Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version) Medical Gas Supply System Description and intended use The BeaconMedaes Medical Gas Supply System, mainly consists of cylinder manifold system, manifold changeover system, header & tailpipe and other necessary components, is intended to work in conjunction with other devices within the medical gas pipeline system (MGPS) to assist with maintaining continuity of supply throughout the hospital to local connection points for eventual introduction into the human body. The device controls the gas pressure within the specification as required by the MGPS. In all cases other devices outside of this scope are connected downstream of the MGPS are required to control the final gas pressure and flow requirements for safe supply to the patient. Therefore, this device is not intended to control supply parameter hazards directly to the patient, only to the MGPS. The device including Cylinder Manifold System and Manifold Changeover System (abbreviate as manifold below) is principally designed for use as a primary or secondary source of supply, or for emergency backup. The Medical Gas Supply System supplies one of the following medical gases to a piped distribution system, Oxygen, Nitrous Oxide, O2/N2O 50%: 50%, Medical Air, Surgical Air, Nitrogen & Carbon Dioxide. BeaconMedaes • No.45 Ximei Road, Xinwu District, Wuxi City, Jiangsu Province, 214111, China • www.beaconmedaes.com MAT-S Medical Gas Supply System
Open the catalog to page 1BeaconMedaes MAT-S Medical Gas Supply System The distinctive features of MAT-S A constant supply of medical gases such as oxygen, nitrous oxide, and medical air is critical to ensure safe, quality patient care. That is why we designed the new MAT-S Medical Gas Supply System to deliver superior reliability. Our medical gas supply system offers a higher utilization rate and uptime for using it more frequently and for longer periods. In addition, its smart features make it easy and efficient to operate the net result: is premium performance and durability and lower operational and maintenance costs....
Open the catalog to page 2SPECIFICATION Control System The control of the various changeover system conforms to NHS Health Technical Memorandum 06-01(HTM06-01) Electrical services supply and distribution. Following the Chapter 11 requirement, the manufacturer will provide evidence of Electromagnetic compatibility (EMC) for the manifold, e.g., an EMC test certificate. The pressure sensor monitors the pressure in each bank of cylinders and bank pressure gauges attached to each bank regulator (1st stage). A pressure sensor measures delivery pressure, and a delivery pressure gauge is attached to each line (2nd stage) regulator...
Open the catalog to page 3Output Flow MAT-S output flow is tested according to ISO 10524-2. MAT-S flow performance Volumetric Flow (l/min) Based on air and assuming sufficient cylinders are connected and zero pipeline pressure loss: Static = 0 l/min at 4.8 bar ‘X’ = 600 l/min at 4.5 bar ‘Y’ = 1080 l/min at 4.3 bar ‘Z’ = 1450 l/min at 4 bar Nominal delivery / Line output pressure MAT-S flow performance Volumetric Flow (l/min) Based on air and assuming sufficient cylinders are connected and zero pipeline pressure loss: Static = 0 l/min at 11.5 bar ‘X’ = 1250 l/min at 10.8 bar ‘Y’ = 1769 l/min at 10.5 bar ‘Z’ = 2050 l/min...
Open the catalog to page 4Typical Automatic Changeover Manifold Schematic Hospital Pipeline Relief Valve Exhaust Lockable Ball Valve Sampling Outlet Main Gas Line Control Gas Line Ball Valve Non-Return Valve Line Pressure Sensor Exhaust Line Vent Valve Service Valve Line Relief Valve Service Relief Valve Dome Regulator* Ball Valve Service Relief Valve Ball Valve Pressure Gauge * Pressure Gauge Ball Valve Ball Valve Service Relief Valve Header With Non-Return Valve Which Have Gas Specific Connection Intermediate Relief Valve Non-Return Valve Intermediate Relief Valve Service Relief Valve Non-Return Valve Pressure Gauge...
Open the catalog to page 5Typical Cylinder Header & Extension Layout Details Note : 1.Manifold weight is 35kg. 2.Dimension: Length*Width*height=460mm*200*1100 Headers (with rack and all components) / DETAIL OF MOUNTING BRACKET Gas specific nonreturn valves Cylinder Header Mounting Bracket Gas specific nonreturn valve Mounting Bracket Chain pin Wall mounting points NOTE - 3/8” plug and bonded seal is required on the last cylinder header on each bank. Cylinder Header Wall mounting points Note : 3/8" plug and bonded seal is required on the last cylinder header on each bank. Cylinder chain pin hook points Copper Washer Drawing...
Open the catalog to page 6Manifold Designation e.g.,MAT-S-HH-O2-A A-B-C-D-E Variable Allowable Value Allowable Value Description Model name Fully Automatic Manifold Smartbox & MyMedgas enabled Low-pressure liquid cylinder Input type Nominal Delivery / Line pressure Low Changeover Pressure Part No High Changeover Pressure Part No Headers (with rack and all components)/ both sides Gas Type *If request headers quantity is above 2x6 need to add up with available configuration. For example: for 2x7 may order 2x3 and 2x4. BeaconMedaes • No.45 Ximei Road, Xinwu District, Wuxi City, Jiangsu Province, 214111, China • www.beaconmedaes.co...
Open the catalog to page 7Tail pipe Gas Type Bull nose (ISO5145) Side entry Bull nose (BS341) Top entry Bull nose (BS341) Side entry Bull nose (BS341) Extended Chinese Bullnose Nitrogen (N2) Carbon Dioxide (CO2) Note: 1.Bull nose tailpipes (except Chinese type) are according to the following B.S. standards: Oxygen, Air, Nitrogen: BS: 341-1 No. 3; Carbon dioxide: BS: 341-1 No. 8; Nitrous oxide: BS: 341-1 No. 13. 2.Mixture N2O - O2 (registered trade name Entonox BOC) low-pressure cylinder “G” type has a Pin-indexed connector according to standard BS EN ISO 407, and 230 bar cylinder “E.W.” type has Bull nose connector according...
Open the catalog to page 8Free standing configuration Post for manifold support with Dimension shown below. The illustration below shows the new free standing option. The free standing assembly is a contained, optimised and space saving configuration. Main parts are: Descrition Note: To be supplied with kit [4109626500] Post for manifold support Post for Cylinder rack support Manifold Header Assembly Post for Cylinder rack support with Dimension shown below Descrition Part Number Post for manifold support Post for Cylinder rack support Drop Down Connector Post joint to Cylinder rack Note: To be supplied with kit [4109626600]...
Open the catalog to page 93 Pages
4 Pages
4 Pages
4 Pages
8 Pages
4 Pages
2 Pages
4 Pages
8 Pages
4 Pages
4 Pages
4 Pages
8 Pages
6 Pages
8 Pages
2 Pages
2 Pages
12 Pages
5 Pages
4 Pages
4 Pages
4 Pages
4 Pages
2 Pages
2 Pages
4 Pages
4 Pages
2 Pages
2 Pages
2 Pages
39 Pages
6 Pages
6 Pages