Overview: The HI VAC II steam sterilizer is designed for general sterilization, lactary, and special cycles, offering chamber sizes from 140 to 891 liters with stainless steel construction.
Selection and Specifications: The sterilizer features chamber thickness options of 4.763 mm and 7.938 mm, made from stainless steel 316L or 316Ti, with polished or electropolished finishes.
Command and Control: It includes a high-tech command panel with Siemens technology, matrix or thermal printers, and supervisory control systems, along with sensors for chamber and loading.
Door and Steam Configuration: Available with one or two doors, it can be fed by direct steam or a steam generator, with vacuum systems including liquid ring or dry vacuum pumps.
Construction and Standards: The chamber is constructed with a double wall and complies with international standards like EN 285:2006, ISO 17665, and IEC 61010.
Applications: Suitable for hospitals, clinics, and laboratories, it handles surgical instruments, glassware, and liquids, supporting Bowie & Dick and Leak Tests.
Features: User configuration and programming are supported, with a guillotine-type door, touch screen interface, and optional printer and recorder.
Construction Details: The chamber and jacket are tested for high pressure, with thermal insulation and carbon steel support with anti-corrosive treatment.
Sterilization Cycles: Offers multiple programs with specific cycles for different materials, including Program B for packages and latex, Program C for heavy instruments, and Program HL for lactary sterilization.
Cycle Descriptions: Configurable parameters include vacuum depth, sterilization temperature, and exposure time, with specific cycles for materials like latex and heavy instruments.
Components and Accessories: Configurable with various vacuum pumps, steam generators, and air compressors, supporting dual steam supply and different power configurations.
Safety Features: Includes automatic door locking, emergency routines for power failures, and alarms for water, steam, and air failures.
Installation and Engineering Data: Provides detailed installation specifications, emphasizing proper drainage and electrical protection, with dimensions and weight for different models.
Technical Specifications: A table outlines technical specifications for various models, including chamber dimensions, capacity, power requirements, and connections for water, steam, and compressed air.
See more

Catalog excerpts

Steam Sterilizer HI VAC II-1

Steam Sterilizer HI VAC II SELECTION Useful space of the chamber (Medium) □ B0110-140 - 460 x 460 x 660 mm □ B0110-200 - 460 x 660 x 660 mm □ B0110-263 - 460 x 460 x 1245 mm □ B0110-370 - 460 x 660 x 1245 mm □ B0110-418 - 660 x 660 x 960 mm □ B0110-542 - 660 x 660 x 1245 mm □ B0110-716 - 660 x 660 x 1645 mm □ B0110-890 - 660 x 660 x 2045 mm Thickness of the chamber □ 4,763 mm (3/16”) □ 7,938 mm (5/16”) Chamber Material □ Stainless steel 316L □ Stainless steel 316Ti Camera Finish □ Polished (standard) □ Electropolished Application □ General sterilization □ Lactary □ Special cycles Command □ Hi Technology □ Siemens Printer □ Matrix Printer □ Thermal Printer Recorder □ Recorder □ Supervisory control system Control Sensor □ Chamber additional sensor □ Sensors with RBC calibration certificate □ Loading Sensor Electric Supply Power □ 220 V □ 50 Hz □ 380 V □ 60 Hz Door configuration □ One door □ Two doors Steam Feeding □ Direct steam (Standard) □ Switch (direct steam more steam generator) □ Steam Generator 18 kW / 30 Kg/h □ Steam Generator 36 kW / 60 Kg/h □ Steam Generator 44 kW / 73 Kg/h □ Steam Generator 52 kW / 86 Kg/h □ Steam Generator 72 kW / 120 Kg/h □ Steam Generator with industrial steam pipe coil Vacuum System □ Liquid ring vacuum pump 1,5 hp □ Liquid ring vacuum pump 3,0 hp □ Liquid ring vacuum pump 4,0 hp □ Liquid ring vacuum pump 5,0 hp □ Liquid ring vacuum double stage pump 3,0/5,0 hp □ Dry vacuum pump 3,0 hp Piping □ Brass □ Stainless steel Proportional Integral Derivative Valve □ PID Heat Exchanger □ Heat exchanger Structure □ Painted finish □ Stainless steel Front Cabinet □ Painted finish (default) □ Stainless steel Lateral closing □ Without closing (standard) □ Painted finish - Single □ Stainless steel - Single □ Stainless steel - Articulate Upper closing □ Without closing (standard) □ Painted finish □ Stainless steel Back Closure □ Without closing (standard) □ Painted finish □ Stainless steel Tghtness □ No Tightness □ Hospitalar - Painted □ Hospitalar - Stainless steel □ Biological - Stainless steel Air Compressor □ Compressed air system (standard) □ Compressor R0100-170 (separate) Accessories □ Load rack □ Tray for rack □ Rack trolleys □ Wire mesh baskets □ Floor support □ Wall support □ Casters Installation Components □ Steam valve system □ Sanitary filter for steam □ Water valve □ Water treatment □ Compressed air valve □ Electric panel Services □ Points installation □ Replacement Parts Kit □ Validation □ Preventive Maintenance □ Training □ Extended Warranty □ Qualification □ Gamp 5 Documentation APPLICATION Sterilizer designed to meet the most demanding sterilization parameters and normatives of production and national and international safety, being able to dry and sterilize density materials such as fields, surgical clothes, instruments and utensils wrapped or not, glasses, gloves, syringes, rubber, liquid in airtight jars or not with calculation of F0, especially necessary in quality control laboratories with specific program for Bowie & Dick and Leak Test. Suitable for use in hospitals, emergency and ambulatory care units, medical clinics, testing laboratories and clinical laboratories for quality control. Using saturated steam from pressurized water with conditioning load phase with air removal through vacuum and steam pressure pulses and vacuum drying. MODELS AND CAPACITY HI VAC II line is composed of small, medium and large sterilizers. B0110-140 - capacity: 140 l B0110-200 - capacity: 200 l B0110-263 - capacity: 263 l B0110-370 - capacity: 378 l B0110-418 - capacity: 418 l B0110-542 - capacity: 542 l B0110-716 - capacity: 717 l B0110-890 - capacity: 891 l MAIN CHARACTERISTICS Configuration by the user: The HI VAC II allows the configuration of the equipment at purchasing time, aiming a better adequacy of the equipment to the needs of the user. User Accessible programming:

