-resolution touch-screen display makes LYRA x1 mounted on a trolley your choice for ICU applications, as well as a high performance ventilator...
Open the catalog to page 2LYRA x1 Technical Specifications Physical Specication Dimensions: 336 mm x 330 mm x 345 mm (L x W x H): 664 mm x 600 mm x 1370 mm (with trolley) Weight: A pproximately 9.5 kg, Approximately 31.0 kg (with trolley) Screen Display Size: 12.1 Color active matrix TFT touch Display Resolution (H) x (V): 1280 x 800 pixels Brightness: Adjustable Ventilation Specications Patient Type: Adult, Pediatric, Neonate Invasive Ventilation Mode: VCV (Volume Control Ventilation) PCV (Pressure Control Ventilation) VSIMV (Volume Synchronized Intermittent Mandatory Ventilation) PSIMV (Pressure Synchronized Intermittent...
Open the catalog to page 3Apnea Ventilation Vtapnea: Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL / Neonate: 2-300 mL (increments of 1 mL) APapnea: 5-80 cm H2O (increments of 1 cm H2O) Fapnea: 1-80 bpm (increments of 1 bpm) Apnea Tinsp: 0.20-10 s (increments of 0.05 s) Interval: 20 s-180 min (increments of 1 s from 20 to 59 s, increments of 1 min from 1 to 180 min) Cycles Sigh: 1-20 (increments of 1 ) Tube Type: ET Tube, Trach Tube, Disable STRC Tube I.D.: Adult: 5.0 -12.0 mm (increments of 0.5 mm) / Pediatric: 2.5 - 8.0 mm (increments of 0.5 mm) Compensate: 0-100% (increments of 1%) Expiration Compliance...
Open the catalog to page 4Monitoring Accuracy Airway pressure (Ppeak, Pplat, Pmean, PEEP, PAP, EPAP): ±(2 cm H2O + 4% of the actual reading) Tidal Volume: (Tvi, Tve, TVe/IBW, TVe spn): 0 ml-100 ml: ±(10 ml + 3% of the actual reading) (BTPS) 100 ml-4000 ml: ±(3 ml +10% of the actual reading) (BTPS) Minute Volume (MV, MVspn, Mvleak): ±0.3 L/min or ±8% of the actual reading, Alarm settings Tidal Volume: High / Low Minute Volume: High / Low Airway pressure: High / Low Frequency: High / Low Inspired Oxygen (FiO2): High / Low etCO2: High / Low Apnea alarm time: 5-60 s whichever is greater (BTPS) Frequency (ftotal, fmand, fspn):...
Open the catalog to page 5LYRA x1 Technical Specifications High Flow Oxygen Therapy Controlled Parameters O2%: 21-100% (increments of 1%) Flow: 4-60 L/min Controlled Accuracy O2%: ±(3 vol.% +1% of setting) Flow: ±(2 L/min +10% of setting) (BTPS) Environmental specifications Temperature: 5-40 °C (operating); -20 to 60 °C (storage and transport, O2sensor: -20 to 50 °C ) Relative Humidity: 10-95% (operating); 10-95% (storage and transport) Barometric Pressure: 62-106 kPa (operating); 50-106 kPa (storage and transport) Power Battery Backup External AC power supply Input voltage: 100-240 V Input frequency: 50/60 Hz Input current:...
Open the catalog to page 6LYRA x2 Top notch performance of Non-Invasive and Invasive Ventilation LYRA x2 is a premium non-invasive turbine driven ventilator with no compromise on the performance in invasive ventilation. User can easily switch between NIV- and IV-modes by UI operation only. Comprehensive parameter monitoring describes the full scenario of patient’s status to the care giver. In a busy ICU it is imperative to give the desired mechanical ventilation to the patient. An 18.5 inch vertical layout touchscreen display makes operating of the ventilator smooth & easy.
Open the catalog to page 7LYRA x2 Technical Specifications Physical Specification Dimensions: 327 mm x 310 mm x 493 mm (L x W x H): 664 mm x 600 mm x 1520 mm (with trolley) Weight: Approximately 12.0 kg, Approximately 33.0 kg (with trolley) Screen Display Size: 18.5“ Color active matrix TFT touch Display Resolution (H) x (V): 1080 x 1980 pixels Brightness: Adjustable Ventilation Specifications Patient Type: Adult, Pediatric, Neonate Invasive Ventilation Modes: VCV (Volume Control Ventilation) PCV (Pressure Control Ventilation) VSIMV (Volume Synchronized Intermittent Mandatory Ventilation) PSIMV (Pressure Synchronized...
Open the catalog to page 8Ramp time: 5-45 min (increments of 5 min), Off Min P (VS minimum IPAP): 5-30 cm H2O (increments of 1 cm H2O) Max P (VS maximum IPAP): 6-40 cm H2O (increments of 1 cm H2O) Max P (PPV maximum pressure limit): 5-40 cm H2O (increments of 1 cm H2O) Max V (PPV maximum volume limit): 200-3500 mL (increments of 5 ml) Max E: 0-100 cm H2O/L (increments of 1 cm H2O/L) Max R: 0-50 cm H2O/L (increments of 1 cm H2O/L) PPV%: 0%-100% (increments of 1%) Apnea Ventilation Vtvapnea: Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL / Neonate: 2-300 mL (increments of 1 mL) ΔPapnea: 5-80 cm H2O (increments...
Open the catalog to page 9LYRA x2 Technical Specifications Control Accuracy O2%: ±(3 vol.% + 1% of setting) TV: ±(10 mL + 10% of setting) (BTPS) Tinsp: ±0.1 s or ±10% of setting, whichever is greater I: E: 2:1 to 1:4: ±10% of setting, other range: ±15% of setting f: ±1 bpm fSIMV: ±1 bpm Tslope: ±(0.2 s + 20% of setting) PEEP: ±(2.0 cm H2O + 5% of setting) ΔPinsp: ±(2.0 cm H2O + 5% of setting) ΔPsupp: ±(2.0 cm H2O + 5% of setting) Phigh: ±(2.0 cm H2O + 5% of setting) Plow: ±(2.0 cm H2O + 5% of setting) Thigh: ±0.2 s or ±10% of setting, whichever is greater Tlow: ± 0.2 s or ±10% of setting, whichever is greater Pressure...
Open the catalog to page 10Alarm settings Tidal Volume: High / Low Minute Volume: High / Low Airway pressure: High / Low Frequency: High / Low Inspired Oxygen (FiO2): High / Low etCO2: High / Low Apnea alarm time: 5-60 s Trend Type: Tabular, Graphic Length: 72 hours Content: Monitor Parameters, Setting Parameters (Setting Ventilation mode and Parameters) High Flow Oxygen Therapy Controlled Parameters O2%: 21-100% (increments of 1%) Flow: 4-60 L/min Controlled Accuracy O2%: ±(3 vol.% +1% of setting) Flow: ±(2 L/min +10% of setting) (BTPS) Environmental specifications Temperature: 5-40°C (operating); -20 to 60 °C (storage...
Open the catalog to page 1112 Pages
24 Pages
12 Pages
7 Pages