1. Catalogs
  2. BeaconMedaes
  3. Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)
video corpo

Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)

Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)
1 / 10 PagesView full catalog

Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)

Product catalog summary
Medical Gas Supply System Overview
The BeaconMedaes Medical Gas Supply System is engineered to ensure a continuous supply of medical gases such as oxygen, nitrous oxide, and medical air in hospitals. It includes components like a cylinder manifold system and a manifold changeover system, serving as a primary or secondary supply source or emergency backup. The system operates within the medical gas pipeline system (MGPS) to maintain gas pressure but does not directly control supply parameters to patients.
Distinctive Features of MAT-S
The MAT-S system is noted for its high reliability and efficiency, offering features that reduce operational and maintenance costs. It provides two variants for 4 bar systems: a high flow delivery option and a cost-saving option with different changeover pressures. Key features include a lightweight ABS plastic cover, dome-loaded forged regulators, automatic changeover pressure settings, individual service valves, a built-in test terminal unit, and a smart digital controller.
Manifold Design
The manifold design incorporates two stages of regulation to maintain high peak flow rates. It features integrated forged brass housing, non-return valves, and a robust ABS plastic case for durability. The system is designed for easy maintenance with 'O' ring sealed connections and a removable control panel.
Control System
The control system complies with NHS HTM06-01 standards and includes features like a color graphical interface, alarm logs, and a mechanical pressure gauge backup. It supports RS485 and BACnet communication and includes a 7-inch touch screen for user interaction. The system facilitates easy maintenance and operation with features like a maintenance timer and smartbox connectivity.
Power Supply and Operation
The system operates on a wide power range and includes over-voltage and overload protection. It automatically switches between duty and standby banks to ensure continuous supply and includes manual changeover options for servicing.
Specifications and Materials
The MAT-S system is tested according to ISO 10524-2 standards, with flow performance detailed for different pressure levels. All materials used are halogen-free and compatible with medical gases. The modular header manifold includes non-return valves and corner connectors for flexible installation.
CE Marking
The BeaconMedaes MAT-S manifold system is CE marked as a Class IIb medical device under the EU Medical Device Regulation 2017/745.
Specifications
  • Nominal Delivery/Line Output Pressure: Available in 4.0±0.1 bar, 8.0±0.1 bar, and 11.0±0.1 bar.
  • Volumetric Flow: Ranges from 495 l/min at 4.5 bar to 1800 l/min at 4 bar.
  • Input Pressure Range: 0-230 bar.
  • Regulator Test Pressure: As per ISO 10524-2 with a 10% pressure drop.
  • Changeover Pressure: Ranges from 10±1 bar to 20±1 bar.
Procedures and Configurations
  • Automatic Changeover Manifold Schematic: Includes components like dome regulators, solenoid valves, and pressure sensors.
  • Installation Proposal: Details on manifold weight, dimensions, and mounting requirements.
  • Free Standing Configuration: Describes a space-saving setup with specific parts and kits required.
Standards and Recommendations
  • ISO Standards: Compliance with ISO 10524-2 for regulator test pressures.
  • BS Standards: Bull nose tailpipes conform to BS standards for various gases.
Key Components and Part Numbers
  • Various part numbers are provided for different configurations and gas types, including oxygen, nitrous oxide, and medical air.
  • Specific kits and connectors are listed for manifold support and cylinder rack support.
Conclusion
The document outlines the technical details and configurations for the MAT-S manifold system, emphasizing compliance with international standards and providing detailed part numbers for ordering and assembly.
See more

Catalog excerpts

Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-1

Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version) Medical Gas Supply System Description and intended use The BeaconMedaes Medical Gas Supply System, mainly consists of cylinder manifold system, manifold changeover system, header & tailpipe and other necessary components, is intended to work in conjunction with other devices within the medical gas pipeline system (MGPS) to assist with maintaining continuity of supply throughout the hospital to local connection points for eventual introduction into the human body. The device controls the gas pressure within the specification as required by the MGPS. In all cases other devices outside of this scope are connected downstream of the MGPS are required to control the final gas pressure and flow requirements for safe supply to the patient. Therefore, this device is not intended to control supply parameter hazards directly to the patient, only to the MGPS. The device including Cylinder Manifold System and Manifold Changeover System (abbreviate as manifold below) is principally designed for use as a primary or secondary source of supply, or for emergency backup. The Medical Gas Supply System supplies one of the following medical gases to a piped distribution system, Oxygen, Nitrous Oxide, O2/N2O 50%: 50%, Medical Air, Surgical Air, Nitrogen & Carbon Dioxide. BeaconMedaes • No.45 Ximei Road, Xinwu District, Wuxi City, Jiangsu Province, 214111, China • www.beaconmedaes.com MAT-S Medical Gas Supply System

 Open the catalog to page 1
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-2

BeaconMedaes MAT-S Medical Gas Supply System The distinctive features of MAT-S A constant supply of medical gases such as oxygen, nitrous oxide, and medical air is critical to ensure safe, quality patient care. That is why we designed the new MAT-S Medical Gas Supply System to deliver superior reliability. Our medical gas supply system offers a higher utilization rate and uptime for using it more frequently and for longer periods. In addition, its smart features make it easy and efficient to operate the net result: is premium performance and durability and lower operational and maintenance costs....

