video corpo

Local alarm

Local alarm

Local alarm

Product catalog summary
Medical Gas Area Alarm Overview
The document outlines the specifications and features of the BeaconMedæs Medical Gas Area Alarm, designed to monitor up to six medical gas services using pressure sensors. It complies with HTM2022, HTM 02-01, and BS EN ISO 7396-1 standards.

Specifications
  • The alarm panel is equipped with a 7” LCD touch screen displaying pipeline pressure and status, with both normal and dark mode settings.
  • It includes a silver antimicrobial additive for user-accessible parts.
  • Gas service status is indicated by color-coded badges: green for normal, red flashing for faults, and "ERR" for sensor communication loss.
  • Configurable for various 4-20mA sensor inputs.

Alarm Features
  • An audible alarm with a mute function that reactivates after 15 minutes or upon a new alarm condition.
  • A test function to verify internal circuits and alarm status.
  • Editable alarm service location legend and password-protected access to alarm logs and setup menus.

Power and Connectivity
  • Operates on 110V-230V, 50/60Hz AC power with a current requirement of 3.0 amps.
  • Includes relay output for integration with other alarms or building management systems.
  • Remote audible connection for external buzzers.

Data Logging and Networking
  • Data logging capability for up to 1000 events, accessible through password-protected screens.
  • Networking capability for repeating alarm signals at other locations.

Installation and Part Numbers
  • Detailed installation instructions with wiring terminal connections provided.
  • Part numbers for various components and related products are listed for ordering.

Compliance and Certification
  • CE marked under the Medical Devices Directive 93/42/EEC as Class IIb Medical Devices.
See more

Catalog excerpts

Local alarm-1

Medipoint 26 Digital (MP26D) Local Alarm Medical Gas Area Alarm SPECIFICATION Medical Gas Area Alarm Gas Service Status Badges Each medical gas area alarm panel shall be capable of monitoring 6 medical gas services by means of pressure sensors, which detect deviations from the normal operating limits of either pressure or medical vacuum. The medical gas area alarm shall fully comply with the requirements of HTM2022, HTM 02-01 and BS EN ISO 7396-1. The gas status bar shall indicate Green steady for normal condition, and change to red flashing for fault conditions. Loss of comunication with the gas sensor is indicated by changing the pressure reading to “ERR” and gas ID bar to Red steady. The gas pressure is displayed for each service along with a vertical bar showing the alarm set points relative the the live pressure reading. The cover, backbox and bezel (if required) shall be polyester powder coated in a RAL9010 30% gloss finish. A single tamperproof fastener shall be used to gain access to the hinged door. The hinge shall operate through a minimum of 120° to provide adequate access. Pressure Fault Sensor Fault Antimicrobial Additive User accessible parts such as the screen protector, key pad fascia shall include a silver antimicrobial additive for inherent antimicrobial protection. Alarm Status Screen Each gas service shall be displayed by coloured 7” LCD touch screen to display the pipeline pressure and status for normal and fault conditions in line with HTM02-01 and HTM2022 requirements. Screen graphics are avaialble in normal and dark mode settings for reduced brightness. Alternative Sensor Configuration Each input can be configured for many common 4-20mA sensors inputs including alarm trigger points from the on screen setup menus. Typical examples of customised sensors Typical normal mode view Audible Alarm and Mute function An audible warning shall sound simultaneously with any failure indication and a mute facility shall be provided. Following a mute selection the audible will resound after approximately 15 minutes, or shall operate simultaneously should a further alarm condition occur. A “Mute” switch shall be provided inside the panel for use during any maintenance resulting in prolonged pipeline or plant shutdown. This facility shall automatically reset when the gas service returns to normal. Icon Typical dark mode view Mute Button Press to activate to silence the audible alarm. Test Function The alarm panel shall have a ‘Test’ facility to prove the integrity of the internal circuits, Gas status and audible warning. Icon Description Test Button Press to activate the alarm test button. BeaconMedæs • Telford Crescent, Staveley, S43 3PF • United Kingdom • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com MEDIPOINT 26 AREA ALARM

 Open the catalog to page 1
Local alarm-2

Editable Alarm Service Location Legend Power on and System Fault LED Indicators The top strip on the screem display has a field for displaying the details for the deparment location servered by the gases that can be edited through the setup menus. Each alarm shall provide a green LED to indicate that electrical power is available at the panel and a red LED to indicate ‘System Alarm’. In the event of a system fault or electrical power supply failure the ‘System Alarm’ LED shall illuminate (flashing). Main Screen Interactive Icons There is interatable icons for accessing the alarm information screen,...

 Open the catalog to page 2
Local alarm-3

Relay Output Part Numbers The alarm includes a relay output connection to allow the overal status of the alarm to other alarms or building management systems. Any fault condition on the alarm will change the status of the relay output. For complete alarm select 1st fix, 2nd fix and appropriate number of pressure & vacuum transducers. Part Number Main Product group Remote Audible The alarm includes a connection for a remote buzzer. Any condition that activates the alarm’s internal audible warning will also activate the remote buzzer. Mute functions will silence the remote buzzer as per the internal...

 Open the catalog to page 3
Local alarm-4

In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Installation Details C 2 x 20 CABLE GLAND KNOCKOUTS VIEW A 5 x 20 CABLE GLAND KNOCKOUTS Wiring Terminal Connections Remote Audible Connection for remote buzzer for audible alarm. N/O Volt free Relay output Normally open relay output triggered from any alarm fault condition to to another alarm or Building Management System. Gas sensor or alternative 4-20mA transmitter input connection...

 Open the catalog to page 4

All BeaconMedaes catalogs and technical brochures

  1. Central alarm

    4  Pages

  2. Magnis MSV

    2  Pages

  3. mVAC

    12  Pages

Archived catalogs

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.