 Open the catalog to page 1
Steam Sterilizer HI VAC II-2

Steam Sterilizer HI VAC II parameters through command interface and password. Guillotine type door: with vertical movement, automatically driven with no physical effort. Equipment may be configured with single or double door for installation at sanitary barrier. Chamber: with central fold to increase the withdrawal of condensate, optimizing process and saving water. Piping: All the existing piping is AISI 316 stainless steel (optionally brass / bronze). No Tightness or With Tightness: collapsible, making easier to handle through elevator doors, hallways, and standard dimension door-post during transport...

 Open the catalog to page 2
Steam Sterilizer HI VAC II-3

Steam Sterilizer HI VAC II printer, eliminating the need of a separate line recorder. Command Valves: control the entry of steam into the chamber and the drain outlet, independent and pneumatic driven, built in AISI 316 stainless steel, solenoid valves actuated by compressed air. Electric solenoid valves for controlling the other electrical supply lines. CONSTRUCTION Chamber: built with double wall, with central crease and drop. Internal Chamber: built in AISI 316L Stainless steel or AISI 316 Ti, with sanitary polish (optionally, it may be supplied with electropolishing). Jacket: built in AISI...

 Open the catalog to page 3
Steam Sterilizer HI VAC II-4

Steam Sterilizer HI VAC II □ Exposure time between 0 and 999 min. □ Drying time from 0 to 999 min. □ End of cycle temperature - between 80°C and 98°C. Available only for cycle of liquids. COMPONENTS E ACCESSORIES Configurable for all the HI VAC II models. Vacuum system: through a monoblock water-ring vacuum pump, optionally it can be supplied with liquid ring vacuum double stage pump or dry vacuum pump which provides up to 70% more economy of filtered water. The system is responsible for the vacuum pulses in the initial load conditioning and final drying. Dimensioned to comply with ABNT NBR 11816. The...

 Open the catalog to page 4
Steam Sterilizer HI VAC II-5

Steam Sterilizer HI VAC II TABLE WITH EQUIPMENT PLANT, VIEW AND PERSPECTIVE MINIMUM DISTANCE ONLY ON LEFT SIDE LOAD SIDE UNLOADING SIDE LOADING SIDE FRONT VIEW SIDE VIEW SUGESTED POSITION SUPPLY POINTS f2 REVERSE OSMOSIS OPTIONAL PARTITION POSITION MASORY POSITION 2,00 m Minimum operating Distance DRAIN POSITION 2,00 m Minimum operating Distance LOADING SIDE PLANT VIEW