Overview
This document provides detailed information on medical information technology, specifically focusing on the standards and products offered by Baaske Medical GmbH & Co.KG. It is intended for manufacturers, resellers, and users of medical PCs and related equipment.

Medical PC Specifications
A Medical PC is designed to meet medical standards and is equipped with a medically approved power supply according to IEC 60601-1 Safety standards. It is protected against electromagnetic interferences (EMI) and does not emit any, complying with IEC 60601-1-2 EMC standards. Each unit is individually tested for compliance with IEC 601 during production.

Patient Environment
The 'Patient Environment' is defined according to IEC 60601-1:2006 Subclause 3.79. This section outlines the safety requirements for medical devices used in proximity to patients, ensuring minimal leakage current and safe operation.

Leakage Current and Safety
The document discusses the importance of managing leakage current in medical environments. It provides examples of normal and single fault conditions, emphasizing the need for protective grounding and isolation to prevent hazardous currents.

Product Offerings
  • Ethernet Isolation: The MI 1005 Medical Isolator is designed for use in Ethernet networks, blocking unwanted currents and voltages.
  • Power Supply: Medical-grade isolation transformers and desktop adapters are available to meet strict leakage current requirements.
  • Interfaces: USB and RS232 isolators provide electrical isolation and extended range capabilities.
  • Medical PC Systems: These systems comply with low voltage and EMC directives, certified under DIN EN 60601-1 and DIN EN 60601-1-2.


Case Studies
The document includes case studies based on room plans, illustrating the application of medical PCs and electrical safety products in various medical settings, such as surgeries and waiting rooms.

Conclusion
Baaske Medical GmbH & Co.KG offers a range of products designed to enhance electrical safety in medical environments, ensuring compliance with international standards and providing reliable solutions for medical information technology.
See more

Catalog excerpts

Medical Information technology-1

Medical Information Technology Medical Information technology Information for manufacturers, resellers and users Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 1
Medical Information technology-2

Medical Information Technology Medical PC … definition A short trip into the world of electrotechnology Case study based on a room - plan Baaske Medical Products for electrical safety Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 2
Medical Information technology-3

Medical Information Technology A „Medical PC“ … . . . Is desinged to meet medical standards . . . is equipped with a medical approved power supply according to (IEC 60601-1 Safety) . . . is protected against electromagnetical interferences (EMI) ……(IEC 60601-1-2 EMC) . . . does not emit any electromagnetical interferences ……(IEC 60601-1-2 EMC) . . . is individually tested for IEC 601 in line production Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 3
Medical Information technology-4

Medical Information Technology … can be combined with active medical devices (new System) … no medical device according to MDD 93/42 EEC … allowed to be used in ‚patients environment‘defined …. in IEC 60601-1 Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 4
Medical Information technology-5

Medical Information Technology The ‚Patient Environment‘ is defined in IEC 60601-1:2006 Subclause 3.79 Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 5
Medical Information technology-6

Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 6
Medical Information technology-7

Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 7
Medical Information technology-8

Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected] Patientenumgebung und Ableitstrom – DIN EN 60601

 Open the catalog to page 8
Medical Information technology-9

Medical Information Technology The Leakage Current Normal Condition Single Fault Condition (Protective Earth broken) Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 9
Medical Information technology-10

Example for missing protective Ground Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 10
Medical Information technology-11

Example for missing protective ground Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 11
Medical Information technology-12

Medical Information Technology Leakage Current in normal case (NC) 2. Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 12
Medical Information technology-13

Medical Information Technology Leakage current in single fault condition (SFC) 2. Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 13
Medical Information technology-14

Medical Information Technology Example: Medical electrical system (MES) Ethernet Patients Monitor Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 14
Medical Information technology-15

Medical Information Technology Resistance of the human body Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 15
Medical Information technology-16

Threshold levels for current conduction Medical Information Technology Threshold level unhand Ventricular level fibrillation level Example for the current conduction between hand and both feet for an adult person. The unhand level will be reached after a conduction Period of 10 seconds at 11 mA. Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske - Bacmeister Str. 3- 32312 Liibbecke - Phone (0 57 41) 236 270 - Email: [email protected]

 Open the catalog to page 16
Medical Information technology-17

Example: Stress ECG with bicycle ergometer Medical Information Technology Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 17
Medical Information technology-18

MES: Reducing current with Ethernet Isolation Medical Information Technology Ethernet Isolation Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 18
Medical Information technology-19

MES: Reducing Current with Monitor Power Set Medical Information Technology Ehternet Isolation Patients Cable Set Office PC Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 19
Medical Information technology-20

MES: Reducing Current using a Medical PC Medical Information Technology Ehternet Isolation Applied Part Created by: Baaske Medical GmbH & Co.KG - Andreas Baaske – Bacmeister Str. 3– 32312 Lübbecke – Phone (0 57 41) 236 270 – Email: [email protected]

 Open the catalog to page 20
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.