Main Parts
The EO exhausted gas scrubber system consists of several components including an EO gas-water separator, Level I, II, and III acid washing towers, a gas filter unit, an alkali neutralization tank, and an acid liquid match tank. The system is equipped with anti-acid and anti-alkali pumps, valves, tubes, fans, air channels, and a controller system to facilitate acid liquid spraying, washing, acid-alkali neutralization, exhausted gas emission, and liquor transfer functions. It is integrated with the EO sterilizer washing process, initiating residual gas treatment upon receiving a signal from the sterilizer.

Automation
The entire scrubber system is automatically controlled by a PLC and dynamically displayed on a touch screen.

Main Working Process
When the sterilizer begins washing, the system starts. The input EO passes through the gas-water separator and into the Level I washing tower where it reacts with acid and water, absorbing the EO gas. The residual EO then enters Level II and Level III washing towers for further absorption. The final gas is filtered and released into the atmosphere. Saturated ethylene glycol from the washing towers is pumped into the neutralization tank where NaOH is added until the pH reaches 7, after which the liquor is transferred to the spend liquor tank for management.

Working Principle
The EO gas is absorbed in the pickling tower, reacting with acid liquid to form glycol as per the reaction: C2H4O + H2O → C2H4(OH)2.

Key Factors
Two main factors must be controlled: the pH of the acid liquid and the saturation degree of glycol. The acid liquid typically consists of 1%-5% dilute sulfuric acid with a pH less than 1.5. If the pH exceeds this, more acid is added. One cubic meter of water can absorb a specific amount of EO, necessitating a change of acid liquid if EO consumption exceeds this.

Efficiency and Customization
The system achieves a clarification rate of over 99.9% for EO. It is customized based on customer requirements and local government standards, with different models available based on processing capacity and purifying rate:
  • Level I: Processes about 20kgs EO gas per day with a 90% purifying rate.
  • Level II: Processes about 40kgs EO gas per day with a 96% purifying rate.
  • Level III: Processes about 60kgs EO gas per day with a 99.9% purifying rate.
See more

Catalog excerpts

AXIS INDUSTRIAL ETHYLENE OXIDE GAS SCRUBBERS-1

AXIS INDUSTRIAL ETHYLENE OXIDE GAS SCRUBBERS Main Parts EO exhausted gas scrubber system is composed by below parts      EO gas-water separator Level I, Level II & Level III acid washing tower Gas Filter Unit Alkali Neutralization tank Acid Liquid Match Tank, etc. .The system allocated kinds functions anti-acid(anti alkali) pump,valve,tube,fan,air channel and controller system. To finish the acid liquid spraying and washing,acid-alkali neutralization,exhausted gas emission, liquor transfer functions.This system is linkage with EO Sterilizer washing process. That is when the gas exposure stage is finished,the sterilizer is in washing process, the sterilizer must send a single(standard single,such as switch single,coded single,etc),the residual gas treatment will start after received this

 Open the catalog to page 1
AXIS INDUSTRIAL ETHYLENE OXIDE GAS SCRUBBERS-2

The whole scrubber system is automatically controlled by PLC, and dynamic displayed on the touch screen.

 Open the catalog to page 2
AXIS INDUSTRIAL ETHYLENE OXIDE GAS SCRUBBERS-4

Main working process When the sterilizer begin washing,this system is started.The input EO must be through the gas-water separator,then into the Level I washing tower, EO contact fully with the acid,and react with the water,EO gas is absorbed. After the Level I process, the residual EO enter into the Level II washing tower, to be absorbed by second time .After twice absorption,then enter into Level III washing tower.Then the final gas go through the filter unit and fan system and enter into the atmosphere. When the product of the washing tower ethylene glycol is saturated, the acid liquid is...

 Open the catalog to page 4
AXIS INDUSTRIAL ETHYLENE OXIDE GAS SCRUBBERS-5

2.There are two main factors must be controlled:The PH of the acid liquid and the degree of the saturation for the glycol. The acid liquid is generally 1%~5% dilute sulphuric acid,the PH is less than 1.5,if bigger than this,there needs add the acid.Generally 1m³ water can absorb a determined value EO,when the EO consume is bigger than this,there need change the acid liquid in the acid tank. 3. After the treatment system,the clarify rate for the EO is more than 99.9% IV. The treatment system is customized according to customers’ requirements and the local governments’ standard.Different requirements...