 Open the catalog to page 2
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-3

SPECIFICATION Control System The control of the various changeover system conforms to NHS Health Technical Memorandum 06-01(HTM06-01) Electrical services supply and distribution. Following the Chapter 11 requirement, the manufacturer will provide evidence of Electromagnetic compatibility (EMC) for the manifold, e.g., an EMC test certificate. The pressure sensor monitors the pressure in each bank of cylinders and bank pressure gauges attached to each bank regulator (1st stage). A pressure sensor measures delivery pressure, and a delivery pressure gauge is attached to each line (2nd stage) regulator...

 Open the catalog to page 3
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-4

Output Flow MAT-S output flow is tested according to ISO 10524-2. MAT-S flow performance Volumetric Flow (l/min) Based on air and assuming sufficient cylinders are connected and zero pipeline pressure loss: Static = 0 l/min at 4.8 bar ‘X’ = 600 l/min at 4.5 bar ‘Y’ = 1080 l/min at 4.3 bar ‘Z’ = 1450 l/min at 4 bar Nominal delivery / Line output pressure MAT-S flow performance Volumetric Flow (l/min) Based on air and assuming sufficient cylinders are connected and zero pipeline pressure loss: Static = 0 l/min at 11.5 bar ‘X’ = 1250 l/min at 10.8 bar ‘Y’ = 1769 l/min at 10.5 bar ‘Z’ = 2050 l/min...

 Open the catalog to page 4
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-5

Typical Automatic Changeover Manifold Schematic Hospital Pipeline Relief Valve Exhaust Lockable Ball Valve Sampling Outlet Main Gas Line Control Gas Line Ball Valve Non-Return Valve Line Pressure Sensor Exhaust Line Vent Valve Service Valve Line Relief Valve Service Relief Valve Dome Regulator* Ball Valve Service Relief Valve Ball Valve Pressure Gauge * Pressure Gauge Ball Valve Ball Valve Service Relief Valve Header With Non-Return Valve Which Have Gas Specific Connection Intermediate Relief Valve Non-Return Valve Intermediate Relief Valve Service Relief Valve Non-Return Valve Pressure Gauge...

 Open the catalog to page 5
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-6

Typical Cylinder Header & Extension Layout Details Note : 1.Manifold weight is 35kg. 2.Dimension: Length*Width*height=460mm*200*1100 Headers (with rack and all components) / DETAIL OF MOUNTING BRACKET Gas specific nonreturn valves Cylinder Header Mounting Bracket Gas specific nonreturn valve Mounting Bracket Chain pin Wall mounting points NOTE - 3/8” plug and bonded seal is required on the last cylinder header on each bank. Cylinder Header Wall mounting points Note : 3/8" plug and bonded seal is required on the last cylinder header on each bank. Cylinder chain pin hook points Copper Washer Drawing...

 Open the catalog to page 6
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-7

Manifold Designation e.g.,MAT-S-HH-O2-A A-B-C-D-E Variable Allowable Value Allowable Value Description Model name Fully Automatic Manifold Smartbox & MyMedgas enabled Low-pressure liquid cylinder Input type Nominal Delivery / Line pressure Low Changeover Pressure Part No High Changeover Pressure Part No Headers (with rack and all components)/ both sides Gas Type *If request headers quantity is above 2x6 need to add up with available configuration. For example: for 2x7 may order 2x3 and 2x4. BeaconMedaes • No.45 Ximei Road, Xinwu District, Wuxi City, Jiangsu Province, 214111, China • www.beaconmedaes.co...

 Open the catalog to page 7
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-8

Tail pipe Gas Type Bull nose (ISO5145) Side entry Bull nose (BS341) Top entry Bull nose (BS341) Side entry Bull nose (BS341) Extended Chinese Bullnose Nitrogen (N2) Carbon Dioxide (CO2) Note: 1.Bull nose tailpipes (except Chinese type) are according to the following B.S. standards: Oxygen, Air, Nitrogen: BS: 341-1 No. 3; Carbon dioxide: BS: 341-1 No. 8; Nitrous oxide: BS: 341-1 No. 13. 2.Mixture N2O - O2 (registered trade name Entonox BOC) low-pressure cylinder “G” type has a Pin-indexed connector according to standard BS EN ISO 407, and 230 bar cylinder “E.W.” type has Bull nose connector according...

 Open the catalog to page 8
Medical Gas Supply System HTM02-01 / HTM2022 / ISO7396-1 MAT-S (HTM/ISO Version)-9

Free standing configuration Post for manifold support with Dimension shown below. The illustration below shows the new free standing option. The free standing assembly is a contained, optimised and space saving configuration. Main parts are: Descrition Note: To be supplied with kit [4109626500] Post for manifold support Post for Cylinder rack support Manifold Header Assembly Post for Cylinder rack support with Dimension shown below Descrition Part Number Post for manifold support Post for Cylinder rack support Drop Down Connector Post joint to Cylinder rack Note: To be supplied with kit [4109626600]...

 Open the catalog to page 9

All BeaconMedaes catalogs and technical brochures

  1. Central alarm

    4  Pages

  2. Magnis MSV

    2  Pages

  3. mVAC

    12  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.