TECHNICAL SPECIFICATIONS
2Pages

{{requestButtons}}

Catalog excerpts

TECHNICAL SPECIFICATIONS - 1

TECHNICAL SPECIFICATIONS Laser Type Diode-pumped solid state laser (DPSSL) Pulse Frequency Pulse Duration Power Supply Power Outlet Heat Load L: 81.7 in (without patient bed) L: 82.7 in (with patient bed) W: 32.5 in H: 65.9 in System Components Visualization Patient Interface Main laser unit Patient bed included Sterile patient interface kit Live-action, high-contrast OCT High-resolution video microscope Optional external microscope Intelligent Pressure Sensors Curved interface with separate suction clip Capsulotomy Cataract Phacofragmentation Procedures Corneal incisions Arcuate incisions Corneal Procedures LASIK flap Room Conditions Temperature 64.4°F to 75.2°F, controlled to +/- 2°F Humidity 30% to 50%, noncondensing Free of dust and particles; no carpet No solvent, chemical liquids, or fumes Floor under laser supports 15.6 lb/in2 Room Dimensions

Open the catalog to page 1
TECHNICAL SPECIFICATIONS - 2

VICTUS Technical Specifications ® VICTUS System Dimensions Installation Requirements All corridors and doorways leading to the laser room must be wider than 33 inches (no doorstops less than 79 inches behind doors) Corridors narrower than 47.5 inches must have no 90° corners Top View If an elevator must be used, it must have a minimum length of 79 inches and a minimum width of 33 inches, and accept a minimum load of 1874 lb Floors must not have gaps > 1 inch If a ramp is necessary to overcome stairs, an angle of 20° should not be exceeded The floor leading to and in the room must support...

Open the catalog to page 2