# Medical Terminal Unit - Gem Shield BS 5682 SPECIFICATION BeaconMed/esPart of the Atlas Copco Group 8102341037.06 26/11/19 Page 1 of 6 Medical Terminal Unit (Gem Shield ) Medical terminal units are located at the final delivery points in medical gas piping systems to provide quick connection to the system. Terminal units provide for single handed insertion and removal of the medical gas probe. The medical gas terminal units shall conform to BS EN ISO 9170-1 and accept probes to BS5682. Terminal units shall be capable of single-handed insertion and removal of the medical gas probe. The anaesthetic gas scavenging (AGS) terminal unit shall conform to BS6834. Medical gas terminal units consist of two parts: First fix - the wall mounted assembly consists of brass pipeline termination block with copper stub pipe secured between a back plate and a gas specific plate to allow 130 degrees radial movement of the copper stub to align with the pipeline. The second fix plastic components shall be manufactured with the pin index permanently moulded into the gas specific socket. The socket assembly shall retain a capsule assembly, containing the check valve and probe 'O' ring seals. Second fix terminal units shall be supplied with the anti-rotation pin loose and bagged to be fitted as required. The socket is fixed to the base block with a design that insures profile is always in the correct position. Probes are gas specific connectors that couple to the medical gas terminal unit sockets to provide access to gas systems. The replaceable capsule assembly shall enable all working parts subject to wear through usage to be replaced as a factory tested assembly, thereby reducing maintenance time. Each termination block assembly shall be pressure tested by the pressure decay method. Gas Specificity Terminal units shall be gas specific and only accept the correct medical gas probe. Gas specific components shall be pin-indexed to ensure that a correct gas specific assembly is achieved so that in normal course of dismantling for repair or maintenance, parts from other gases cannot inadvertently be used. Wall mounted and bedhead terminal units shall incorporate an anti-rotation pin to engage with connected downstream medical equipment ensuring correct orientation. Materials All screws, probe roller pins, locking springs and the anti-rotation pin shall be manufactured from stainless steel. The second fix assembly shall incorporate three injection moulded parts in fire-retardant nylon 66. All wetted parts (except seals) shall be brass or copper. Copper stubs pipes shall be manufactured from phosphorous de-oxidised non-arsenical copper to BS EN 1412 grade CW024A, manufactured to metric outside diameters in accordance with BS EN 13348 R250 (half hard). All elastomeric seals shall be manufactured from Viton with a Shore hardness of 75. Antimicrobial Additive All user accessible parts, 2nd fix, gas ID ring, plaster box, trim plate and inks shall include Biomaster silver ion antimicrobial additive for inherent antimicrobial protection. Verimaster additive shall also be included to enable the end user to verify the product is a genuine BeaconMedss product. * Biomaster and Verimaster are registered products of Addmaster (UK). * Absolute pressure CE Marking The standard range of BeaconMedss Gem Medical Gas Terminal Units and Gem Conversions are 'CE' marked under the Medical Devices Directive 93/42/EEC with approval from notified body no. 0088 (Lloyd's Register Quality Assurance). Under this directive, the specified products are classified as Class IIb Medical Devices. Warranty The terminal units shall have a three-year warranty, subject to the recommended minimum routine maintenance operations being carried out by correctly competent persons. Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.beaconmedaes.com
Open the catalog to page 1
Gem Shield Terminal Unit Installation Dimensions BS EN ISO 9170-1 First fix "Revolve" Gas 1. Backplate includes two slotted fixing holes (bottom) in order to provide adjustment of the terminal within the bedhead unit. 2. First fix "Revolve" type is identical for BS5682 and DIN 13260-2 standard. 3. Gem Shield 2nd fix does not require disassembling during installation due to the fastener access points. Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.beaconmedaes.com
Open the catalog to page 2
Second Fix Gem Shield Terminal Unit (with Antimicrobial Protection) Second Fix Gem Shield Stainless steel. Thread Connection Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.beaconmedaes.com
Open the catalog to page 3
Range of First fixes There are many possible terminal unit configurations types available on request to suit almost any applications. Modular/Wall. Straight stub Pendant Rear NIST Retractable Pendant Flexible Pendant Hose Barb Rigid Pendant Curved Rear Stub 1. Flexible pendant supplied complete: first+second fix. Shroud supplied with hose assembly. 2. Terminal units with a barb connection are not recommended for UK market. 3. There are many possible terminal unit configurations types available on request to suit almost any application. Typical examples are barbed hose connection, DISS...
Open the catalog to page 4
Hose assemblies to suit GEM pendant outlets Gas Note: CO2 and N2 outlets include a NIST style connection as there is no defined probe size in the BS standard. Atlas Copco Ltd. trading as Atlas Copco Medical Unit 18 Nuffield Way, Abingdon, Oxfordshire, UK OX14 1RL www.beaconmedaes.com
Open the catalog to page 5
In an effort to continuously improve our products, the right is reserved to change the specification of the items described herein at any time. Please contact us for further information and up to date specifications. Accessories Part Number Drawings detailing different views and all dimensional information are available upon request - please refer to table above. Other documents Mounting Box Adaptor plate Terminal Unit Bezel and Plaster Shield kit Plug2 Trim Plate Kit Size Document number Other 3D images are available upon request - please contact our technical department. Jig Plates - for...
Open the catalog to page 6All BeaconMedaes catalogs and technical brochures
-
Series B Premium
3 Pages
-
Local alarm
4 Pages
-
Central alarm
4 Pages
-
Digital Medical Gas Alarms
4 Pages
-
Zone Valve Boxes
4 Pages
-
mVAC Medical Vacuum Systems
8 Pages
-
Medical Scroll Air Retrofit
4 Pages
-
Ceiling Columns
2 Pages
-
Medical Vacuum Filters
4 Pages
-
Claw Medical Vacuum
8 Pages
-
Hose Assemblies
2 Pages
-
Biomaster Additive
4 Pages
-
Medical Gas Outlets
4 Pages
-
Lifeline Manifolds
4 Pages
-
Medical Equipment Air
8 Pages
-
Medical Instrument Air
4 Pages
-
ZMED Medical Air Systems
6 Pages
-
Scroll Medical Air Systems
8 Pages
-
Medical Air Dryers
2 Pages
-
Magnis MSV
2 Pages
-
mVAC
12 Pages
-
Lifeline® MCS
5 Pages
-
VAV05D-060H-D_V
4 Pages
-
VHV05D-060H-D_V
4 Pages
-
Medical Gas Retrofit Alarm
2 Pages
Archived catalogs
-
Medical Gas Alarm Systems
4 Pages
-
Medical Gas Outlets_2009
4 Pages
-
Oxygen and Air Flowmeters
2 Pages
-
Suction & Oxygen Therapy
2 Pages
-
Medical Air Desiccant Dryers
2 Pages
-
Instrument Air White Paper
39 Pages
-
Z MED Medical Air Systems
6 Pages
-
Claw Vacuum System
6 Pages