1. Catalogs
  2. BeaconMedaes
  3. Medipoint Alarm Panels HTM/ISO Brochure
video corpo

Medipoint Alarm Panels HTM/ISO Brochure

Medipoint Alarm Panels HTM/ISO Brochure

Medipoint Alarm Panels HTM/ISO Brochure

Product catalog summary
Alarm Panels Overview
The document provides detailed information about BeaconMedæs alarm panels, focusing on line contact monitoring, pressure monitoring, and various types of alarm panels.

Line Contact Monitoring
Line contact monitoring circuits ensure the integrity of input sensors and wiring circuits.

Pressure Monitoring
High-specification pressure switches with dual double throw contacts monitor both high and low pressure. A miniature PCB with line contact monitoring is included in the switch cover.

Local Alarm Panels
The Medipoint 26 Medical Gas Area Alarm monitors up to six medical gas services for normal, high, and low pressure conditions, providing warnings of abnormal operating pressure.

Central Alarm Panels
The Medipoint 125 Medical Gas Central Alarm system monitors the complete medical gas system, displaying up to 5 gas sources and 20 individual point alarms.

Key Benefits
  • Fully populated LED light boards for additional gas monitoring.
  • Surface mounted components for reduced heat and extended product life.
  • Biomaster antimicrobial protection.
  • RoHs2 compliant boards and pressure switches.
  • CE certified and compliant with BS EN 60601-1 Safety Standards.

Surface Mount Technology
Advanced printed circuit boards with surface mount technology ensure cooler operating temperatures and increased reliability.

Universal Symbols and Security
Includes easy-to-understand symbols and anti-tamper proof fittings for authorized access only.

Adjustable Volume Control
Volume control feature allows setting the alarm volume to suit the installation area.

Expandable System
Features like pre-punched boxes and interchangeable legends allow for easy system expansion.

Simple Reprogramming
The Medipoint 26 Alarm can be easily reprogrammed on-site for modifications to the medical gas system.

Repeating Facility
Slave panels can be installed to repeat status conditions in other hospital areas.

Accessories
  • Finishing Bezel for semi-flush installation.
  • Signal Booster for excessive cable runs.
  • Remote Audible Speaker for additional alarm positioning.
  • Relay interface panel for signal conversion.
  • Cable for connecting pressure sensors and alarms.
  • Pressure Switch connection kits.

Biomaster Antimicrobial Additive
The alarm fascia includes an antimicrobial additive to reduce cross-infection, using Silver Ion technology for lasting protection.

Verimaster Safety Taggant
A safety taggant is included for quality control, ensuring the product is genuine and protected.
See more

Catalog excerpts

Medipoint Alarm Panels HTM/ISO Brochure-1

Alarm Panels for more information contact us

 Open the catalog to page 1
Medipoint Alarm Panels HTM/ISO Brochure-2

LOCAL ALARM PANELS The BeaconMed^s Medipoint 26 Medical Gas Area Alarm is designed to monitor up to six medical gas services in the Normal, High and Low pressure conditions. They are normally used to provide a warning of abnormal operating pressure downstream of departmental area valve service units. CENTRAL ALARM PANELS The BeaconMed^s Medipoint 125 Medical Gas Central Alarm system is designed to provide full monitoring of the complete medical gas system installation. An individual Medipoint 125 alarm panel displays up to 5 gas sources in the 'Normal' and 4 'stage' alarm conditions, or may be...

 Open the catalog to page 2
Medipoint Alarm Panels HTM/ISO Brochure-3

SURFACE MOUNT TECHNOLOGY The Medipoint alarm comes with advanced printed circuit boards using surface mount technology as standard. This allows cooler operating temperatures to be maintained, increasing the reliability and substantially lengthening service life. All boards are entirely touch proof, with any low voltage components or circuitry either shielded or mounted behind the board for added safety of the user. UNIVERSAL SYMBOLS Simple, easy to understand universal symbols are included for ease of operation. A unique anti-tamper proof fitting on the door panel is also provided to ensure only...

 Open the catalog to page 3
Medipoint Alarm Panels HTM/ISO Brochure-4

BIOMASTER ANTIMICROBIAL ADDITIVE The alarm fascia incorporates an antimicrobial additive to minimise cross infection. Independent testing proves a reduction against E coli and MRSA bacteria. Since the additive is included in the mixture of the plastic, the inherent properties last the lifetime of the product. Biomaster protected surface Biomaster Antimicrobial additives use a Silver Ion based technology. This prevents the growth of harmful bacteria, lasts the life-time of the product and is safe and non toxic. the bacteria can’t survive Biomaster binds to the cell wall disrupting growth Permanent...

 Open the catalog to page 4

All BeaconMedaes catalogs and technical brochures

  1. Central alarm

    4  Pages

  2. Magnis MSV

    2  Pages

  3. mVAC

    12  Pages

Archived catalogs

  1. Local alarm

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.